Skip to main content
. 2001 Oct;39(10):3750–3752. doi: 10.1128/JCM.39.10.3750-3752.2001

TABLE 3.

Comparison of PCR results with culture results

Protocol Result No. with culture result
Sensitivity Specificity Positive predictive value Negative predictive value
Positive Negative
PCR Positive 30 6 0.71
Amplicor (n = 573) Negative 12 525 0.99 0.83 0.93
PCR Positive 4 3 0.67
Silica membrane (n = 75) Negative 2 66 0.96 0.57 0.97