TABLE 10.

Reproducibility of the AMPLICOR and COBAS AMPLICOR HCV v2.0 testsa

No. of HCV RNA copies/ml Correct result AMPLICOR
COBAS AMPLICOR
No. of specimens tested % Correct result 95% CIb No. of specimens tested % Correct result 95% CI
0 Negative 153 99.3 96.4–99.9 171 100 97.9–100
200 Positive 38 100 90.1–100 43 100 91.8–100
300 Positive 38 100 90.8–100 42 100 91.6–100
500 Positive 38 97.4 86.2–99.9 43 97.7 87.1–99.9
50,000 Positive 38 100 90.8–100 43 100 91.8–100
a

Pooled data from four test sites (EU, UM, UW, and Roche Molecular Systems). Results for five PI panel members were excluded from the data analysis (one in AMPLICOR and four in COBAS AMPLICOR). 

b

CI, confidence interval.