TABLE 2.

Sample carryover test results of HBsAg NEXTa

Run Mean S/CO of:
S/CO difference between unprotected and protected negative samples P valueb (unprotected versus protected)
Protected negative High positive sample Unprotected negative
1 0.20 946.68 0.21 0.01 0.21
2 0.19 954.15 0.18 −0.01 0.54
a

Data were generated from two runs, each with 5 replicates of each sample. Protected negative and unprotected negative samples were run preceding and following the high-positive sample, respectively. S/CO, signal over cutoff.

b

P value was determined by t test. A P value of <0.05 was considered statistically significant.