Table 1:
Technique | Commercial Availability |
Equipment Cost | Lateral Spatial Resolution | Axial Spatial Resolution | Temporal Resolution | Sample penetration depth in standard setup | Specialized fluorophores | Difficulty of Data Analysis |
---|---|---|---|---|---|---|---|---|
STORM/PALM | High | Medium | ~20–50 nma | ~40–100 nmb | Seconds-Minutes | Several μm | Yes | High |
DNA-PAINT | Medium | Medium | ~20–50 nma | ~40–100 nmb | Minutes-Hoursc | Several μm | No | High |
ROSE/SIMFLUX | Low | Medium/High | ~5–20 nma | N/A (implemented in 2D) | Minutes-Hours | Several μm | Nod | High |
ROSE-Z | Low | Medium/High | ~2–5 nma | ~2–5 nm | Minutes-Hours | Several μm | Nod | High |
ModLoc | Low | Medium/High | ~3–5 nma | ~7–10 nm | Minutes-Hours | Several μm | Nod | High |
Expansion Microscopy | High | Low-Mediume | ~25–70 nmf | ~25–70 nm | Seconds-Minutes | Variableg | No | Low |
STED | High | High | ~20–50 nm | ~40–100 nmh | 10 ms-Minutesi | Typically 10–15 μm, up to 120 μmj | No | Low |
RESOLFT | Medium | High | ~20–50 nm | ~40–100 nm | Seconds-Minutesi | Several μm | Yes | Low |
Parallelized RESOLFT | Low | High | ~45–65 nmk | ~50–80 nmk | Seconds-Minutes | Several μm | Yes | Low-Medium |
MINFLUX | Medium | High | ~1–3 nm | ~1–3 nm | Down to 80 μsecondsi | Several μm | Yes | High |
SIM | High | Medium | ~100–150 nm | ~250–350 nm | Sub-seconds | Several μml | No | Medium |
TIRF-SIM | High | Medium | ~85–120 nm | ~100–200 nm | Sub-seconds | 200 nm | No | Medium |
ISIM | High | Medium | ~140–200 nm | ~300–400 nm | Down to 10 ms | 20 μm | No | Medium |
Airyscan | High | Medium | ~120–180 nm | ~350–450 nm | Milliseconds-Secondsi | Several μm | No | Low |
SOFI | High | Low | ~75–175 nmm | ~200–300 nmm | Sub-seconds-Seconds | Several μm | Yes | High |
SRRF | High | Low | ~60–150 nm | Diffractionlimited (no improvement) | Seconds | Several μm | no | high |
Values depend on many parameters, importantly the localization precision, which is dependent on the number of photons collected and the fluorophores used. Ranges correspond to most commonly used and bright fluorophores.
Values correspond to the most commonly used PSF engineering implementations (e.g. cylindrical lens) and can further be improved using interferometric approaches (iPALM, 4Pi-SMS and W-4Pi-SMSN)
Values are dependent on number of targets to image; with improved buffers and DNA-sequences this can be decreased to seconds-minutes
These methods can be implemented with STORM fluorophores or DNA-PAINT approaches. When implemented with DNA-PAINT, they do not require specialized fluorophores.
ExM can be done on a simple widefield microscope, but generally a confocal microscope is used due to sample size and thickness
70 nm refers to the most common ~4.5X sample expansion. With improved gel recipes or iterative expansions, larger expansions are possible, bringing resolution down to 25 nm
ExM imaging generally requires many Z-stacks based on increased sample size
Values correspond to most common implementation of 3D STED and can further be improved with the use of 4Pi geometry (isoSTED)
Temporal resolution inversely scales with the size of field of view.
Deep sample penetrance has been achieved with high N.A glycerol immersion lenses (Urban et al., 2011)
Lateral and axial resolution based on numbers reported in (Masullo et al., 2018) and (Bodén et al., 2021), respectively.
Up to 80 um depth has been demonstrated with adaptive optics (Lin et al., 2021).
The resolution here is based on 2nd to 4th order statistical analysis, the most common range for SOFI.