Table 4. Mean difference in G6PD activity for Intermediate and High controls by each Biosensor device between Phases A and B.

Low controls were not directly comparable between Phases as these were from different lots.

Device Intermediate: mean difference in U/gHb*, (95% CI, p) High: mean difference in U/gHb*, (95% CI, p)
1 -0.1 (-0.1 to 0.2, 0.481) 0.0 (-0.5 to 0.5, 0.993)
2 -0.3 (-0.4 to -0.1, <0.001) 0.1 (-0.3 to 0.5, 0.657)
3 -0.3 (-0.5 to -0.1, 0.016) -0.5 (-1.1 to 0.0, 0.050)
4 -0.3 (-0.5 to -0.2, <0.001) -1.5 (-2.1 to -1.0, <0.001)
5 -0.1 (-0.6 to 0.3, 0.559) -0.2 (-0.6 to 0.2, 0.399)
6 -0.3 (-0.4 to -0.2, <0.001) -0.2 (-0.7 to 0.2, 0.326)
7 0.1 (-0.1 to 0.3, 0.184) 0.0 (-0.4 to 0.4, 0.906)
8 -0.4 (-0.6 to -0.2, <0.001) -0.8 (-1.3 to -0.3, 0.003)
9 -0.2 (-0.3 to 0.0, 0.032) -0.1 (-0.7 to 0.5, 0.765)
10 -0.2 (-0.4 to -0.04, 0.018) -0.3 (-0.8 to 0.2, 0.224)
Pooled -0.2 (-0.3 to -0.1, <0.001) -0.4 (-0.5 to -0.2, <0.001)

*Phase A–Phase B