. 2022 Jan 7;35(1):9–20. doi: 10.1007/s10278-021-00553-z

Table 5.

Technical accuracy. Median luminal diameters, absolute deltas (Δ) in millimeter, and relative deltas in percent for aorta, main branch, and smallest procedure-relevant artery including the interquartile range (IQR). No significant differences were found between the printed models and their original CT (each p > 0.05) using Friedman’s test (1) and Dunn’s multiple comparison test for the diameters, as well as Wilcoxon matched-pair signed rank test for the deltas

Aorta Main branch Smallest artery
Diameter Median diameter p Median diameter p Median diameter p
Original CT 19.5 (± 6.3) mm 0.58 1 5.9 (± 2.4) mm 0.21 1 3.2 (± 1.8) mm 0.30 1
Clear resin (V4) 18.9 (± 5.3) mm 0.62 6.1 (± 2.6) mm  > 0.99 3.0 (± 1.8) mm 0.50
Flexible resin (80A) 19.0 (± 5.5) mm 0.94 5.9 (± 2.5) mm 0.22 3.0 (± 1.6) mm 0.30
Absolute delta Δ (mm) Median delta (mm) p Median delta (mm) p Median delta (mm) p
CT vs. clear resin 0.48 (± 0.55) mm 0.18 0.40 (± 0.36) mm 0.43 0.23 (± 0.25) mm 0.54
CT vs. flexible resin 0.67 (± 0.54) mm 0.30 (± 0.36) mm 0.18 (± 0.21) mm
Relative delta Δ (%) Median error (%) p Median error (%) p Median error (%) p
CT vs. clear resin 2.66 (± 2.70)% 0.15 6.31 (± 5.31)% 0.65 6.52 (± 10.28)% 0.56
CT vs. flexible resin 3.63 (± 3.17)% 6.27 (± 7.93)% 7.21 (± 8.46)%