. 2001 Dec;39(12):4380–4386. doi: 10.1128/JCM.39.12.4380-4386.2001

TABLE 6.

Comparison of Isolator system to BACTEC Plus Aerobic/F bottle

Potential pathogen(s) or contaminant(s) No. of isolates with the following result:
P valueb
Positive with Isolator system only Positive with BACTEC Plus Aerobic/F bottle only Positive with both Positive with neithera Total
Pathogens
 All microorganisms 75 128 245 130 578 0.0003
Staphylococcus aureus 7 17 69 8 101 0.0639
Staphylococcus spp., coagulase negative 10 41 49 39 139 0.0001
Streptococcus pneumoniae 0 4 13 2 19 0.125
 Viridans group streptococci 1 13 10 15 39 0.002
Enterococcus spp. 7 5 5 8 25 0.774
Enterobacteriaceae family 27 26 58 14 125 1.0
Pseudomonas spp. 1 3 14 1 19 0.625
 Obligately anaerobic bacteria 0 1 3 23 27 1.0
 Other bacteria 10 13 14 14 51 0.678
Mycobacterium tuberculosis complex 1 0 0 2 3 1.0
Candida spp. 8 5 10 3 26 0.581
Cryptococcus neoformans 1 0 0 1 2 1.0
Histoplasma capsulatum 2 0 0 0 2 0.50
Contaminants
 All microorganisms 133 9 0 16 158 0.0001
Bacillus spp. 20 0 0 4 24 0.0001
Corynebacterium spp., not C. jeikeim 15 4 0 6 25 0.019
Propionibacterium spp. 1 1 0 4 6 1.0
Staphylococcus spp., coagulase negative 74 4 0 2 80 0.0001
 Viridans group streptococci 6 0 0 0 6 0.031
 Other bacteria 17 0 0 0 17 0.0001
a

Refers to the number of isolates detected with only the BACTEC MYCO/F Lytic bottle or the BACTEC Anaerobic Lytic/10 bottle. 

b

Refer to Materials and Methods section for calculation of P values.