. 2001 Dec;39(12):4380–4386. doi: 10.1128/JCM.39.12.4380-4386.2001

TABLE 7.

Bloodstream infections detected by the Isolator system and BACTEC Plus Aerobic/F bottle

Potential pathogen(s) No. of bloodstream infections detected with:
P valuea
Isolator system only BACTEC Plus Aerobic/F bottle only Both
All microorganisms 54 88 145 0.005
Staphylococcus aureus 5 2 30 0.453
Staphylococcus spp., coagulase negative 9 34 35 0.0002
Streptococcus pneumoniae 0 3 8 0.25
Viridans group streptococci 1 10 5 0.012
Enterococcus spp. 7 3 3 0.344
Enterobacteriaceae family 17 16 40 1.0
Pseudomonas spp. 1 3 10 0.625
Obligately anaerobic bacteria 0 1 1 1.0
Other bacteria 6 12 8 0.24
Mycobacterium tuberculosis complex 1 0 0 1.0
Candida spp. 4 4 5 1.0
Cryptococcus neoformans 1 0 0 1.0
Histoplasma capsulatum 2 0 0 0.5
a

Refer to Materials and Methods section for calculation of P values.