Table 5.

Median volume computed tomography (CT) dose index (CTDIvol) by body region, patient age, manufacturer and phantom, and relative CTDIvol comparing 16-cm with 32-cm phantoms

Body region Age category CTDIvol in mGy
Canon GE Siemens
Phantom Relative dose Phantom Relative dose Phantom Relative dose
32-cm 16-cm 32-cm 16-cm 32-cm 16-cm
Brain <1 y
1–4 y 5 17 3.5
5–9 y 10 27 2.8
10–14 y 9 31 3.3 12 29 2.4
15–17 y 11 37 3.3 8 35 4.1
Skull < 1 y
1–4 y 3 17 5.8
5–9 y 6 11 1.8 2 18 7.9 3 11 4.4
10–14 y 13 27 2.1 4 20 4.6 4 14 3.4
15–17 y 17 29 1.7 13 23 1.8 11 15 1.4
Chest <1 y 1 4 2.5
1–4 y 2 6 3.1
5–9 y 2 7 3.1
10–14 y 4 10 2.3
15–17 y 6 9 1.5
Abdomen <1 y 2 3 1.7
1–4 y 2 6 2.5 2 4 2.0
5–9 y 4 7 1.9 4 5 1.2
10–14 y 6 8 1.4
15–17 y 6 11 1.8

Values are not shown when there were fewer than 5 CT examinations performed by age and body region using each phantom (numbers of scans can be derived from the n and percent values of Table 3). y years