Table 6.
WaveForm C-CGM System In-Clinic Accuracy Performance After Post Hoc Analysis With Hybrid Algorithm.
In-clinic (compared to YSI) |
|
---|---|
Number of samples | 17 381 |
Mean ARD (%) 100-400 mg/dL | 9.9 |
Median ARD (%) 100-400 mg/dL | 7.9 |
Mean AD (mg/dL) 40-100 mg/dL | 14.5 |
Median AD (mg/dL) 40-100 mg/dL | 11.3 |
C-CGM, Cascade continuous glucose monitoring; YSI, Yellow Spring Instruments.