TABLE 1:

Scan Parameters for Conventional Knee MRI and the qDESS Protocol

Sequence 2D
3D Sagittal qDESS
Axial Proton Density Fat-Saturated Coronal T1-Weighted Sagittal T2-Weighted Fat-Saturated Sagittal Proton Density Coronal Proton Density Fat-Saturated
TR/TE 4000/35 600/18 5441/54 4000/35 4000/35 18.6/5.9
Flip angle (°) 142 142 142 142 142 20
Echo-train length 9 3 11 9 9 1
No. of signals acquired 3 3 4 2 2 1
Bandwidth (± kHz) 50 50 50 50 50 41.7
FOV (mm) 140 × 140 140 × 140 140 × 140 140 × 140 140 × 140 160 × 160
Acquisition matrix 512 × 224 416 × 224 384 × 192 384 × 224 384 × 224 416 × 512
Resolution (mm2) 0.27 × 0.63 0.34 × 0.63 0.36 × 0.72 0.36 × 0.63 0.36 × 0.63 0.34 × 0.27
Slice thickness (mm) 3 2.5 2.5 2.5 2.5 1.6
Slice gap (mm) 0.3 0 0 0 0 0
No. of slices 36 44 40 40 44 80
Parallel imaging 2 × 1 NA 2 × 1 NA 2 × 1 2 × 1
Scan time (min:s) 2:56 2:18 3:43 3:28 2:00 5:00

Note—qDESS = quantitative double-echo steady-state, NA = not applicable.