. 1999 Apr;37(4):944–949. doi: 10.1128/jcm.37.4.944-949.1999

TABLE 3.

Summary of sensitivities and specificities of several automated bacterial identification systems tested with food-borne pathogens

Pathogen and system % Sensitivitya % Specificityb
Listeria species
 WalkAway 40 100 100
 MIS 90 100
 VITEK 97.5 100
 MicroLog 97.5 100
S. aureus
 WalkAway 40 97.5 100
 MIS 47.5 100
 VITEK 95 95
B. cereus
 MIS 55 97.5
 VITEK 42.5 97.5
C. jejuni, MIS 72.5 32.5
Salmonella species
 WalkAway 40 97.5 100
 MIS 85 100
 VITEK 100 100
 Replianalyzer 95 100
 MicroLog 95 100
E. coli
 WalkAway 40 100 100
 MIS 52.5 97.5
 VITEK 100 97.5
 Replianalyzer 90 100
 MicroLog 72.5 100
a

Proportion of reference positive strains which were correctly identified with an acceptable confidence rating. 

b

Proportion of reference negative strains which were not identified as the pathogen of concern with an acceptable confidence rating.