Table 1.

Treatment planning system beam modeling parameters requested via IROC Houston surveys and their range of dose effects (based on the reported spread in values), as previously determined by phantom dose calculations, for a common base Varian linac model equipped with Millennium120 MLC (e.g. Trilogy, 2100iX, etc.) using 6 MV photons [13,20].

TPS Parameter Estimated Dose Effects
2.5th Percentile 10th Percentile 90th Percentile 97.5th Percentile
Parameter Value Dose Effect (vs. 50th percentile) Parameter Value Dose Effect (vs. 50th percentile) Parameter Value Dose Effect (vs. 50th percentile) Parameter Value Dose Effect (vs. 50th percentile)
Eclipse AAA
Effective Target Spot Size X and Y [mm] 0.0000 0.0% 0.0000 0.0% 0.5000 0.0% 1.0000 0.0%
MLC Transmission Factor 0.0118 −1.1% 0.0134 −0.7% 0.0200 +0.8% 0.0200 +0.8%
Dosimetric Leaf Gap [cm] 0.1000 −3.6% 0.1388 −1.5% 0.2000 +1.2% 0.2300 +2.8%
RayStation
Primary Source X Width and Y Width [cm] 0.05000 0.0% 0.04000 0.0% 0.09700 0.0% 0.12345 0.0%
MLC Transmission 0.0070 −4.0% 0.0070 −4.0% 0.0250 +2.3% 0.0250 +2.3%
Tongue and Groove [cm] 0.0100 +1.1% 0.0100 +1.1% 0.0500 −0.3% 0.0500 −0.3%
Leaf Tip Width [cm] 0.1770 −1.6% 0.1860 −1.4% 0.5000 +1.9% 0.5000 +1.9%
MLC Position Offset [cm] 0.0000 −3.6% 0.0000 −3.6% 0.1160 +6.7% 0.1160 +6.7%
MLC Position Gain 0.0000 0.0% 0.0000 0.0% 0.0150 0.0% 0.0150 0.0%
MLC Position Curvature [1/cm] 0.0000 0.0% 0.0000 0.0% 0.0010 +0.2% 0.0010 +0.2%