Table 1.
Results of heating tests of the needle shims at a position 15 cm right of isocenter in two orientations, perpendicular to B0 (0,0,0 degrees) and parallel to B0 ( 0,−90,0 degrees) (A) With shim insert inside needle (B) Only shim insert, without the needle, (C): In-scanner Reference with no needle or shim insert, (D) On the bench, outside the scanner with temperature probes directly on shim coils and the shim insert immersed vertically at the center of the phantom.
A. Temperature Rise in °Celsius With Needle (Probes on Needle Surface) | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Angle | 3D FSE | Continuous 2D SSFP | ||||||||||||
No Current | 1A in CN0 | 1A in CN90 | No Current | 1 A in CN0 | 1 A in CN90 | |||||||||
P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | |||
0,0,0. (Perpendicular to B0) | 1.50 | 0.30 | 1.60 | 0.50 | 1.50 | 0.76 | 0.60 | 0.30 | 0.80 | 0.60 | 0.90 | 0.70 | ||
0, −90,0. (Parallel to B0) | 1.90 | 0.90 | 2.79 | 1.47 | 2.00 | 1.00 | 1.17 | 0.51 | 1.40 | 0.79 | 1.34 | 0.63 | ||
B. Temperature Rise in °Celsius Without Needle (Probes on shim coils directly) | ||||||||||||||
Position | Angle | 3D FSE | Continuous 2D SSFP | |||||||||||
No Current | 1A in CN0 | 1A in CN90 | No Current | 1 A in CN0 | 1 A in CN90 | |||||||||
P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | |||
0,0,0. (Perpendicular to B0) | 1.67 | 1.74 | 2.34 | 2.60 | 2.20 | 2.49 | 1.63 | 1.29 | 1.9 | 1.46 | 1.76 | 1.60 | ||
0, −90,0. (Parallel to B0) | 1.44 | 0.43 | 2.34 | 1.24 | 2.04 | 0.76 | 1.14 | 0.36 | 1.70 | 0.6 | 1.51 | 0.70 | ||
C. in-scanner Reference No needle or shim Insert, No Current P1 at center, P2 at off-center position |
D. Phantom On Bench Reference Probes on shim coils. |
|||||||||||||
3D FSE | 2D SSFP | 1 A in CN0 | 1 A in CN90 | |||||||||||
P1 | P2 | P1 | P2 | P1 | P2 | P1 | P2 | |||||||
0.20 | 1.11 | 0.20 | 0.39 | 0.21 | 0.23 | 0.27 | 0.40 |