Published in final edited form as: Magn Reson Med. 2021 Aug 19;87(1):541–550. doi: 10.1002/mrm.28977

Table 1.

Results of heating tests of the needle shims at a position 15 cm right of isocenter in two orientations, perpendicular to B0 (0,0,0 degrees) and parallel to B0 ( 0,−90,0 degrees) (A) With shim insert inside needle (B) Only shim insert, without the needle, (C): In-scanner Reference with no needle or shim insert, (D) On the bench, outside the scanner with temperature probes directly on shim coils and the shim insert immersed vertically at the center of the phantom.

A. Temperature Rise in °Celsius With Needle (Probes on Needle Surface)
Angle 3D FSE Continuous 2D SSFP
No Current 1A in CN0 1A in CN90 No Current 1 A in CN0 1 A in CN90
P1 P2 P1 P2 P1 P2 P1 P2 P1 P2 P1 P2
0,0,0. (Perpendicular to B0) 1.50 0.30 1.60 0.50 1.50 0.76 0.60 0.30 0.80 0.60 0.90 0.70
0, −90,0. (Parallel to B0) 1.90 0.90 2.79 1.47 2.00 1.00 1.17 0.51 1.40 0.79 1.34 0.63
B. Temperature Rise in °Celsius Without Needle (Probes on shim coils directly)
Position Angle 3D FSE Continuous 2D SSFP
No Current 1A in CN0 1A in CN90 No Current 1 A in CN0 1 A in CN90
P1 P2 P1 P2 P1 P2 P1 P2 P1 P2 P1 P2
0,0,0. (Perpendicular to B0) 1.67 1.74 2.34 2.60 2.20 2.49 1.63 1.29 1.9 1.46 1.76 1.60
0, −90,0. (Parallel to B0) 1.44 0.43 2.34 1.24 2.04 0.76 1.14 0.36 1.70 0.6 1.51 0.70
C. in-scanner Reference
No needle or shim Insert, No Current
P1 at center, P2 at off-center position
D. Phantom On Bench Reference
Probes on shim coils.
3D FSE 2D SSFP 1 A in CN0 1 A in CN90
P1 P2 P1 P2 P1 P2 P1 P2
0.20 1.11 0.20 0.39 0.21 0.23 0.27 0.40