Abstract
Anemia, characterized by low blood hemoglobin level, affects about 25% of the world’s population with the heaviest burden borne by women and children. Anemia leads to impaired cognitive development in children, as well as high morbidity and early mortality among sufferers. Anemia can be caused by nutritional deficiencies, oncologic treatments and diseases, and infections such as malaria, as well as inherited hemoglobin or red cell disorders. Effective treatments are available for anemia upon early detection and the treatment method is highly dependent on the cause of anemia. There is a need for point-of-care (POC) screening, early diagnosis, and monitoring of anemia, which is currently not widely accessible due to technical challenges and cost, especially in low- and middle-income countries where anemia is most prevalent. This review first introduces the evolution of anemia detection methods followed by their implementation in current commercially available POC anemia diagnostic devices. Then, emerging POC anemia detection technologies leveraging new methods are reviewed. Finally, we highlight the future trends of integrating anemia detection with the diagnosis of relevant underlying disorders to accurately identify specific root causes and to facilitate personalized treatment and care.
Graphical Abstract
In this critical review, emerging non-invasive and minimally invasive point-of-care anemia detection technologies are discussed, highlighting future opportunities and the need for multiplexed approaches and integrated disease etiology identification.
1. Introduction
1.1. Global Scope of Anemia
Hemoglobin (Hb) is an iron-containing oxygen-transport metalloprotein in red blood cells (RBCs) that carries oxygen to tissues and transports carbon dioxide back to the lungs [1, 2]. Anemia, defined as low blood hemoglobin level, is a major global health problem. This disease affects about 25% of the world’s population, with the heaviest burden borne by women and children, in countries throughout the world across all income levels [3]. Anemia has significant adverse health consequences and negative impact on social and economic development, which include reduced performance in school and work productivity [3]. Anemia causes numerous symptoms including weakness, fatigue, and dizziness in milder cases as well as life-threatening cardiovascular collapse in more severe cases [4, 5].
The Institute for Health Metrics and Evaluation reported, in the Global Burden of Disease (GBD) Study conducted in 2013, that anemia affected 27.0% of the world’s population (1.93 billion people) [6]. Based on data collected between 1993 and 2005, anemia affected 41.8% of pregnant women, 30.2% of non-pregnant women, 47.4% of preschool children, and 25.4% of school age children (which equates to 56.4 million, 468.4 million, 293.1 million, and 304.6 million people in each sub-group respectively) [7]. Detailed definition of anemia has been provided by the WHO based on Hb level thresholds that are categorized into 5 distinct groups: adult males, adult non- pregnant females, adult pregnant females, male children, and female children depending on subject’s gender, age, and pregnancy status (Table 1) [1].
Table 1.
Hemoglobin thresholds (g/dL) by age, sex, and pregnancy status used in classifying anemia cases as mild, moderate, and severe in Global Burden of Disease 2013
| Severity of Anemia | Severity of Anemia | ||
|---|---|---|---|
| Variable | Mild | Moderate | Severe |
| Age < 1 mo | |||
| Males | 13.0 – 14.9 | 9.0 – 12.9 | < 9.0 |
| Females | 13.0 – 14.9 | 9.0 – 12.9 | < 9.0 |
| Age 1 mo – 5 y | |||
| Males | 10.0 – 10.9 | 7.0 – 9.9 | < 7.0 |
| Females | 10.0 – 10.9 | 7.0 – 9.9 | < 7.0 |
| Age 5 – 14 y | |||
| Males | 11.0 – 11.4 | 8.0 – 10.9 | < 8.0 |
| Females | 11.0 – 11.4 | 8.0 – 10.9 | < 8.0 |
| Age ≥ 5 y | |||
| Males | 11.0 – 12.9 | 8.0 – 10.9 | < 8.0 |
| Females, non-pregnant | 11.0 – 11.9 | 8.0 – 10.9 | < 8.0 |
| Females, pregnant | 10.0 – 10.9 | 7.0 – 9.9 | < 7.0 |
| Disability weights | |||
| All cases | 0.005 (0.002–0.011) | 0.058 (0.038–0.086) | 0.164 (0.112–0.228) |
Table adopted and modified from The World Health Organization (WHO) reference [1]
1.2. Etiology of Anemia
The etiology of anemia can range from nutritional deficiencies (e.g., iron deficiency anemia) [8], infections (e.g., malaria, hookworm disease, HIV) [9, 10], chronic medical conditions that indirectly cause anemia (e.g., chronic kidney disease, inflammatory/ autoimmune disorders) [9, 10], hemoglobin or red cell disorders (e.g., sickle cell disease (SCD), thalassemia, myelodysplastic syndromes) [4], and pharmaceutical drug treatment (e.g., cancer chemotherapy) [4, 5]. A more extensive list of anemia etiologies is summarized in Table 2 by Kassebaum et. al. [11]. Among the major causes listed in Table 2, iron deficiency, malaria, and hemoglobin disorders constitute the leading causes for anemia.
Table 2.
Major Causes for Anemia
| Causes for which allocation of the anemia envelope was based on prevalence results from GBD 2010* |
| Malaria |
| Hookworm |
| Schistosomiasis |
| Peptic ulcer disease |
| Gastritis and duodenitis |
| Maternal hemorrhage |
| Sickle cell disorders |
| Homozygous sickle cell disease (HbS-HbS or Hb SS) |
| Heterozygous sickle cell disease (HbS-HbA or Hb SA) |
| Compound heterozygous sickle cell disease |
| Compound heterozygous sickle cell disease-thalassemia (HbS-Hbβ or Hb Sβ) |
| G6PD deficiency |
| Homozygous class I |
| Homozygous class II |
| Heterozygous (female only) |
| Thalassemias |
| β thalassemia major (Hbβ-Hbβ) |
| Heterozygous β thalassemia (Hbβ-HbA) |
| Compound heterozygous “E-β” thalassemia I (HbE-Hbβ) |
| Heterozygous hemoglobin E (HbE-normal) |
| Hemoglobin H disease (genotype = [− −/−alpha]) |
| CKD (Chronic Kidney Disease) |
| Due to diabetes mellitus |
| Due to hypertension |
| Other and unspecified |
| Causes for which allocation of anemia envelope was based on systematic redistribution methods* |
| IDA (Iron-deficiency anemia) |
| Other infectious disease |
| Other neglected tropical diseases |
| Other hemoglobinopathies and hemolytic anemias |
| Uterine fibroids |
| Other gynecologic disorders |
| Other endocrine, nutrition, blood and immune disorders |
Table adopted and modified from reference by Kassebaum et. al. 2014 [13].
Iron deficiency anemia can be caused by low dietary iron intake, chronic blood loss or loss of blood due to intestinal worm colonization (i.e., hookworm infection), and iron malabsorption, which is most prevalent in low- and middle-income countries [8, 12]. In addition, functional iron deficiency can also be caused by cancer due to inflammation induced sequestration/decreased utilization of iron and the blood loss from tumor sites, malignant invasion of normal tissue and bone marrow [13]. Malaria is also a major cause of anemia in endemic areas [14]. In high transmission settings, malaria induces anemia especially in young children. At all levels of transmission, malaria has been found to contribute to maternal anemia during pregnancy as well as to poor birth outcomes [14]. Inherited hemoglobin disorders such as SCD and thalassemia can cause hemolysis of red blood cells (RBCs) and severe, chronic anemia [11, 15].
Anemia can be mitigated by early diagnosis followed by timely intervention [16]. Effective treatment of anemia requires timely and accurate identification of the specific cause [17–19]. Misdiagnosis of the cause of anemia will lead to the incorrect intervention, which may result in severe clinical events [9]. For example, iron deficiency anemia can be treated by providing micro-nutrient supplementation. However, the WHO recommends that iron supplements be used only where malaria prevention and control systems are available [20]. Additionally, patients with hemoglobinopathies are at the risk of ron overload from hemolysis and recurrent multiple transfusions in patients with SCD, as well as from imbalanced iron homeostasis in patients with thalathemia [21–23]. Consequently, there is a significant need for rapid and accurate diagnosis of anemia and identification of the underlying disease etiology.
1.3. Anemia Detection and Anemia Cause Identification – Tests within Centralized Laboratory Environment and Unmet Needs at the Point of Care
Optimal management of anemia requires early diagnosis and monitoring using blood tests that measure hemoglobin levels [11, 24]. Even in high resource nations, anemia remains a major public health issue. Based on data collected between 2003 & 2012 for the US population, it was estimated 5.6 % met the criteria for anemia and 1.5% could be classified as moderate-severe anemia [25]. Laboratories in hospitals and clinical offices are utilized to aid in identifying the underlying cause of anemia in an individual patient [26, 27]. The current gold standard in laboratories for anemia detection and hemoglobin level measurement is complete blood count (CBC) using an automated hematology analyzer (AHA) on a venous blood sample in a centralized clinical hematology laboratory environment [26, 27]. The current gold standard laboratory methods for detecting 1) iron deficiency, 2) malaria, and 3) hemoglobin variant are 1) Levels of serum ferritin and transferrin, mean corpuscular volume, mean corpuscular Hb level [12], 2) microscopy visualization [28], and 3) High Performance Liquid Chromatography (HPLC), respectively. These detection methods are expensive and labor intensive, typically taking an extended period of time before results are available and are not accessible to patients all over the world, especially in low- and middle resource settings, where anemia is the most prevalent. In a 2019 report, the WHO listed diagnosis and monitoring of anemia, iron deficiency, malaria and SCD as essential in vitro diagnostic (IVD) tests for primary care use in low and middle income countries [29, 30]. POC technologies have been developed for detecting iron deficiency [31–33], malaria [34–36], and SCD [37–39] and can be found in previous published review articles.
The purpose of this review is to provide a useful comprehensive resource to researchers working on developing novel, clinically useful POC technologies. We aim to provide an overview of the methodologies for measuring Hb level and to review the current commercially available as well as emerging technologies for anemia detection. Section 2 introduces a list of methods and theoretical foundations for minimally-invasive and non-invasive anemia detection. Section 3 reviews current commercially available technologies (both minimally invasive and non-invasive) and provides a summary of the reported performance from recent publications. Section 4 reviews minimally invasive technologies (4.1) using electrical impedance (4.1.1) and optical detectors (4.1.2), as well as non-invasive technologies using optical detectors including smartphone-based technologies (4.2). In Section 5, we provide recommendations for the development and testing of future POC anemia detection platforms. The objective of this review is to provide a summary of (1) major Hb measurement methods, (2) current commercially available (minimally invasive and non-invasive) technologies and (3) emerging technologies (minimally invasive and non-invasive) for anemia detection. Based on the reviewed literature, we attempt to recommend practical solutions to develop and test integrated technologies for screening, monitoring and diagnosis of anemia and identification of the major underlying causes for the disease with high reproducibility, sensitivity, and specificity.
2. Methods
Combinations of key words including ‘anemia’, ‘anaemia’, ‘hemoglobin’, ‘haemoglobin’, ‘point-of-care’, ‘microfluidics’, ‘non-invasive’, ‘near-infrared’, and ‘photoplethysmography’ were searched using Google Scholar and PubMed. For commercially available technologies, the maximum and minimum reported value of sensitivity, specificity and Bias ± standard deviation for each technology were summarized along with the test population, threshold Hb level used for determining anemia, and reference method (Table 3). Commercially available technologies evaluated by less than 3 publications were introduced but not included in the performance summary. Similar information was also summarized for the emerging technologies reported after 2015, although only single set of test result was available for most of these studies (Table 4). Finally, we summarized test time and cost for commercially available technologies (Table 5).
Table 3.
Performance of commercially available technologies for anemia detection
| Technology | Lowest Sensitivity Reported | |||
|---|---|---|---|---|
| Reported Sensitivity at (Indicated Hb cutoff level) | Test Population | Reference Method | Reference | |
| WHO Hemoglobin Color Scale | 70.9% (11.0g/dL) | Pregnant women (194 subjects) | Hematology analyzer (Advia 2120i) | [69] |
| HemoCue® 201 | 56.00% (12.0g/dL) | Blood donor (969 subjects) | Hematology analyzer (ABX Pentra 60) | [80] |
| HemoCue® 301 | 25.0% (Male: 13.5 g/dL), 22.6% (Female: 12.5 g/dL) | Blood donor (553 subjects) | Hematology analyzer (Advia 2120i) | [81] |
| Pronto-7® | 54.5% (Male: 13.5 g/dL), 65.2% (Female: 12.5 g/dL) | Blood donor (553 subjects) | Hematology analyzer (Advia 2120i) | [81] |
| NBM-200 | 33.70% (11.0g/dL) | Pregnant women (269 subjects) | Hematology analyzer (Sysmex XP-100) | [122] |
| Haemospect | 66.7% (Male: 13.5 g/dL and Female: 12.5 g/dL) | Blood donor (353 subjects) | Hematology analyzer (Sysmex XE-2100) | [234] |
| Technology | Highest Sensitivity Reported | |||
| Reported Sensitivity at (Indicated Hb cutoff level) | Test Population | Reference Method | Reference | |
| WHO Hemoglobin Color Scale | 99.3% (12.0g/dL) | Patients attending the general outpatient department (200 subjects) | Cyanmethemoglobin method | [65] |
| HemoCue® 201 | 93% (11.0g/dL) | Pregnant women (102 subjects) | Hematology analyzer (Sysmex XS 1000i) | [83] |
| HemoCue® 301 | 90% (11.0g/dL) | Pregnant women (102 subjects) | Hematology analyzer (Sysmex XS 1000i) | [83] |
| Pronto-7® | 93% (Male: 13.5 g/dL and Female: 12.5 g/dL) | Blood donor (445 subjects) | Cell Counter (Beckman Coulter) | [105] |
| NBM-200 | 96.36% (12.5g/dL) | Blood donor (445 subjects) | Hematology analyzer (Beckman LH500) | [77] |
| Haemospect | 94.60% (12.5g/dL) | Blood donor (161 subjects) | Hematology analyzer (Advia 2120i) | [115] |
| Technology | Lowest Specificity Reported | |||
| Reported Specificity at (Indicated Hb cutoff level) | Test Population | Reference Method | Reference | |
| WHO Hemoglobin Color Scale | 27.69% (12.0 g/dL) | Patients attending the general outpatient department (200 subjects) | Cyanmethemoglobin method | [65] |
| HemoCue® 201 | 76% (11.0 g/dL) | Pregnant women (102 subjects) | Hematology analyzer (Sysmex XS 1000i) | [83] |
| HemoCue® 301 | 80% (11.0 g/dL) | Pregnant women (102 subjects) | Hematology analyzer (Sysmex XS 1000i) | [83] |
| Pronto-7® | 75.4% (Male: 13.8 g/dL and Female: 11.6 g/dL) | Convenience sample of patients in the emergency department (350 subjects) | Hematology analyzer (Tosoh G8) | [101] |
| NBM-200 | 91.43% (12.5 g/dL) | Blood donor (506 subjects) | Hematology analyzer (Beckman LH500) | [77] |
| Haemospect | 0.00% (12.5 g/dL) | Blood donor (161 subjects) | Hematology analyzer (Advia 2120i) | [115] |
| Technology | Highest Specificity Reported | |||
| Reported Specificity at (Indicated Hb cutoff level) | Test Population | Reference Method | Reference | |
| WHO Hemoglobin Color Scale | 49.1% (11.0 g/dL) | Pregnant women (643 subjects) | Hematology analyzer (Advia 2120i) | [69] |
| HemoCue® 201 | 99.80%(12.0 g/dL) | Blood donor (969 subjects) | Hematology analyzer (ABX Pentra 60) | [80] |
| HemoCue® 301 | 99.3% (Male: 13.5 g/dL), 98.9% (Female: 12.5 g/dL) | Blood donor (553 subjects) | Hematology analyzer (Advia 2120i) | [81] |
| Pronto-7® | 95% (11.5 g/dL) | Healthy children (110 subjects) | Hematology analyzer (Coulter LH780) | [109] |
| NBM-200 | 99.20% (12.5 g/dL) | Blood donor (993 subjects) | Hematology analyzer (Sysmex KX-21N) | [110] |
| Haemospect | 77.10% (12.5 g/dL) | Blood donor (353 subjects) | Hematology analyzer (Sysmex XE-2100) | [234] |
| Technology | Lowest Bias±SDD Reported | |||
| Reported Bias±SDD at (Indicated Hb cutoff level) | Test Population | Reference Method | Reference | |
| WHO Hemoglobin Color Scale | 0.2±1.96 (13.1 g/dL) | Patients in secondary care hospital (100 subjects) | Cyanmethemoglobin method | [71] |
| HemoCue® 201 | 0.56±0.099 (12.0 g/dL) | Pregnant women (969 subjects) | Hematology analyzer (ABX Pentra 60) | [80] |
| HemoCue® 301 | − 0.25±0.85 (11.0 g/dL) | Pregnant women (102 subjects) | Hematology analyzer (Sysmex XS 1000i) | [83] |
| Pronto-7® | 0.1±0.7 (12.5g/dL) | Blood donor (993 subjects) | Hematology analyzer (KX-21N) | [110] |
| NBM-200 | 0.18±0.09 (11.0 g/dL) | Blood donor (110 subjects) | Hematology analyzer (KX-21) | [90] |
| Haemospect | − 0.23±1.21 (Male: 13.5 g/dL and Female: 12.5 g/dL) | Blood donor (1483 subjects) | Hematology analyzer (Tosoh G8) | [114] |
| Technology | Highest Bias±SDD Reported | |||
| Reported Bias±SDD at (Indicated Hb cutoff level) | Test Population | Reference Method | Reference | |
| WHO Hemoglobin Color Scale | 1.121±1.87 (12.0 g/dL) | Patients attending the general outpatient department (200 subjects) | Cyanmethemoglobin method | [65] |
| HemoCue® 201 | − 0.53 ±1.01(11.0 g/dL) | Pregnant women (102 subjects) | Hematology analyzer (Sysmex XS 1000i) | [83] |
| HemoCue® 301 | 0.43±1.243 (Male: 13.5 g/dL and Female: 12.5 g/dL) | Blood donor (553 subjects) | Hematology analyzer (Advia 2120i) | [81] |
| Pronto-7 ® | − 0.59±1.98 (Male: 13.5 g/dL and Female: 12.5 g/dL) | Blood donor (553 subjects) | Hematology analyzer (Advia 2120i) | [81] |
| NBM-200 | 0.7±0.8 (12.5 g/dL) | Blood donor (993 subjects) | Hematology analyzer (KX-21N) | [110] |
| Haemospect | − 0.6±1.96 (12.5 g/dL) | Blood donor (353 subjects) | Hematology analyzer (Sysmex XE-2100) | [234] |
Table 4.
Performance analysis of emerging anemia detection technologies
| Technology | Sensitivity | Specificity | R2, r and p value | Reference method | Threshold value | Test subjects | Reference |
|---|---|---|---|---|---|---|---|
| Computer vision and deep learning assisted electrophoresis | 100% | 92.30% | p < 0.05 | Complete blood count (Not Specific) | 11.0 g/dL | Clinical and healthy patient samples (46 subjects) | [147] |
| Multi-Model Stacking Regressor | N/A | N/A | r=0.81, p<0.01 | N/A | N/A | Blood donors (1583 subjects) | [193] |
| Physical examination | 64.30% | 83.60% | p = 0.04 | Automated cell counter (Not Specific) | 11 g/dL | Physicians, residents and students in hospital (10Physicians, 18 residents and 16 students) | [159] |
| Simple Paper-based Assay | 87.50% | 100% | r = 0.909 | N/A | 11 g/dL | Patients covering a wide spectrum of anemia (32 subjects) | [235] |
| Computer vision techniques | 100% | 82.40% | N/A | N/A | 13 g/dl | Recruited random participants (77 subjects) | [198] |
| Digital Photographs of the Conjunctiva | 74% (Apple iPhone 5S), 57% (Panasonic DMC-LX5) | 71% (Apple iPhone 5S), 83% (Panasonic DMC-LX5) | R2=0.397, p< 0.0001 (Panasonic DMC LX5); p< 0.0001 (Apple iPhone 5S); | Hematology analyzer (Sysmex XE-2100) | 11.0 g/dL | Patients in Wellington Blood and Cancer Centre (47 Subjects) | [203] |
| A Kalman Filtering and Nonlinear Penalty Regression Approach | 76.19% | 80.85% | N/A | N/A | 11 g/dL | N/A | [200] |
| Digital Camera-Based Spectrometry | N/A | N/A | r = 0.96, P<0.0001 | Hematology analyzer (KX-21NTM) | N/A | Patient samples from the outpatient department in Nil Ratan Sircar Medical College and Hospital, Kolkata (300 subjects) | [141] |
| Medical Imaging, m-Health and Emerging Communication Systems | 83% | 82% | N/A | N/A | 12 g/dL | Blood donor (33 subjects) | [236] |
| Combined reflectance spectroscopy and stochastic modeling approach | 86% | N/A | N/A | N/A | N/A | Patients were enrolled at the Emergency Department (32 subjects) | [237] |
| Machine learning techniques with the PPG signal’s characteristics features | N/A | N/A | r=0.98918 (Classification and regression trees), r=0.96729 (Support vector regression), p < 0.01 | Hemocue Hb-201TM | N/A | Blood donor (33 subjects) | [192]. |
| Spectrophotometry | N/A | N/A | r=0.77, P<0.001 | Hematology analyzer (SP 1000i®) | N/A | Adult patients scheduled for potentially hemorrhagic major urologic surgery (44 subjects) | [238] |
| Smartphone app with Artificial Neural Network | 94% | 96% | R2 = 0.93 | N/A | N/A | Patients with an age range from 20 to 56 years (75 subjects) | [210] |
| Dynamic Spectrum | N/A | N/A | R2=0.8399 (Experimental group), R2=0.6686 (Control group) | N/A | N/A | Volunteers at the Physical Examination Center of Tianjin People’s Hospital (187 subjects) | [180] |
| Photoplethysmography System | N/A | N/A | R2=0.43, r=0.61, p< 0.001 (partial least squares); R2=0.42, r=0.62, p< 0.001 (backpropagation artificial neural network); | Hematology analyzer (Sysmex XS-1000i) | 12g/dL | Blood donor (238 subjects) | [185] |
| Smartphone app with patient-sourced photos | 97% | 76% | r =0.82 | Hematology analyzer (Advia 2120i) | 12.5 g/dL | Patients with various anemia etiologies (265 subjects) | [239] |
| Micro-Hct using HemataSTAT II | 39.50% | 93.20% | r = 0.533, p < 0.001 | Hematology analyzer (ABX Pentra 60) | 12.0 g/dL | Unselected potential female donors (969 subjects) | [80] |
| 3D printed auto-mixing chip | 81.2% (Severe anemia),100% (Mild anemia) | 100% (Severe anemia),83.3% (Mild anemia) | r=0.808, p<0.01 | N/A | 12g/dL (Mild anemia), 8g/dL (Severe anemia) | Patient and healthy donors (22 subjects) | [233] |
| Retinal fundus photos | 96.90% | 91.70% | p=0.019 | Image Structure Clustering (Not Specific) | N/A | N/A | [240] |
| Finger probe with five PPG signals | N/A | N/A | R2 = 0.964 (five wavelengths PPG system), R2 = 0.688 (three wavelengths PPG system) | N/A | N/A | N/A | [241] |
| Visual and color-based point-of-care anemia self-testing | 90.20% | 83.70% | r = 0.864 | Hematology analyzer (Advia 2120i) | 11 g/dL | Patients with various anemia etiologies (238 subjects) | [242] |
| Microfluidic approach | N/A | N/A | N/A | Complete blood count (Not Specific) | 8.0 g/dL (severe anemia) | N/A | [140] |
| Novel multi wavelength sensor | N/A | N/A | R2=0.99 (Cold hands), R2=0.82 (previous heat) | HemoCue® and a blood gas analyzer (Radiometer ABL 80) | 13.5 g/dL (male), 12.5 g/dL (female) | Blood donor (1008 subjects) | [243] |
| HemaApp with smartphone cameras | 85.70% | 76.50% | r = 0.69 (White+970nm LEDS), r = 0.74 (White+ 970nm LEDS+Incandescent), r = 0.82 (White+880nm or 970nm LEDS+Incandescent) | Complete blood count (Not Specific) | N/A | Patients ranging from 6 – 77 years of age (31 subjects) | [213] |
| Near-infrared spectrophotometric system with InGaAs detector array and plane grating spectrometer | N/A | N/A | (r=0.74, r=0.81 in predicting test 1), (r=0.60, r=0.81 in predicting test 2) | N/A | N/A | Blood donor (91 subjects) | [164] |
| Imaging cytometry platform | N/A | N/A | r=0.98 | Hematology analyzer (Sysmex KN21) | N/A | Blood sample from UCLA Blood and Platelet Center (No subject report) | [142] |
Table 5.
Time and cost analysis of commercially available anemia detection technologies
| Technology | Test Time | Cost |
|---|---|---|
| WHO Haemoglobin Colour Scale | 30 seconds [65] | $0.01/test |
| HemoCue® 201 | 60 seconds [244] | Device: $698 Consumable: $1.5/test |
| Diaspect ™ | within 2 seconds [90] | Device: $225 Consumable: |
| HemoControl | 25 seconds [245] | Device: $425 Consumable: $1.14/test |
| URIT-12 | 10 to 15 seconds [246] | Device: $159 Consumable: $0.7/test |
| Anemo Check | 60 to 120 seconds [234] | $1/test |
| Pronto-7® | 30 to 45 seconds [101] | $629 |
3. Hemoglobin Level Measurement Methods
3.1. Minimally invasive Methods
Various methodologies have been previously developed for Hb level measurement and anemia detection. Minimally invasive anemia detection methods require blood sampling using minimally invasive techniques such as finger pricks. The cyanmethemoglobin, or hemiglobincyanide (HiCN) method, is currently used as a reference method for evaluating new instruments and alternative methods for Hb level measurement in a standardized guideline provided by Clinical and Laboratory Standards Institute titled CLSI H15-A3 [40] and International Committee for Standardization in Haematology (ICSH) [41]. Vanzetti’s azide methemoglobin method demonstrates comparable sensitivity and specificity for Hb level measurement to HiCN method and provided the foundation of the first generation of POC hemoglobin devices [42]. Other minimally invasive methods including Tallquist method [43], Sahli’s method [44], Alkaline Hematin Detergent method [45], and Copper-Sulfate Specific Gravity method [46]. These methods, although mostly being inexpensive, simple and rapid, suffer from one or multiple deficiencies including significant measurement errors [42, 47], lower sensitivity and reliability [48–51], relying on subjective visual judgement [42], lacking quality control, or requiring biohazardous chemical disposal [46, 52, 53]. Currently, these methods are mostly used in low- and middle-resource settings, where expensive and complex methods are impractical due to limited resources [42].
3.2. Non-invasive Methods
Compared to minimally invasive methods, Hb level measurement based on non-invasive techniques is advantageous for obvious reasons of eliminating the need for a minimally invasive finger prick or venous blood draw. To date, the majority of commercialized non-invasive Hb level measurement devices rely on spectrophotometry or photoplethysmography (PPG) to identify the spectral pattern of Hb in underlying blood vessels to determine Hb level. The working principle of both methods is based on the Beer-Lambert law, I0 = Ie−α·C·D, where I0 is output light intensity, I is incident light intensity, α is the light absorption coefficient, C is the concentration of Hb, and D is the light path.
In spectrophotometry, near infrared (NIR) light transmit through or reflect from tissue and blood differentially depending on the properties of a living tissue [54]. The variation of these light transmission or reflection properties are associated with the shape, volume, refractive index of Hb, and angular distribution of scattered light, which captures the absorption properties of blood and tissue and enables non-invasive Hb measurement [55]. Light absorption (or light attenuation) is correlated to oxygenated Hb and deoxygenated Hb by modified Beer-Lambert Law, , where Aλ is the light absorption (in optical densities) at specific light wavelength, λ, DPF is the Differential Pathlength Factor, and G is a parameter of scattering [56]. One specialized spectrophotometer, CO-oximeter, has been applied by commercially available technologies for measuring total hemoglobin concentration based on the fact that hemoglobin and all of its derivatives including oxygen-carrying hemoglobin, non-oxygen-carrying normal hemoglobin, carboxyhemoglobin and methemoglobin are colored proteins which absorb light at specific wavelengths and thus have a characteristic spectrum [57, 58].
Photoplethysmography (PPG) signal consists of alternating current (AC) and direct current (DC) components [59]. The AC component represents the light absorption by changes of pulsatile component of artery blood, while the DC component represents the light absorption by tissues, non-pulsatile component of artery blood and continuous venous blood [60]. The absorption by tissue (T), plasma (P), and Hb under certain wavelength λ can be then expressed as: ; and the AC and DC components are expressed as: , and , where D1 is the path length for Hb and plasma in AC signal (pulsatile component of artery blood), D2 is the path length for Hb and plasma in DC signal, and DT is the path length for tissue. The ratio between the AC and DC components, , where ΔD = D1 – D2 is the path length that affects only the Hb and plasma under wavelength of λ, is thus calibrate toward the effect of tissue and can be used to determine Hb level [59, 61]. The readers may refer to published reviews for a comprehensive understanding on non-invasive methods for Hb measurement [62].
4. Commercially Available Point-of-Care Testing Technologies for Hb Level Measurement
Performance of commercially available technologies including WHO Haemoglobin Colour Scale, Hemocue®, NBM-200, and Pronto-7® have been evaluated by multiple groups, while fewer publications are available for evaluation of other technologies. Performance of emerging technologies have been mainly reported by the inventors in the publications where the technology was reported. The majority of identified publications have reported mean bias (or mean difference) and standard deviation (or 95% confidence interval, or standard error) when comparing the Hb level measured using the reported/tested technology to the Hb level measured by various reference methods, as well as sensitivity and specificity for anemia detection. Fewer publications included the correlation of Hb level measured by reported/tested technology and reference methods (i.e., R2, R, or Pearson correlation coefficient), and even fewer included the receiver operating characteristic (ROC) curves. Furthermore, large discrepancies exist among various publications reporting the performance of the same technology. For example, one report published by the WHO found that the WHO Haemoglobin Colour Scale to have 95.0% sensitivity and 99.6% specificity in determining the degree of anemia [63]. However, another study published by Lindblade et al. reported 24.0% sensitivity and 97.0% specificity [64], while one other study published by Madakshira et al. reported 99.3% sensitivity and 27.7% specificity [65]. These discrepancies exist in the published studies for most methods, probably due to the variations among test population (adults vs. children, male vs. female, females with and without pregnancy, healthy donors vs. patients with various of diseases), reference method (Cyanmethemoglobin method and a range of different models of AHAs), applied threshold hemoglobin level for defining anemia, as well as testers (researchers, clinicians, nurses, and laboratory technicians). In this review, we compare the performance of the most extensively tested technologies by summarizing the minimum and maximum values of reported sensitivity, specificity, as well as mean bias (mean difference) and standard deviation, along with the test information including test population and the applied reference method (Table 3). Reported performance for other technologies will be individually discussed, due to the limited data available.
4.1. Commercially Available Minimally Invasive Technologies
The World Health Organization (WHO) hemoglobin color scale (HCS) (Fig. 1a) is a quantitative method developed as a commercial kit by WHO under license agreement by COPAC (Oststeinbek, Germany). HCS determines Hb level by comparing blood induced color change on a chromatography paper with the standard card containing 6 standard colors which correlates to Hb levels from 2 to 14 g/dL displayed in increments of 2 g/dL [66–68]. The performance of HCS has been evaluated in numbers of publications as summarized in Table 3 [51, 63–65, 69, 70]. Overall, WHO color scale is an inaccurate method to screen anemia [42, 71].
Figure 1. Minimally invasive point-of-care technologies for Hb level measurement and anemia testing.

(A) The WHO Haemoglobin Colour Scale (Reproduced with permission from [230]). (B) HemoCue® Hb 301(Reproduced with permission from Hemocue, Inc, HemoCue America), (C) DiaSpect™ (Reproduced with permission from EKF Diagnostics, Inc), (D) Mission® Hemoglobin Plus Hb (Reproduced with permission from ACON Laboratories, Inc), and (E) AnemoCheck™ (Reproduced from with permission from [98]).
HemoCue developed various models of hemoglobin photometers (HemoCue® Hb 201 DM System, HemoCue® Hb 201+ System, HemoCue® Hb 301 System (Fig. 1A)) for POC Hb measurement. HemoCue® System consists of a photometer to determine hemoglobin level by converting hemoglobin to azidemethemoglobin at 565 nm with compensation for turbidity at 880 nm [72, 73]. HemoCue® is by far one of the most extensively evaluated technology for anemia detection [74–83]. According to these studies, HemoCue demonstrated overall satisfactory performance for detecting anemia, while still hold variability among operators [84] and groups of tests [83] and test environment such as humidity [75]. Additional articles are available that review literature on the performance of HemoCue® as compared with laboratory reference tests [85].
Other commercially available minimally invasive technologies for Hb level measurement including Diaspect™, HemoControl, Mission® Plus Hb, URIT Hemoglobin meter, and AnemoCheck™ have also been evaluated, although fewer test data are available than HCS and HemoCue®. The DiaSpect™ system and HemoControl (Fig. 1C) by EKF Diagnostics utilizes multichromatic sensor which measures the whole blood absorbance report Hb level within 5 seconds [86–88]. Both of the technologies have been tested by a few studies as summarized in Table 3 [86, 89–92]. Mission® Plus Hb (Fig. 1D) developed by ACON Laboratories uses the principle of reflectance photometry. Hb is converted to methemoglobin on a disposable test strip and measures Hb level measurement, and Hematocrit [93]. We only identified two studies for evaluating the performance of this technology [93, 94]. AnemoCheck™ (Fig. 1E) developed by Sanguina, LLC uses a reagent solution containing hydrogen peroxide and 3,3’,5,5’-tetramethylbenzidine that react with 10 μL of blood sample [95]. The blood sample and reagent reacts within a small vial, and the Hb level is measured by comparing the reactant color change to a standard color scale [95–97]. A modified AnemoCheck-LRS can be used in limited-resource settings to permit testing and diagnosis of severe anemia (roughly 2 – 8 g/dL) which can be common in this setting [95, 97, 98]. Further tests would be necessary to better evaluate these technologies under various conditions on different test populations. Overall, minimally-invasive anemia detection technologies demonstrate reasonable reproducibility and accuracy for Hb level measurement as well as sensitivity and specificity for anemia detection. However, blood sampling may cause discomfort to the patient and carry an infection risk.
4.2. Commercially Available Non-invasive Technologies
Pronto-7® developed by Masimo (Fig. 2A) allows intermittently records and displays spectrophotometry measurements [99, 100]. There are recent studies that have been conducted to evaluate the performance of Pronto-7® [81, 99, 101–110] as summarized in Table 3. Multiple publications have suggested to take general caution on making clinical decisions based on these measurements alone [99, 100, 106, 108, 111, 112].
Figure 2. Non-invasive point-of-care Hb level measurement technologies.

(A) Pronto-7® (Reproduced from online resource [231], Masimo, Inc), (B) Haemospect® (Reproduced from online resource [232]), and (C) NBM-200 (Reproduced with permission from OrSense U.S., LLC)
Haemospect® (Fig. 2B) developed by MBR Optical Systems measures Hb level leveraging transcutaneous reflection spectroscopy [58]. The device projects white light into underlying tissue and the transmitted and reflected light are broken down to its separate wavelengths by a spectrometer to calculate the subject’s Hb level [112]. Compared to Pronto-7® and NBM-200, fewer evaluation studies have been conducted on Haemospect® as summarized in Table 3 [58, 79, 111–117]. The observed sensitivity ranges from 3.5% [79] to 94.6% [115]. Additional studies on the performance of Haemospect® for Hb level measurement and anemia detection may further verify the validity of the system. Additionally, Bosch has recently developed a portable Hemoglobin Monitor Solution in 2021. We have not identified any clinical evaluation studies on this new technology as of this review.
NBM-200 developed by OrSense (Fig. 2C) utilizes a reusable, ring-shaped pneumatic sensor probe to perform optical absorption measurements of temporary blood-flow occlusion on a subject’s finger. Hb level is calculated using a portable desktop monitoring system from a multi-wavelength light source generated by the occluded blood within 60 seconds [88, 111]. This method is similar to normal spectroscopy, except it has a high signal-to-noise ratio due to the obstruction of blood flow, which improves sensitivity of Hb concentration and pulse rate measurements [118–120], while it is intended to spot check Hb level instead of continuous monitoring. The performance of NBM-200 has been evaluated by multiple studies [77, 79, 89, 90, 105, 107, 110, 120–124] as summarized in Table 3. Similar to Pronto®–7, it is also suggested that general caution should be taken if this technology is used to prescreen blood donors for eligibility [77].
5. Emerging Technologies for Hemoglobin Level Measurement and Anemia Detection
In this section, we introduce the emerging approaches and technologies for anemia detection. We categorize the emerging technologies into minimally invasive, including electrochemical impedance and optical based methods, as well as non-invasive methods, including near-infrared, photoplethysmography, and smartphone-based technologies. We attempt to summarize the key perspectives or components for each technology that the researchers have focused on to improve for each category. We also refer the readers to previously published review articles for further information under each sub-category.
5.1. Emerging Minimally Invasive Anemia Detection Technologies
5.1.1. Emerging minimally invasive Anemia Detection Using Electrical and Electrochemical Sensors
Electrical and electrochemical sensors enable the estimation of hematocrit and Hb level by measuring key electrical parameters including impedance, capacitance and electric current of blood constituents. Electrochemical sensors are normally comprised of a working electrode, counter electrode, and with a third reference electrode [125], where WE measures electrical response of the applied blood sample, the CE supplies the required current, and the RE tracks the electrical signal.
StatStrip Hb/Hct developed by Nova Biomedical leverages the electrochemical method for measuring Hb level and hematocrit (Hct) using 1.6 μL of blood in 40 seconds [126]. As of the time of this review, this technology is not yet available in the USA or Canada thus we included it under ‘emerging technologies’. Chakraborty et al have developed a silicon based microfluidic device employing a two electrodes system for measuring Hct levels (Fig. 3A) [127]. Punter-Villagrasa et al. developed a three electrodes system for measuring Hct levels based on the electric impedance (Fig. 3B) [128, 129]. Phillips et al. demonstrated a non-invasive platform to measure Hb level using a customized capacitance sensor (Fig. 3C) [130]. Overall, these technologies are still at relatively early stages and have only been tested using either small sample sizes or Hb solutions instead of actual clinical blood samples.
Figure 3. Electrical and electrochemical based Hb level measurement methods.

(A) Silicon based two electrodes microfluidic system for Hb level measurement (Reproduced with permission from [127]). (B) Three-electrodes system for electrochemical impedance spectroscopy (EIS) based Hb level measurement (Reproduced with permission from [129]). (C) Customized capacitance sensor for measuring Hb levels non-invasively (Reproduced with permission from [130]). (D) Measurement of red blood cell count and mean corpuscular volume using electrochemical collision events (Reproduced with permission from [131]).
Ho et al. developed a unique system employing electrochemical detection of both Hb level and RBC mean corpuscular volume (MCV) [131]. The working principle of this technology relies on the detection of decrease in staircase current caused by collision of RBC to electrode surface, which hinders electrochemical oxidation of an electroactive redox species, potassium ferricyanide (Fig. 3D). This technology uniquely enables measurement of RBC MCV in addition to Hb concentration, which can potentially allow identification of anemia-related diseases caused by abnormal RBC size such as megaloblastic anemia and microcytic anemia.
Electrical and electrochemical based technologies have the potential to provide anemia detection platforms that are portable, label free, easy to use, high throughput, and without complex optical instrumentation [132]. However, some of the current electrochemical platforms still cannot achieve high reproducibility, accuracy and stability as published by the CLSI guidelines (EP05-A3 [133] and EP25-A [134]) [135]. Additionally, variations in other blood components such as white blood cells (WBCs), platelets, plasma proteins and plasma ions may affect the accuracy of the reading thus additional blood pre-processing is required. Readers are referred to recently published review articles for additional information regarding electrochemical sensors [132, 136, 137].
5.1.2. Emerging Minimally Invasive Anemia Detection Using Optical Sensors
Optical-sensing based anemia detection relies on the sensitive detection of photon emission from blood cells, hemoglobin or reactant from hemoglobin such as methemoglobin. Researchers have employed microfluidic technologies to improve anemia detection with various perspectives including blood sample processing, mixing, detection wavelength selection, whole blood processing, and assay multiplexing. Kim et al. developed a disposable polymethyl methacrylate microfluidic cuvette with a single sidewall microchannel at customized channel depth and a dry film resist as adhesive to efficiently blood sampling by capillary action [138]. Plevniak et al. developed microfluidic auto-mixer, iPOC3D, using 3D printing. The developed iPOC3D automatically mix whole blood with reagents including 3,3’,5,5’-tetramethylbenzidine and hydrogen peroxide (H2O2) via capillary force, which is then captured and analyzed calorimetrically using smartphone camera for Hb levels measurement (Fig. 4A) [139]. These sample processing techniques can simplify sample preparation, eliminate the requirement of laboratory infrastructure and external off-chip operation, and increase detection accuracy by preventing air trapping and cell aggregation.
Figure 4.

Emerging optical sensing based minimally invasive Hb level measurement methods. (A) iPOC3D automatic blood mixer (Reproduced with permission from [233]). (B) Microfluidic cartridges for blood count (Reproduced with permission from [142]). (C) Paper-based thermometer-style channel for determining hematocrit (Reproduced with permission from [144]). (D) Gazelle Hb Variant/Anemia paper-based microchip electrophoresis for integrated anemia detection and hemoglobin variant identification (Hemex Health Inc, Portland, Oregon) [145–147].
While Hb levels in lysed blood samples can be measured relatively easily solely based on light absorption, anemia detection using unprocessed whole blood can further simplify sample preparation steps and more importantly, can potentially allow additional blood tests for determining the causes of detected anemia with the same sample [140]. Taparia et al. and Halder et al. developed microfluidic systems for anemia detection using non-lysed whole blood samples instead of lysed hemoglobin [140, 141]. The membrane of intact RBCs in non-lysed blood causes light scattering which create deviation from the linear relationship between light absorption and hemoglobin concentration. To maintain the accuracy of Hb measurement, Taparia et al. incorporated parameters including absorption and scattering coefficient and fractional hematocrit in a more generalized model and enabled hemoglobin level measurement based on the calibrated non-linear fitting curve [140]. Halder et al. associated hemoglobin level with the difference between the absorption at 570 nm (isosbestic point of oxy- and deoxy-hemoglobin) and 630 nm (deoxy-hemoglobin) [141].
Zhu et al. developed three types of microfluidic cartridges that can be individually adapted to the same manifold which can incorporate smartphone for imaging the cartridges (Fig. 4B) [142, 143]. Each microchip enables one type of measurement including: (1) WBC counting based on fluorescent labeling, (2) RBC counting using pre-diluted RBCs, and (3) Hb level measurement using lysed blood. The capability of performing WBC and RBC counting in addition to Hb level measurement is highly significant. Future efforts on eliminating sample pre-processing steps or integrating with whole blood processing microchips can further increase the platform usability, especially in middle- or low-resource settings [142, 143].
Paper-based microfluidics have also been implemented for anemia detection. Berry et al. developed a paper-based microfluidic device with selected paper properties and customized device treatment and geometry to enable controlled transport of RBCs. The traveled distance of RBCs is then used to determine hematocrit of whole blood based on the in a thermometer-styled channel (Fig. 4C) [144]. Our group has developed a paper-based microchip electrophoresis platform that allows hemoglobin variant identification [145] and a two-step hemoglobin separation [146, 147]. The first step separates total Hb and a standard calibrator based on the major difference in electrophoretic mobility. The Hb level is determined by measuring the relative abundancy of Hb band and standard calibrator band to allow anemia detection. The second step further separates Hb variants due to their finer mobility differences to allow identification of abnormal Hb variants, including sickle hemoglobin, hemoglobin C, and hemoglobin E (Fig. 4D) [145–152].
5.2. Emerging Non-Invasive Anemia Detection
In certain cases, utilization of minimally invasive methods for anemia detection may be challenging, for example for infants, elderly, pregnant women, as well as for patients who may need continuous Hb level monitoring [61, 153]. Frequent blood sampling can create discomfort to the patient and can be relatively expensive, which is specifically challenging in low- and middle-resource settings. As a result, there is great interest of developing methods and tools that enables Hb level measurement non-invasively, with reduced costs, both in the lab and at the point-of-care [61, 153].
In clinical the environment, rapid non-invasive screen for anemia at the primary care level are conducted by clinicians by examining palmar, nailbed, and conjunctivae for pallor [154, 155]. Although these initial subjective screenings can be useful, especially for identifying severe anemia [154, 156], this method overall demonstrates low sensitivity and specificity for anemia detection and suffers from poor inter-observer agreement [154, 157–159]. The commercially available non-invasive technologies for Hb measurement (Section 3.2) are mainly based on photospectroscopy and photoplethysmography analysis on these body sites [56, 102, 106, 108, 160–164]. However, most of these technologies suffer from one or more of the following limitations: challenging data collection methods; complex data analysis and feature extraction processes; affordability and portability; and relatively low accuracy [165].
5.2.1. Near Infrared (NIR) Spectroscopy Based Technologies
NIR-based Hb level measurement can be traced back to early 2000 [55, 166–169]. However, most of those systems suffered from low detection accuracy. Ongoing research efforts in NIR spectroscopy-based anemia detection have focused on addressing the insufficient signal-to-noise ratio (SNR) of measuring instruments and influences of individual differences remain a challenge [170]. Emerging approaches and technologies have been developed to improve measurement accuracy from the perspectives of light sources (i.e., LED, tungsten halogen light), focus lens, optical collector, beam-splitting system, data collection system, and data analysis algorithm [160, 164, 171–176]. Tian et al. used Equidistant Combination Multiple Linear Regression method and Partner wavelength method to select wavelengths for Hb level measurement in whole blood and concluded that using only two wavelengths combination (1143 nm and 1298 nm) as performed as fine as full-spectrum model [177]. Yuan et al. implemented indium gallium arsenide array detector to avoid the influence from blood flowing, as well as incorporated multiple independent amplifier circuit to obtain higher SNR and sampling rate [164]. Ding et al. have used nine-wavelength broadband light source and silicon photodiode array and developed a high-performance spectrophotometry detectors, with implemented back propagation artificial neural network to analyze the principal components of the spectra extracted [160]. In a most recently published work, Askari et al. implemented novel periodic metamaterial structure that perfectly absorbs the electromagnetic wave in a narrow frequency band in NIR region for Hb level measurement [173]. Readers should refer to recently published review articles for additional information on NIR based sensors [174–176].
5.2.2. Photoplethysmography (PPG) Based Technologies
Numerous commercially available and emerging PPG-based technologies have been developed during the past decade and have demonstrated significant progress. However, the non-invasive Hb level measurement demonstrated inconsistent or compromised accuracy during various tests as summarized by us and others. Ongoing research efforts in PPG-based anemia detection have focused on selection of light sources and wavelengths (i.e., LED, lasers, or lights of different spectra [178–181]), reducing the power consumption, design of photodetectors, and data processing methods. Timm et al. implemented three wavelengths (670, 810, and 1300 nm) to measure Hb level, pulse, and oxygenation, and a fourth wavelength at 1150 nm to measure water absorption to facilitate the calculation of Hb level [182]. Doshi et al. and Nirupa et al. independently developed non-contact sensor using combinations of red and infrared light wavelengths at 660, 940 nm [183] and 624, 850 nm [184] for Hb level measurement. Liu et al. implemented eight wavelength LEDs with broadband light source combined with temporally constrained independent component analysis and adaptive filters to derive high-quality PPG signals [185]. To reduce the device power consumption, Glaros et al. developed a fully integrated pulse oximeter front-end with sub-mW power consumption [186]. Various of photodetectors were developed to improve signal detection. Schidl et al. implemented standard CMOS process to build multi-junction device [187]. Suzaki et al. implemented silicon photodiode combined with indium gallium arsenide [188]. Costa et al. implemented vertically stacked double junction photodiodes [189]. Yuan et al. developed a multi-sensor approach combining the PPG probe, an accelerometer, and a wearable chest respiration monitor to minimize signal noise including motion artefact, blood pressure and respiration variations [190].
Recent advances in digital image processing and machine learning methods can potentially offer solutions to improve these non-invasive anemia screening methods, as summarized here. For example, Golap et al. utilized symbolic regression of multigene genetic programming to facilitate data analysis [191]. Liu et al. developed models based on partial least squares regression and backpropagation artificial neural network to process PPG signals [185]. Kavsaoğlu used eight regression machine learning algorithms to recognize the characteristic features of PPG signals [192]. Acharya et al. implemented Stacked Regressor Model machine learning model comprising of a two-layer stack of regressors including Least Absolute Shrinkage and Selection Operator, Ridge, Elastic Net, Adaptive Boost and Support Vector Regressors [193]. The readers can refer to recently published reviews for ongoing research efforts on PPG-based Hb measurement and other applications [176, 194–196].
5.2.3. Smartphone Based Technologies
With recent advances in high performance photographic sensors in smartphones, smartphone-based tests have demonstrated significant advantages of being easily accessible and making anemia screening low cost [61, 153]. However, the variety of photographic sensors installed in smartphones poses challenges in the standardization of methods and in the development of universal tools for image acquisition and data analysis. The majority of smartphone-based technologies analyze tissue colors using captured images from conjunctiva [197–209], fingertip [210–215], nailbed [216], or retinal fundus [217]. Collings et al. determined erythema index calculated from smartphone images of palpebral conjunctiva taken under ambient light conditions can be used to measure Hb levels (Fig. 5A) [203]. Dimauro et al. developed an accessory that consists of a special spacer and a macro lens to acquire images using smartphone camera without impact from ambient light (Fig. 5B) [201]. Park et al. leveraged spectral super-resolution spectroscopy to mathematically reconstruct high-resolution spectra of blood Hb acquired by the built-in camera of a smartphone (Fig. 5C) [218]. Wang et al. developed HemaApp involved white and IR LED for anemia detecting over fingertip [213], which has been improved to require only white LED through hardware configuration [212]. Mannino et al. developed machine learning based smartphone app to measure Hb levels from images of the fingernail bed (Fig. 5D) [216]. Mitani et al. developed deep learning algorithm to measure Hb levels using images of retinal fundus, which can be captured in teleretinal disease screening (Fig. 5E) [217]. This method can be valuable for studying the association of anemia and ocular disease, for which fundus images are already available [217].
Figure 5.

Emerging smartphone based non-invasive anemia detection. (A) Erythema index measurement for determining Hb level from the palpebral and forniceal portions of the conjunctiva, with in-frame color calibration target (Reproduced with permission from [203]). (B) Accessory to eliminate impact from ambient light during non-invasive Hb level test (Reproduced with permission from [201]). (C) Spectral super-resolution spectroscopy (SSR) based Hb level measurement using inner eyelid as accessible sensing site. The smartphone application collects red, green and blue color information and applies SSR to construct spectra used for determining Hb levels (Reproduced with permission from [218]). (D) Fingernail bed based Hb level measurement (Reproduced with permission from [216]). (D) Retinal fundus based Hb level measurement using red and white areas in each saliency map. This technology has the potentially to determine the association of anemia and ocular disease (Reproduced with permission from [217]).
In a most recently published review article, Hasan et al. recommended to use fingertip as data collection site due to easy access, use of three different NIR lighting sources, specific signal processing techniques and feature selection methods, and region of interest selection methods, for the optimal development of an accurate Hb measurement model [61]. In addition to these, Dimauro et al. also recommended to increase data sharing and data availability [153]. The readers may refer to these articles for further details regarding smartphone-based technologies [61, 153, 176].
6. Discussion and Future Perspectives
The WHO has declared anemia as a public health priority and has listed anemia testing in the essential list of in vitro diagnostics [29]. Recently developed novel micro-technologies offer simple, rapid, and affordable screening and diagnosis of anemia in minimally invasive and non-invasive manner. These technologies have the potential to facilitate universal diagnosis and screening in low-resource settings. Smart and cautious utilization and wide-spread application of these technologies may revolutionize anemia diagnosis and screening in resource-limited settings and dramatically decrease early mortality due to anemia.
Minimally invasive technologies overall demonstrate relatively higher reproducibility and accuracy in Hb level measurement and have demonstrated the potential to be integrated with additional tests for identifying anemia etiology. Current and future improvement considerations on minimally invasive anemia detection technologies include whole blood processing, reagent mixing, selection of detection wavelength, and assay multiplexing. Integration with blood pre-processing modules or microchips may facilitate efficient sample handling for Hb level measurement as demonstrated by Berkel et al. [219]. Review articles are available on microfluidic-based plasma separation [220] and blood cell sorting [221] technologies that could potentially be available for integration. Potential challenges when integrating these technologies may arise from device cost, usability, portability and the relatively strict requirements for operating environment [219].
Non-invasive technologies have obvious advantage of avoiding discomfort associated with blood sampling, eliminating biohazardous specimen handling, and enabling real-time Hb level monitoring. There have been more than one hundred articles published in recent years reporting various technologies for non-invasive anemia detection. Current research interests are oriented towards the combination of smartphone-based data collection and machine learning algorithm-based image analyses. Current and future improvement considerations on non-invasive anemia detection technologies include selection of body sites for signal acquisition, the type of light source, image sensors, additional accessories, data analysis algorithms, and data availability [61, 153, 176].
Reproducible and accurate Hb level measurement is necessary for clinicians to establish efficient treatment plana for individual patients. To address the need for convenience and accuracy, we envision that one option for future POC anemia detection technologies is the integrated use of non-invasive and minimally invasive tests (Fig. 6). Such an approach would allow rapid and easy to use non-invasive screening with high sensitivity, followed by minimally invasive diagnosis with high specificity to provide accurate confirmation and anemia severity determination to facilitate accurate clinical decisions.
Figure 6.

There is a need for multiplexed approaches and integrated disease etiology identification in point-of-care anemia testing. Future efforts should focus on cascading non-invasive and minimally invasive diagnostic methods to enhance widespread use and diagnostic accuracy, as well as multiplexing additional test capabilities to identify the root-cause of anemia, such as integrated tests for red cell disorders, iron deficiency, and malaria.
Upon detection, timely and efficient treatment of anemia require identification of specific disease etiology [6, 222] thus integrating tests for the root cause would be helpful (Fig. 6). For example, policy in Kenya states that women should be tested for multiple diseases at first antenatal care including anemia, malaria, HIV and syphilis [223]. Iron deficiency, hemoglobinopathy (such as SCD), and malaria are the three major causes for anemia and are typically most prevalent in the same areas [222]. These diseases are also listed by the WHO as essential items for in vitro diagnostics and are suggested to be assessed with anemia testing [29]. For example, the WHO recommends that iron supplements should be used only where malaria prevention and control systems are available [20]. Also, the United States Agency for International Development recognized the need for pairing of malaria testing with complementary anemia testing [224, 225]. As a result, there is a significant need for affordable, easy-to-use POC platforms that can perform integrated detection of anemia, iron deficiency, malaria, and sickle cell disease. Current detection of these diseases, however, would require involvement of multiple single-purpose technologies. For example, Smart et al. in 2018 conducted a study for screening anemia and sickle cell disease in Tanzania using two POC technologies [226], while Young et al. in 2018 conducted a study for screening anemia and malaria in western Kenya implementing two separate POC methods [227]. Recent collaborative efforts from the U.S. National Institute of Biomedical Imaging and Bioengineering of National Institute of Health, and Bill & Melinda Gates Foundation have emphasized the need for developing non-invasive POC technologies for integrated testing of anemia, sickle cell disease, and malaria. Such integrated platforms can be used to rapidly screen large populations as well as provide physicians with a practical tool for optimizing therapy in individual patients. In a few of the emerging technologies reviewed in this article have already demonstrated the potential to perform integrated test for anemia and anemia etiology as well as other diseases such as megaloblastic anemia and WBC count [142, 143], microcytic anemia [131], ocular disease [217], and SCD [146, 147], and iron deficiency [33]. Comprehensive review articles on POC detection of iron deficiency [31–33], malaria [34–36], and sickle cell disease [37–39] are available for investigators who are interested in integrating these test capabilities to provide a diagnostic platform for anemia surveillance.
Finally, we would like to emphasize the importance of testing and publishing on new technologies following standard guidelines. We have noticed that the studies conducted for evaluating the performance of these technologies lack consistency in data reporting. Conducting and reporting diagnostic accuracy studies following published guidelines would enable researchers or clinicians to more easily compare and contrast different technologies. Guidelines for clinical study design, study participants, sample size calculations, details on test methods, as well as data reporting are available, such as the Reference and Selected Procedures for the Quantitative Determination of Hemoglobin in Blood [40], Evaluation of Precision of Quantitative Measurement Procedures [133], Evaluation of Stability of In Vitro Diagnostic Reagents [134] published by National Committee for Clinical Laboratory Standards, Standards for Reporting Diagnostic Accuracy [228], and Guidance Document on Management of Point-of-Care Testing published by the American Association for Clinical Chemistry [229].
Significance.
In this critical review, emerging non-invasive and minimally invasive point-of-care anemia detection technologies are discussed, highlighting the future directions and the need for multiplexed approaches and integrated disease etiology identification.
Acknowledgments
Authors acknowledge National Heart Lung and Blood Institute Small Business Innovation Research Program (R44HL140739, R41HL151015), National Heart Lung and Blood and the National Center for Complementary & Integrative Health (NCCIH) 1U54HL143541, National Institute of Diabetes and Digestive and Kidney Diseases Small Business Innovation Research Program (R41DK119048), and NHLBI R01HL133574 and T32HL134622. This article’s contents are solely the responsibility of the authors and do not necessarily represent the official views of the National Institutes of Health.
Footnotes
Conflicts of interest
RA, CP, JAL, UAG, and Case Western Reserve University have financial interests in Hemex Health Inc. JAL, EK, UAG, and Case Western Reserve University have financial interests in BioChip Labs Inc. UAG and Case Western Reserve University have financial interests in Xatek Inc. UAG has financial interests in DxNow Inc. Financial interests include licensed intellectual property, stock ownership, research funding, employment, and consulting. Hemex Health Inc. offers point-of-care diagnostics for hemoglobin disorders, anemia, and malaria. BioChip Labs Inc. offers commercial clinical microfluidic biomarker assays for inherited or acquired blood disorders. Xatek Inc. offers point-of-care global assays to evaluate the hemostatic process. DxNow Inc. offers microfluidic and bio-imaging technologies for in vitro fertilization, forensics, and diagnostics. Competing interests of Case Western Reserve University employees are overseen and managed by the Conflict of Interests Committee according to a Conflict-of-Interest Management Plan.
References
- 1.(WHO), T.W.H.O., Haemoglobin concentrations for the diagnosis of anaemia and assessment of severity. WHO reference number: WHO/NMH/NHD/MNM/11.1, 2011. [Google Scholar]
- 2.Thom CS, et al. , Hemoglobin Variants: Biochemical Properties and Clinical Correlates. Cold Spring Harb Perspect Med, 2013. 3(a011858): p. 1–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Stevens GA, et al. , Global, regional, and national trends in haemoglobin concentration and prevalence of total and severe anaemia in children and pregnant and non-pregnant women for 1995–2011: a systematic analysis of population-representative data. The Lancet Global Health, 2013. 1(1): p. e16–e25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Beghe C, Wilson A, and Ershler WB, Prevalence and outcomes of anemia in geriatrics: a systematic review of the literature. Am J Med, 2004. 116 Suppl 7A: p. 3s–10s. [DOI] [PubMed] [Google Scholar]
- 5.Calabrich A and Katz A, Management of anemia in cancer patients. Future Oncol, 2011. 7(4): p. 507–17. [DOI] [PubMed] [Google Scholar]
- 6.Kassebaum NJ and Collaborators, The Global Burden of Anemia. Hematol Oncol Clin N Am, 2016. 30: p. 247–308. [DOI] [PubMed] [Google Scholar]
- 7.McLean E, et al. , Worldwide prevalence of anaemia, WHO Vitamin and Mineral Nutrition Information System, 1993–2005. Public Health Nutrition, 2008. 12(4): p. 444–454. [DOI] [PubMed] [Google Scholar]
- 8.Aspuru K, et al. , Optimal management of iron deficiency anemia due to poor dietary intake. International Journal of General Medicine, 2011. 4: p. 741–50. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Koukounari A, et al. , Relationships between anaemia and parasitic infections in Kenyan schoolchildren: a Bayesian hierarchical modelling approach. Int J Parasitol, 2008. 38(14): p. 1663–71. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Bechir M, et al. , Parasitic infections, anemia and malnutrition among rural settled and mobile pastoralist mothers and their children in Chad. Ecohealth, 2012. 9(2): p. 122–31. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Kassebaum NJ, et al. , A systematic analysis of global anemia burden from 1990 to 2010. Blood, 2014. 123(5): p. 615–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Camaschella C, Iron-Deficiency Anemia. New England Journal of Medicine, 2015. 372(19): p. 1832–1843. [DOI] [PubMed] [Google Scholar]
- 13.Bryer E and Henry D, Chemotherapy-induced anemia: etiology, pathophysiology, and implications for contemporary practice. International Journal of Clinical Transfusion Medicine, 2018. 6: p. 21–31. [Google Scholar]
- 14.White NJ, Anaemia and malaria. Malaria Journal, 2018. 17(1): p. 371. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Kato GJ, et al. , Sickle cell disease. Nature Reviews Disease Primers, 2018. 4: p. 18010. [DOI] [PubMed] [Google Scholar]
- 16.Chalco JP, et al. , Accuracy of clinical pallor in the diagnosis of anaemia in children: a meta-analysis. BMC Pediatr, 2005. 5: p. 46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Short MW and Domagalski JE, Iron deficiency anemia: evaluation and management. Am Fam Physician, 2013. 87(2): p. 98–104. [PubMed] [Google Scholar]
- 18.Malik SG, et al. , Non-nutritional anemia: Malaria, thalassemia, G6PD deficiency and tuberculosis in Indonesia. Asia Pac J Clin Nutr, 2020. 29(Suppl 1): p. S32–s40. [DOI] [PubMed] [Google Scholar]
- 19.Williams TN and Obaro SK, Sickle cell disease and malaria morbidity: a tale with two tails. Trends in Parasitology, 2011. 27(7): p. 315–320. [DOI] [PubMed] [Google Scholar]
- 20.Organization, W.H., Guideline daily iron supplementation in infants and children. 2016: World Health Organization. [PubMed] [Google Scholar]
- 21.Ballas SK, Iron overload is a determinant of morbidity and mortality in adult patients with sickle cell disease. Semin Hematol, 2001. 38(1 Suppl 1): p. 30–6. [DOI] [PubMed] [Google Scholar]
- 22.Aroke D, et al. , Iron supplementation in pregnant sicklers: an opinion. BMC pregnancy and childbirth, 2018. 18(1): p. 256–256. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Mariani R, et al. , Iron metabolism in thalassemia and sickle cell disease. Mediterranean journal of hematology and infectious diseases, 2009. 1(1): p. e2009006–e2009006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Modell B and Darlison M, Global epidemiology of haemoglobin disorders and derived service indicators. Bulletin of the World Health Organization, 2008. 86: p. 480–487. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Le CHH, The Prevalence of Anemia and ModerateSevere Anemia in the US Population (NHANES 2003–2012). PLOS ONE, 2016. 11: p. e0166635, 1–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Buttarello M, Laboratory diagnosis of anemia: are the old and new red cell parameters useful in classification and treatment, how? International Journal of Laboratory Hematology, 2016. 38(S1): p. 123–132. [DOI] [PubMed] [Google Scholar]
- 27.Wallerstein RO Jr., Laboratory evaluation of anemia. The Western journal of medicine, 1987. 146(4): p. 443–451. [PMC free article] [PubMed] [Google Scholar]
- 28.Mathison BA, Pritt BS, Update on Malaria Diagnostics and Test Utilization. Journal of Clinical Microbiology, 2017. 55(7): p. 2009–2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.(WHO), W.H.O., Second WHO Model List of Essential In Vitro Diagnostics. 2019(WHO/MVP/EMP/2019.05).
- 30.Organization, G.W.H., First WHO Model List of Essential In Vitro Diagnostics. 2019. ((WHO Technical Report Series, No. 1017; )). [Google Scholar]
- 31.Meybohm P, The future of iron deficiency diagnostics - Rapid home-use point-of-care test kits. EBioMedicine, 2019. 42: p. 28–29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Sultana GS, et al. , Value of red cell distribution width (RDW) and RBC indices in the detection of iron deficiency anemia. Mymensingh medical journal : MMJ, 2013. 22(2): p. 370–376. [PubMed] [Google Scholar]
- 33.Yap BK, et al. , Potential Point-of-Care Microfluidic Devices to Diagnose Iron Deficiency Anemia. Sensors (Basel, Switzerland), 2018. 18(8): p. 2625. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Agarwal R, et al. , Rapid diagnostic tests for Plasmodium vivax malaria in endemic countries. Cochrane Database Syst Rev, 2020. 11: p. Cd013218. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Kim S, et al. , Malaria rapid diagnostic test as point-of-care test: study protocol for evaluating the VIKIA Malaria Ag Pf/Pan. Malar J, 2015. 14: p. 114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.McBirney SE, et al. , Rapid Diagnostic for Point-of-Care Malaria Screening. ACS Sens, 2018. 3(7): p. 1264–1270. [DOI] [PubMed] [Google Scholar]
- 37.Alapan Y, et al. , Emerging point-of-care technologies for sickle cell disease screening and monitoring. Expert Rev Med Devices, 2016. 13(12): p. 1073–1093. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Ilyas S, Simonson AE, and Asghar W, Emerging point-of-care technologies for sickle cell disease diagnostics. Clin Chim Acta, 2020. 501: p. 85–91. [DOI] [PubMed] [Google Scholar]
- 39.Kaestner L and Bianchi P, Trends in the Development of Diagnostic Tools for Red Blood Cell-Related Diseases and Anemias. Front Physiol, 2020. 11: p. 387. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Standards, N.C.f.C.L., Reference and Selected Procedures for the Quantitative Determination of Hemoglobin in Blood, 3rd Edition. 2000. [Google Scholar]
- 41.Davis BH and Jungerius B, International Council for Standardization in Haematology technical report 1–2009: new reference material for haemiglobincyanide for use in standardization of blood haemoglobin measurements. Int J Lab Hematol, 2010. 32(2): p. 139–41. [DOI] [PubMed] [Google Scholar]
- 42.Srivastava T, et al. , Methods for Hemoglobin Estimation: A Review of “What Works”. Journal of Hematology and Transfusion, 2014. 2(3 : 1028): p. 1–7. [Google Scholar]
- 43.Tallqvist TW, Tallqvist’s Hämoglobin scale. Sharp & Smith: Chicago, Illinois. [Google Scholar]
- 44.Patel NR, Kabra AN, and Vora S, Hemoglobin Estimation by Invasive and Non-Invasive Methods: A Review. International Journal of Science, Engineering and Management, 2019. 4(1): p. 1–7. [Google Scholar]
- 45.Moharram NM, et al. , International collaborative assessment study of the AHD575 method for the measurement of blood haemoglobin. East Mediterr Health J, 2006. 12(6): p. 722–34. [PubMed] [Google Scholar]
- 46.Antwi-Baffour S, et al. , Anemia in prospective blood donors deferred by the copper sulphate technique of hemoglobin estimation. BMC Hematol, 2015. 15(15): p. 1–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Elwood PC and Jacobs A, Haemoglobin estimation: a comparison of different techniques. Br Med J, 1966. 1(5478): p. 20–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Balasubramaniam P and Malathi A, Comparative study of hemoglobin estimated by Drabkin’s and Sahli’s methods. J Postgrad Med, 1992. 38(1): p. 8–9. [PubMed] [Google Scholar]
- 49.Shah VB, Shah BS, and Puranik GV, Evaluation of non cyanide methods for hemoglobin estimation. Indian J Pathol Microbiol, 2011. 54(4): p. 764–8. [DOI] [PubMed] [Google Scholar]
- 50.Shah PP, et al. , Assessing diagnostic accuracy of Haemoglobin Colour Scale in real-life setting. Journal of health, population, and nutrition, 2014. 32(1): p. 51–57. [PMC free article] [PubMed] [Google Scholar]
- 51.Anand H, Mir R, and Saxena R, Hemoglobin color scale a diagnostic dilemma. Indian J Pathol Microbiol, 2009. 52(3): p. 360–2. [DOI] [PubMed] [Google Scholar]
- 52.Okoroma E and Charney E, The copper sulfate methodof screening for anemia in childhood. The Journal of Pediatrics, 1968. 72(3): p. 384–387. [DOI] [PubMed] [Google Scholar]
- 53.Funk M, et al. , Screening for childhood anaemia using copper sulphate densitometry. S Afr Med J, 2002. 92(12): p. 978–82. [PubMed] [Google Scholar]
- 54.Jöbsis FF, Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science, 1977. 198(4323): p. 1264–7. [DOI] [PubMed] [Google Scholar]
- 55.Edwards AD, et al. , Measurement of hemoglobin flow and blood flow by near-infrared spectroscopy. J Appl Physiol (1985), 1993. 75(4): p. 1884–9. [DOI] [PubMed] [Google Scholar]
- 56.Zhou Y, et al. , Measurement of hemoglobin in whole blood using a partial least squares regression model with selected second derivative near infrared transmission spectral signals. Biochemical and Biophysical Research Communications, 2012. 420(1): p. 205–209. [DOI] [PubMed] [Google Scholar]
- 57.Lindner G, Exadaktylos AK, How noninvasive hemoglobin measurement with pulse co-oximetry can change your practice: an expert review. Emergency Medicine International, 2013. 701529. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Avcioglu G, et al. , Comparison of noninvasive and invasive point-of-care testing methods with reference method for hemoglobin measurement. Journal of Clinical Laboratory Analysis, 2018. 32(3 - e22309): p. 1–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Ajmal AA, et al. Development of the Relationship between Hemoglobin Concentration and PPG Signal Characteristics. in 2019 14th Conference on Industrial and Information Systems (ICIIS). 2019. [Google Scholar]
- 60.Al-Baradie RS and Bose ASC. Portable smart non-invasive hemoglobin measurement system. in 10th International Multi-Conferences on Systems, Signals & Devices 2013 (SSD13). 2013. IEEE. [Google Scholar]
- 61.Hasan MK, et al. , HeLP ME: Recommendations for Non-invasive Hemoglobin Level Prediction in Mobile-phone Environment. JMIR mHealth uHealth, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.McMurdy JW, et al. , Noninvasive optical, electrical, and acoustic methods of total hemoglobin determination. Clin Chem, 2008. 54(2): p. 264–72. [DOI] [PubMed] [Google Scholar]
- 63.Organization TWH, Haemoglobin color scale practical answer to a vital need. The World Health Organization, Geneva: WHO, 2001. [Google Scholar]
- 64.Lindblade KA, et al. , Evaluation of the WHO Haemoglobin Colour Scale for diagnosis of anaemia in children and pregnant women as used by primary health care nurses and community health workers in western Kenya. Trop Med Int Health, 2006. 11(11): p. 1679–87. [DOI] [PubMed] [Google Scholar]
- 65.Madakshira M, Joy J, and Dudeja P, World Health Organization hemoglobin color scale: A useful point-of-care test to detect anemia. Medical Journal of Dr. D.Y. Patil Vidyapeeth, 2018. 11(2): p. 133–136. [Google Scholar]
- 66.Stott GJ and Lewis SM, A simple and reliable method for estimating haemoglobin. Bulletin of the World Health Organization, 1995. 73(3): p. 369–373. [PMC free article] [PubMed] [Google Scholar]
- 67.Lewis SM, Stott GJ, and Wynn KJ, An inexpensive and reliable new haemoglobin colour scale for assessing anaemia. Journal of clinical pathology, 1998. 51(1): p. 21–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Timan IS, et al. , Comparison of haemoglobinometry by WHO Haemoglobin Colour Scale and copper sulphate against haemiglobincyanide reference method. Clin Lab Haematol, 2004. 26(4): p. 253–8. [DOI] [PubMed] [Google Scholar]
- 69.Khan AA, Fatmi Z, and Kadir MM, Accuracy and Use of WHO Hemoglobin Color Scale for Diagnosis of Anemia Among Pregnant Women by Health Care Providers in Periurban Settings in Karachi, Pakistan. Asia Pac J Public Health, 2015. 27(6): p. 610–9. [DOI] [PubMed] [Google Scholar]
- 70.Marn H and Critchley JA, Accuracy of the WHO Haemoglobin Colour Scale for the Diagnosis of Anaemia in Primary Health Care Settings in Low-Income Countries: A Systematic Review and Meta-Analysis. Lancet Glob Health, 2016. 4: p. e251–65. [DOI] [PubMed] [Google Scholar]
- 71.Darshana LGT and Uluwaduge DI, Validation of the WHO Hemoglobin Color Scale Method. Anemia, 2014. 2014: p. 4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Adam I, et al. , Comparison of HemoCue(R) hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan. Diagn Pathol, 2012. 7: p. 30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.von Schenck H, Falkensson M, and Lundberg B, Evaluation of “HemoCue,” a new device for determining hemoglobin. Clin Chem, 1986. 32(3): p. 526–9. [PubMed] [Google Scholar]
- 74.Akhtar K, et al. , HemoCue photometer: a better alternative of hemoglobin estimation in blood donors? Indian journal of hematology & blood transfusion : an official journal of Indian Society of Hematology and Blood Transfusion, 2008. 24(4): p. 170–172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Sari M, et al. , Estimating the prevalence of anaemia: a comparison of three methods. Bull World Health Organ, 2001. 79(6): p. 506–11. [PMC free article] [PubMed] [Google Scholar]
- 76.Hinnouho GM, et al. , Comparison of haemoglobin assessments by HemoCue and two automated haematology analysers in young Laotian children. J Clin Pathol, 2018. 71(6): p. 532–538. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Kim MJ, et al. , Comparison of the Accuracy of Noninvasive Hemoglobin Sensor (NBM-200) and Portable Hemoglobinometer (HemoCue) with an Automated Hematology Analyzer (LH500) in Blood Donor Screening. Annals of Laboratory Medicine, 2013. 33(4): p. 261–267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Paddle JJ, Evaluation of the Haemoglobin Colour Scale and comparison with the HemoCue haemoglobin assay. Bulletin of the World Health Organization, 2002. 80(10): p. 813–816. [PMC free article] [PubMed] [Google Scholar]
- 79.Baart AM, et al. , Hemoglobin assessment: precision and practicability evaluated in the Netherlands-the HAPPEN study. Transfusion, 2016. 56(8): p. 1984–93. [DOI] [PubMed] [Google Scholar]
- 80.Mendrone A Jr., et al. , Anemia screening in potential female blood donors: comparison of two different quantitative methods. Transfusion, 2009. 49(4): p. 662–8. [DOI] [PubMed] [Google Scholar]
- 81.Sümnig A, et al. , The impact of noninvasive, capillary, and venous hemoglobin screening on donor deferrals and the hemoglobin content of red blood cells concentrates: a prospective study. Transfusion, 2015. 55(12): p. 2847–2854. [DOI] [PubMed] [Google Scholar]
- 82.Sobhy S, Rogozinska E, and Khan KS, Accuracy of on-site tests to detect anemia during prenatal care. International Journal of Gynecology & Obstetrics, 2017. 139(2): p. 130–136. [DOI] [PubMed] [Google Scholar]
- 83.Yadav K, et al. , Validation of Point of Care Hemoglobin Estimation Among Pregnant Women Using Digital Hemoglobinometers (HemoCue 301 and HemoCue 201+) as Compared with Auto-Analyzer. Indian Journal of Hematology and Blood Transfusion, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Neville RG, Evaluation of portable haemoglobinometer in general practice. Br Med J (Clin Res Ed), 1987. 294(6582): p. 1263–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Sanchis-Gomar F, et al. , Hemoglobin point-of-care testing: the HemoCue system. J Lab Autom, 2013. 18(3): p. 198–205. [DOI] [PubMed] [Google Scholar]
- 86.Goldman M, et al. , Validation and implementation of a new hemoglobinometer for donor screening at Canadian Blood Services. Transfusion, 2012. 52(7 Pt 2): p. 1607–13. [DOI] [PubMed] [Google Scholar]
- 87.Yadav K, et al. , Use of Point of Care Testing (POCT) in measurement of hemoglobin. Indian Journal of Community Health, 2018. 30(01): p. 72–79. [Google Scholar]
- 88.Chaudhary R, Dubey A, and Sonker A, Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions. J Blood Med, 2017. 8: p. 75–88. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Malukani P, Gajjar MD, Gonsai RN, Bhatnagar N, Goswami HM, Study Of Newer Invasive And Non-Invasive Methods Of Haemoglobin Estimation In Blood Donor Screening - A Study On 200 Donors. IJCRR, 2014. 6(4): p. 26–30. [Google Scholar]
- 90.Singh A, et al. , Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors. Blood Transfus, 2015. 13(2): p. 233–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Ranjan V, Rana R, Khillan K, Chauhan K, A comparative quality evaluation of point-of-care methodology for testing hemoglobin in blood donors by two different technologies. Current Medicine Research and Practice, 2020. 10: p. 90–92. [Google Scholar]
- 92.Robertson L, Lewis D, and Osei-Bimpong A, Utility of the DiaSpect Hemoglobinometry System as a Point-of-Care Analyzer. Point of Care, 2011. 10(1): p. 17–21. [Google Scholar]
- 93.Oladele O, et al. , Point-of-Care Testing for Anaemia in Children Using Portable Haematocrit Meter: A Pilot Study from Southwest Nigeria and Implications for Developing Countries. Ethiop J Health Sci, 2016. 26(3): p. 251–258. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Ahmad NA, et al. , Validity of Point-of-Care Testing Mission Plus in Detecting Anemia. International Journal of Biomedicine, 2015. 5(2): p. 91–94. [Google Scholar]
- 95.McGann PT, et al. , An Accurate and Rapid Color-Based Point-of-Care Assay for the Detection of Severe Anemia in Low Resource Settings. Blood, 2014. 124(21): p. 688. [Google Scholar]
- 96.McGann PT, et al. , An accurate and inexpensive color-based assay for detecting severe anemia in a limited-resource setting. American Journal of Hematology, 2015. 90(12): p. 1122–1127. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Perez-Plazola M, et al. , A Simple, Rapid, and Inexpensive Color-Based Hemoglobin Assay As a Robust Screening Test for Severe Anemia in Limited Resource Settings. Blood, 2018. 132(Suppl. 1, 4724): p. 1–7.29976776 [Google Scholar]
- 98.Perez-Plazola M, Tyburski EA, Tyburski EA, Smart LR, Howard TA, Pfeiffer A, Ware RE, Lam WA, and McGann PT, AnemoCheck-LRS: an optimized, colorbased point-of-care test to identify severe anemia in limited-resource settings. BMC Medicine 2020. 18(337): p. 1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Shamah Levy T, et al. , Validation of Masimo Pronto 7 and HemoCue 201 for hemoglobin determination in children from 1 to 5 years of age. PLOS ONE, 2017. 12(2): p. e0170990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Skelton VA, et al. , Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue® and laboratory haemoglobin measurements in obstetric patients*. Anaesthesia, 2013. 68(1): p. 40–45. [DOI] [PubMed] [Google Scholar]
- 101.Osborn ZT, et al. , Accuracy of Point-of-Care Testing for Anemia in the Emergency Department. Respir Care, 2019. 64(11): p. 1343–1350. [DOI] [PubMed] [Google Scholar]
- 102.DeBarros M, et al. , Assessing serum hemoglobin levels without venipuncture: accuracy and reliability of Pronto-7 noninvasive spot-check device. Am J Surg, 2015. 209(5): p. 848–55. [DOI] [PubMed] [Google Scholar]
- 103.Yoshida A, Saito K, Ishii K, Azuma I, Sasa H, Furuya K, Assessment of noninvasive, percutaneous hemoglobin measurement in pregnant and early postpartum women. Medical Devices: Evidence & Research, 2014. 7: p. 11–16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Khalafallah AA, Chilvers CR, Thomas M, Chilvers CM, Sexton M, Vialle M, and Robertson IK, Usefulness of non-invasive spectrophotometric haemoglobin estimation for detecting low haemoglobin levels when compared with a standard laboratory assay for preoperative assessment. British Journal of Anaesthesia, 2015. 114(4): p. 669–676. [DOI] [PubMed] [Google Scholar]
- 105.Belardinelli A, et al. , Noninvasive methods for haemoglobin screening in prospective blood donors. Vox Sanguinis, 2013. 105(2): p. 116–120. [DOI] [PubMed] [Google Scholar]
- 106.Hiscock R, Kumar D, and Simmons SW, Systematic Review and Meta-Analysis of Method Comparison Studies of Masimo Pulse Co-Oximeters (Radical-7™ or Pronto-7™) and HemoCue® Absorption Spectrometers (B-Hemoglobin or 201+) with Laboratory Haemoglobin Estimation. Anaesthesia and Intensive Care, 2015. 43(3): p. 341–350. [DOI] [PubMed] [Google Scholar]
- 107.Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M, Non-Invasive Measurement of Hemoglobin: Assessment of Two Different Point-of-Care Technologies. PLoS ONE, 2012. 7(1): p. 1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108.Hiscock R, et al. , Comparison of Massimo Pronto-7 and Hemocue Hb 201+ with Laboratory Haemoglobin Estimation: A Clinical Study. Anaesthesia and Intensive Care, 2014. 42(5): p. 608–613. [DOI] [PubMed] [Google Scholar]
- 109.Hsu DP, et al. , Evaluation of a Noninvasive Hemoglobin Measurement Device to Screen for Anemia in Infancy. Maternal and Child Health Journal, 2016. 20(4): p. 827–832. [DOI] [PubMed] [Google Scholar]
- 110.Kim J, et al. , Evaluation of Two Noninvasive Hemoglobin Testing Devices. The Korean Journal of Blood Transfusion, 2015. 26: p. 273–281. [Google Scholar]
- 111.Lee J-H, Park Y-H, and Kim J-T, Current Use of Noninvasive Hemoglobin Monitoring in Anesthesia. Current Anesthesiology Reports, 2014. 4(3): p. 233–241. [Google Scholar]
- 112.Ardin S, et al. , Comparison of three noninvasive methods for hemoglobin screening of blood donors. Transfusion, 2015. 55(2): p. 379–87. [DOI] [PubMed] [Google Scholar]
- 113.Crowley C, et al. , Validity and correspondence of non-invasively determined hemoglobin concentrations by two trans-cutaneous digital measuring devices. Asia Pacific Journal of Clinical Nutrition, 2012. 21(2): p. 191–200. [PubMed] [Google Scholar]
- 114.De Clippel D, Van Heddegem L, Vandewalle G, Vandekerckhove P, Compernolle V, Hemoglobin screening in blood donors: a prospective study assessing the value of an invasive and a noninvasive point-of-care device for donor safety. Transfusion, 2017. 57(4): p. 938–945. [DOI] [PubMed] [Google Scholar]
- 115.Debbie Strydom K.v.d.B., Non-invasive technology to determine the haemoglobin level of blood donors at the SANBS. Africa Sanguine, 2016. 18(1). [Google Scholar]
- 116.Begnoche V and O’Reilly M, Noninvasive Hemoglobin Monitoring, in Monitoring Technologies in Acute Care Environments: A Comprehensive Guide to Patient Monitoring Technology, Ehrenfeld JM and Cannesson M, Editors. 2014, Springer New York: New York, NY. p. 299–304. [Google Scholar]
- 117.Rabe H, et al. , Spectroscopic noninvasive measurement of hemoglobin compared with capillary and venous values in neonates. Neonatology, 2010. 98(1): p. 1–5. [DOI] [PubMed] [Google Scholar]
- 118.Pinto M, Barjas-Castro ML, Nascimento S, Falconi MA, Zulli R, and Castro V, The new noninvasive occlusion spectroscopy hemoglobin measurement method: a reliable and easy anemia screening test for blood donors. Transfusion, 2013. 53: p. 766–769. [DOI] [PubMed] [Google Scholar]
- 119.Weinstein A, Herzentien O, Gabis E, Korenberg A, Screening of anemia using occlusion spectroscopy. Transfusion, 2010. 50: p. 91A–92A. [Google Scholar]
- 120.Hadar E, et al. , Precision and accuracy of noninvasive hemoglobin measurements during pregnancy. The Journal of Maternal-Fetal & Neonatal Medicine, 2012. 25(12): p. 2503–2506. [DOI] [PubMed] [Google Scholar]
- 121.Ahankari AS, et al. , Assessment of a non-invasive haemoglobin sensor NBM 200 among pregnant women in rural India. BMJ Innovations, 2016. 2(2): p. 70. [Google Scholar]
- 122.Ahankari AS, et al. , Comparison of the NBM 200 non-invasive haemoglobin sensor with Sahli’s haemometer among adolescent girls in rural India. BMJ Innovations, 2016. 2(4): p. 144–148. [Google Scholar]
- 123.Lotfi R, et al. , A noninvasive strategy for screening prospective blood donors for anemia. Transfusion, 2005. 45(10): p. 1585–1592. [DOI] [PubMed] [Google Scholar]
- 124.Rout D, Sachdev S, Marwaha N, Evaluation of new non-invasive & conventional invasive methods of haemoglobin estimation in blood donors. Indian J Med Res 2019. 149: p. 755–762. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 125.Pradhan R, Mitra A, Das S, Impedimetric characterization of human blood using three-electrode based ECIS devices. J. Electr. Bioimpedance, 2012. 3: p. 12–19. [Google Scholar]
- 126.McGann PT, et al. , Genetic Causes of Anemia in Malawian Children Less Than 5 Years of Age: Results from the Malawi Demographic and Health Survey. Blood, 2016. 128(22): p. 5.27389537 [Google Scholar]
- 127.Chakraborty S, Das S, Das C, Chandra S, Sharma KD, Karmakar A, Chattoapadhyay S, On-chip estimation of hematocrit level for diagnosing anemic conditions by Impedimetric techniques. Biomedical Microdevices: BioMEMS and Biomedical Nanotechnology, 2020. 22(38): p. 1–11. [DOI] [PubMed] [Google Scholar]
- 128.Punter-Villagrasa J, et al. , Toward an anemia early detection device based on 50-muL whole blood sample. IEEE Trans Biomed Eng, 2015. 62(2): p. 708–16. [DOI] [PubMed] [Google Scholar]
- 129.Punter-Villagrasa J, et al. , An instantaneous low-cost point-of-care anemia detection device. Sensors (Basel, Switzerland), 2015. 15(2): p. 4564–4577. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Phillips JP, Hickey M, and Kyriacou PA, Evaluation of Electrical and Optical Plethysmography Sensors for Noninvasive Monitoring of Hemoglobin Concentration. Sensors (14248220), 2012. 12(2): p. 1816–1826. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131.Ho TLT, et al. , Determining mean corpuscular volume and red blood cell count using electrochemical collision events. Biosensors and Bioelectronics, 2018. 110: p. 155–159. [DOI] [PubMed] [Google Scholar]
- 132.Pashchenko O, et al. , A Comparison of Optical, Electrochemical, Magnetic, and Colorimetric Point-of-Care Biosensors for Infectious Disease Diagnosis. Acs Infectious Diseases, 2018. 4(8): p. 1162–1178. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 133.(CLSI), C.L.S.I., Evaluation of Precision of Quantitative Measurement Procedures, 3rd Edition (EP05-A3). 2014.
- 134.(CLSI), C.L.S.I., Evaluation of Stability of In Vitro Diagnostic Reagents (EP25-A). 2009.
- 135.CHen LC, Wang E, Tai CS, Chiu YC, Li CW, Lin YR, Lee TH, Huang CW, Chen JC, Chen WL, Improving the reproducibility, accuracy, and stability of an electrochemical biosensor platform for point-of-care use. Biosensors and Bioelectronics, 2020. 155: p. 1–8. [DOI] [PubMed] [Google Scholar]
- 136.Ronkainen NJ, Halsall HB, Heineman WR, Electrochemical Biosensors. Chemical Society Reviews, 2010. 39: p. 1747–1763. [DOI] [PubMed] [Google Scholar]
- 137.da Silva ETSG, et al. , Electrochemical Biosensors in Point-of-Care Devices: Recent Advances and Future Trends. ChemElectroChem, 2017. 4(4): p. 778–794. [Google Scholar]
- 138.Kim HJ, et al. , Disposable microfluidic blood cuvette for measuring hemoglobin concentration. Microsystem Technologies, 2014. 20(3): p. 499–504. [Google Scholar]
- 139.Plevniak K, Campbell M, and Mei H, 3D printed microfluidic mixer for point-of-care diagnosis of anemia. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference, 2016. 2016: p. 267–270. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Taparia N, et al. , A microfluidic approach for hemoglobin detection in whole blood. AIP Advances, 2017. 7(10): p. 105102. [Google Scholar]
- 141.Halder A, et al. , Digital Camera-Based Spectrometry for the Development of Point-of-Care Anemia Detection on Ultra-Low Volume Whole Blood Sample. IEEE Sensors Journal, 2017. 17(21): p. 7149–7156. [Google Scholar]
- 142.Zhu H, et al. , Cost-effective and rapid blood analysis on a cell-phone. Lab Chip, 2013. 13(7): p. 1282–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 143.Zhu H, et al. , Optofluidic Fluorescent Imaging Cytometry on a Cell Phone. Analytical Chemistry, 2011. 83(17): p. 6641–6647. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 144.Berry SB, et al. , Measurement of the hematocrit using paper-based microfluidic devices. Lab on a Chip, 2016. 16(19): p. 3689–3694. [DOI] [PubMed] [Google Scholar]
- 145.Hasan MN, et al. , Paper-based microchip electrophoresis for point-of-care hemoglobin testing. Analyst, 2020. 145(7): p. 2525–2542. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 146.An R, et al. , Integrated Anemia Detection and Hemoglobin Variant Identification Using Point-of-Care Microchip Electrophoresis. Blood, 2019. 134(Supplement_1): p. 378–378. [Google Scholar]
- 147.An R, et al. , Computer Vision and Deep Learning Assisted Microchip Electrophoresis for Integrated Anemia and Sickle Cell Disease Screening. Blood, 2020. 136(Supplement 1): p. 46–47. [Google Scholar]
- 148.Fraiwan A, et al. , International Multi-Site Clinical Validation of Point-of-Care Microchip Electrophoresis Test for Hemoglobin Variant Identification. Blood, 2019. 134(Supplement_1): p. 3373–3373. [Google Scholar]
- 149.Hasan MN, et al. , Hemechip: An Automated Portable Microchip Electrophoresis Platform for Point-of-Care Sickle Cell Disease Screening. Blood, 2017. 130(Supplement 1): p. 3519–3519. [Google Scholar]
- 150.Hasan MN, et al. Live Demonstration: HemeChip - A Portable Microchip Electrophoresis Technology for Point-of-Care Sickle Cell Disease Screening. in 2018 IEEE Biomedical Circuits and Systems Conference (BioCAS). 2018. [Google Scholar]
- 151.Hasan MN, et al. , Clinical Testing of Hemechip in Nigeria for Point-of-Care Screening of Sickle Cell Disease. Blood, 2018. 132(Supplement 1): p. 1095–1095.29954749 [Google Scholar]
- 152.Ung R, et al. , Point-of-Care Screening for Sickle Cell Disease By a Mobile Micro-Electrophoresis Platform. Blood, 2015. 126(23): p. 3379–3379. [Google Scholar]
- 153.Dimauro G, et al. , Estimate of Anemia with New Non-Invasive Systems—A Moment of Reflection. Electronics, 2020. 9(5): p. 780. [Google Scholar]
- 154.Kalantri A, et al. , Accuracy and reliability of pallor for detecting anaemia: a hospital-based diagnostic accuracy study. PloS one, 2010. 5(1): p. e8545–e8545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 155.Sheth TN, et al. , The relation of conjunctival pallor to the presence of anemia. Journal of general internal medicine, 1997. 12(2): p. 102–106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 156.Stoltzfus RJ, et al. , Clinical pallor is useful to detect severe anemia in populations where anemia is prevalent and severe. J Nutr, 1999. 129(9): p. 1675–81. [DOI] [PubMed] [Google Scholar]
- 157.Butt Z, et al. , Diagnostic accuracy of “pallor” for detecting mild and severe anaemia in hospitalized patients. J Pak Med Assoc, 2010. 60(9): p. 762–5. [PubMed] [Google Scholar]
- 158.Chalco JP, et al. , Accuracy of clinical pallor in the diagnosis of anaemia in children: a meta-analysis. BMC Pediatrics, 2005. 5(1): p. 46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 159.Benseñor IM, et al. , Accuracy of anemia diagnosis by physical examination. Sao Paulo Med J, 2007. 125(3): p. 170–3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 160.Ding H, et al. , Non-invasive prediction of hemoglobin levels by principal component and back propagation artificial neural network. Biomedical Optics Express, 2014. 5(4): p. 1145–1152. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 161.Kraitl J, Ewald H, and Gehring H, An optical device to measure blood components by a photoplethysmographic method. Journal of Optics A: Pure and Applied Optics, 2005. 7(6): p. S318–S324. [Google Scholar]
- 162.Kye Jin J, et al. , Noninvasive total hemoglobin measurement. Journal of Biomedical Optics, 2002. 7(1): p. 45–50. [DOI] [PubMed] [Google Scholar]
- 163.Yoon G, Kim SJ, and Jeon KJ, Robust design of finger probe in non-invasive total haemoglobin monitor. Medical and Biological Engineering and Computing, 2005. 43(1): p. 121–125. [DOI] [PubMed] [Google Scholar]
- 164.Yuan J, et al. , Research on improving the accuracy of near infrared non-invasive hemoglobin detection. Infrared Physics & Technology, 2015. 72: p. 117–121. [Google Scholar]
- 165.Bui N, et al. , PhO2: Smartphone based Blood Oxygen Level Measurement Systems using Near-IR and RED Wave-guided Light, in Proceedings of the 15th ACM Conference on Embedded Network Sensor Systems. 2017, Association for Computing Machinery: Delft, Netherlands. p. Article 26. [Google Scholar]
- 166.Zhang S, et al. , Investigation of Noninvasive in Vivo Blood Hematocrit Measurement Using NIR Reflectance Spectroscopy and Partial Least-Squares Regression. Applied Spectroscopy, 2000. 54(2): p. 294–299. [Google Scholar]
- 167.Dullenkopf A, et al. , Non-invasive monitoring of haemoglobin concentration in paediatric surgical patients using near-infrared spectroscopy. Anaesthesia, 2004. 59(5): p. 453–8. [DOI] [PubMed] [Google Scholar]
- 168.Srinivasan S, et al. , Interpreting hemoglobin and water concentration, oxygen saturation, and scattering measured in vivo by near-infrared breast tomography. Proceedings of the National Academy of Sciences, 2003. 100(21): p. 12349. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 169.Kuenstner JT and Norris KH, Near Infrared Hemoglobinometry. Journal of Near Infrared Spectroscopy, 1995. 3(1): p. 11–18. [Google Scholar]
- 170.Yi X, Li G, and Lin L, Noninvasive hemoglobin measurement using dynamic spectrum. Review of Scientific Instruments, 2017. 88(8): p. 083109. [DOI] [PubMed] [Google Scholar]
- 171.Zhang L, et al. , Performance of calibration model with different ratio of sample size to the number of wavelength: Application to hemoglobin determination by NIR spectroscopy. Spectrochim Acta A Mol Biomol Spectrosc, 2020. 227: p. 117750. [DOI] [PubMed] [Google Scholar]
- 172.Zhang L, et al. , Optimal wavelength selection for visible diffuse reflectance spectroscopy discriminating human and nonhuman blood species. Analytical Methods, 2016. 8(2): p. 381–385. [Google Scholar]
- 173.Askari M, A near infrared plasmonic perfect absorber as a sensor for hemoglobin concentration detection. Optical and Quantum Electronics, 2021. 53(2): p. 67. [Google Scholar]
- 174.Pasquini C, Near infrared spectroscopy: A mature analytical technique with new perspectives – A review. Analytica Chimica Acta, 2018. 1026: p. 8–36. [DOI] [PubMed] [Google Scholar]
- 175.Herold F, et al. , Applications of Functional Near-Infrared Spectroscopy (fNIRS) Neuroimaging in Exercise–Cognition Science: A Systematic, Methodology-Focused Review. Journal of Clinical Medicine, 2018. 7(12): p. 466. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 176.Dimauro G, et al. , A Systematic Mapping Study on Research in Anemia Assessment with Non-Invasive Devices. Applied Sciences, 2020. 10(14): p. 4804. [Google Scholar]
- 177.Tian H, et al. , Optical wavelength selection for portable hemoglobin determination by near-infrared spectroscopy method. Infrared Physics & Technology, 2017. 86: p. 98–102. [Google Scholar]
- 178.AvakhKisomi A, et al. A novel wireless ring-shaped multi-site pulse oximeter. in 2016 IEEE International Symposium on Circuits and Systems (ISCAS). 2016. [Google Scholar]
- 179.Huang C-Y, et al. , Novel wearable and wireless ring-type pulse oximeter with multi-detectors. Sensors (Basel, Switzerland), 2014. 14(9): p. 17586–17599. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180.Li G, et al. , Noninvasive hemoglobin measurement based on optimizing Dynamic Spectrum method. Spectroscopy Letters, 2017. 50(3): p. 164–170. [Google Scholar]
- 181.Azarnoosh M and Doostdar H, Assessment of Photoplethysmography Method in Extraction of Hemoglobin Concentration. Journal of biomedical physics & engineering, 2019. 9(6): p. 711–718. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Timm U, et al. Sensor system for non-invasive optical hemoglobin determination. in SENSORS, 2009 IEEE. 2009. [Google Scholar]
- 183.Doshi R and Panditrao A. Non-Invasive Optical Sensor for Hemoglobin Determination. 2013.
- 184.Nirupa JLA and Kumar VJ. Non-invasive measurement of hemoglobin content in blood. in 2014 IEEE International Symposium on Medical Measurements and Applications (MeMeA). 2014. [Google Scholar]
- 185.Liu H, et al. , Development and Validation of a Photoplethysmography System for Noninvasive Monitoring of Hemoglobin Concentration. Journal of Electrical and Computer Engineering, 2020. 2020: p. 3034260. [Google Scholar]
- 186.Glaros KN and Drakakis EM, A sub-mW fully-integrated pulse oximeter front-end. IEEE Trans Biomed Circuits Syst, 2013. 7(3): p. 363–75. [DOI] [PubMed] [Google Scholar]
- 187.Schidl S, Kaniusas E, and Zimmermann H. CMOS chip with multi junction photo detector for sensing biomedical signals. in SENSORS, 2013 IEEE. 2013. [Google Scholar]
- 188.Suzaki H, et al. , Noninvasive measurement of total hemoglobin and hemoglobin derivatives using multiwavelength pulse spectrophotometry -In vitro study with a mock circulatory system. Conf Proc IEEE Eng Med Biol Soc, 2006. 2006: p. 799–802. [DOI] [PubMed] [Google Scholar]
- 189.Costa J, et al. , Double junction photodiodes for multiwavelength photoplethysmography. 2018. 89. [Google Scholar]
- 190.Yuan H, et al. , Motion artefact minimization from photoplethysmography based non-invasive hemoglobin sensor based on an envelope filtering algorithm. Measurement, 2018. 115: p. 288–298. [Google Scholar]
- 191.Golap MA and Hashem MMA. Non-Invasive Hemoglobin Concentration Measurement Using MGGP-Based Model. in 2019 5th International Conference on Advances in Electrical Engineering (ICAEE). 2019. [Google Scholar]
- 192.Kavsaoğlu AR, Polat K, and Hariharan M, Non-invasive prediction of hemoglobin level using machine learning techniques with the PPG signal’s characteristics features. Applied Soft Computing, 2015. 37: p. 983–991. [Google Scholar]
- 193.Acharya S, et al. , Non-Invasive Estimation of Hemoglobin Using a Multi-Model Stacking Regressor. IEEE J Biomed Health Inform, 2020. 24(6): p. 1717–1726. [DOI] [PubMed] [Google Scholar]
- 194.Castaneda D, et al. , A review on wearable photoplethysmography sensors and their potential future applications in health care. International journal of biosensors & bioelectronics, 2018. 4(4): p. 195–202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 195.Johnson MSS and Mikael Eklund J, A Review of Photoplethysmography-based Physiological Measurement and Estimation, Part 1: Single Input Methods. Annu Int Conf IEEE Eng Med Biol Soc, 2020. 2020: p. 923–927. [DOI] [PubMed] [Google Scholar]
- 196.Johnson MSS and Mikael Eklund J, A Review of Photoplethysmography-based Physiological Measurement and Estimation, Part 2: Multi-input Methods. Annu Int Conf IEEE Eng Med Biol Soc, 2020. 2020: p. 863–866. [DOI] [PubMed] [Google Scholar]
- 197.Suner S, et al. , Non-invasive determination of hemoglobin by digital photography of palpebral conjunctiva. The Journal of emergency medicine, 2007. 33(2): p. 105–111. [DOI] [PubMed] [Google Scholar]
- 198.Bevilacqua V, et al. A novel approach to evaluate blood parameters using computer vision techniques. in 2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA). 2016. IEEE. [Google Scholar]
- 199.Chen Y-M, Miaou S-G, and Bian H, Examining palpebral conjunctiva for anemia assessment with image processing methods. Computer methods and programs in biomedicine, 2016. 137: p. 125–135. [DOI] [PubMed] [Google Scholar]
- 200.Chen Y-M and Miaou S-G, A Kalman filtering and nonlinear penalty regression approach for noninvasive anemia detection with palpebral conjunctiva images. Journal of healthcare engineering, 2017. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 201.Dimauro G, Caivano D, and Girardi F, A new method and a non-invasive device to estimate anemia based on digital images of the conjunctiva. IEEE Access, 2018. 6: p. 46968–46975. [Google Scholar]
- 202.Dimauro G, et al. Automatic segmentation of relevant sections of the conjunctiva for non-invasive anemia detection. in 2018 3rd International Conference on Smart and Sustainable Technologies (SpliTech). 2018. IEEE. [Google Scholar]
- 203.Collings S, et al. , Non-Invasive Detection of Anaemia Using Digital Photographs of the Conjunctiva. PLOS ONE, 2016. 11(4): p. e0153286. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 204.Dimauro G A new image quality metric based on human visual system. in 2012 IEEE International Conference on Virtual Environments Human-Computer Interfaces and Measurement Systems (VECIMS) Proceedings. 2012. [Google Scholar]
- 205.Dimauro G, Altomare N, and Scalera M. PQMET: A digital image quality metric based on human visual system. in 2014 4th International Conference on Image Processing Theory, Tools and Applications (IPTA). 2014. [Google Scholar]
- 206.Tamir A, et al. Detection of anemia from image of the anterior conjunctiva of the eye by image processing and thresholding. in 2017 IEEE Region 10 Humanitarian Technology Conference (R10-HTC). 2017. IEEE. [Google Scholar]
- 207.Anggraeni M and Fatoni A. Non-invasive self-care anemia detection during pregnancy using a smartphone camera. in IOP Conference Series: Materials Science and Engineering. 2017. IOP Publishing. [Google Scholar]
- 208.Sevani N and Persulessy G. Detection anemia based on conjunctiva pallor level using k-means algorithm. in Materials Science and Engineering Conference Series. 2018. [Google Scholar]
- 209.Muthalagu R, Thulasi Bai V, and John S, A Smart (phone) Solution: An effective tool for Screening Anaemia-Correlation with conjunctiva pallor and haemoglobin levels. TAGA Journal, 2018. 14: p. 2611–2621. [Google Scholar]
- 210.Hasan MK, et al. , SmartHeLP: Smartphone-based Hemoglobin Level Prediction Using an Artificial Neural Network. AMIA … Annual Symposium proceedings. AMIA Symposium, 2018. 2018: p. 535–544. [PMC free article] [PubMed] [Google Scholar]
- 211.Wang EJ, et al. Noninvasive hemoglobin measurement using unmodified smartphone camera and white flash. in 2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). 2017. IEEE. [DOI] [PubMed] [Google Scholar]
- 212.Wang EJ, et al. HemaApp IR: noninvasive hemoglobin measurement using unmodified smartphone cameras and built-in LEDs. in Proceedings of the 2017 ACM International Joint Conference on Pervasive and Ubiquitous Computing and Proceedings of the 2017 ACM International Symposium on Wearable Computers. 2017. [Google Scholar]
- 213.Wang EJ, et al. HemaApp: noninvasive blood screening of hemoglobin using smartphone cameras. in Proceedings of the 2016 ACM International Joint Conference on Pervasive and Ubiquitous Computing. 2016. [Google Scholar]
- 214.Ahsan GM, et al. A novel real-time non-invasive hemoglobin level detection using video images from smartphone camera. in 2017 IEEE 41st Annual Computer Software and Applications Conference (COMPSAC). 2017. IEEE. [Google Scholar]
- 215.Hasan MK, et al. , Smartphone-based human hemoglobin level measurement analyzing pixel intensity of a fingertip video on different color spaces. Smart Health, 2018. 5: p. 26–39. [Google Scholar]
- 216.Mannino RG, et al. , Smartphone app for non-invasive detection of anemia using only patient-sourced photos. Nature communications, 2018. 9(1): p. 1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 217.Mitani A, et al. , Detection of anaemia from retinal fundus images via deep learning. Nature Biomedical Engineering, 2020. 4(1): p. 18–27. [DOI] [PubMed] [Google Scholar]
- 218.Park SM, et al. , mHealth spectroscopy of blood hemoglobin with spectral super-resolution. Optica, 2020. 7(6): p. 563–573. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 219.van Berkel C, et al. , Integrated systems for rapid point of care (PoC) blood cell analysis. Lab on a Chip, 2011. 11(7): p. 1249–1255. [DOI] [PubMed] [Google Scholar]
- 220.Mielczarek WS, et al. , Microfluidic blood plasma separation for medical diagnostics: is it worth it? Lab Chip, 2016. 16(18): p. 3441–8. [DOI] [PubMed] [Google Scholar]
- 221.Yu ZTF, Aw Yong KM, and Fu J, Microfluidic Blood Cell Sorting: Now and Beyond. Small, 2014. 10(9): p. 1687–1703. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 222.Kassebaum NJ, et al. , A systematic analysis of global anemia burden from 1990 to 2010. 2014. 123(5): p. 615–624. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 223.(NASCOP), N.A.a.S.C.P., Kenya AIDS Indicator Survey 2012: Final Report. 2014. 2014.
- 224.Milner EM, Kariger P, Pickering AJ, Stewart CP, Byrd K, Lin A, Rao G, Achando B, Dentz HN, Null C, Fernald LCH, Association between Malaria Infection and Early Childhood Development Mediated by Anemia in Rural Kenya. Int J Environ Res Public Health, 2020. 17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 225.Smithson P, Impact of Malaria Control on Mortality and Anemia among Tanzanian Children Less than Five Years of Age, 1999–2010. PLoS ONE, 2015. 10(11). [DOI] [PMC free article] [PubMed] [Google Scholar]
- 226.Smart LR, et al. , Simultaneous point-of-care detection of anemia and sickle cell disease in Tanzania: the RAPID study. Ann Hematol, 2018. 97(2): p. 239–246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 227.Young N, et al. , Integrated point-of-care testing (POCT) of HIV, syphilis, malaria and anaemia in antenatal clinics in western Kenya: A longitudinal implementation study. PLoS One, 2018. 13(7): p. e0198784. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 228.Bossuyt PM, et al. , STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies. Bmj, 2015. 351: p. h5527. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 229.Nichols JH, et al. , AACC Guidance Document on Management of Point-of-Care Testing. J Appl Lab Med, 2020. 5(4): p. 762–787. [DOI] [PubMed] [Google Scholar]
- 230.Fletcher R, et al. Development of smart phone tools for printed diagnostics: Challenges and solutions. in 2016 IEEE Global Humanitarian Technology Conference (GHTC). 2016. [Google Scholar]
- 231. https://www.masimo.cn/pronto-7/index.htm.
- 232. http://www.attikourismedical.com/product/diagnostic/haemospect-device/.
- 233.Plevniak K, et al. , 3D printed auto-mixing chip enables rapid smartphone diagnosis of anemia. Biomicrofluidics, 2016. 10(5): p. 054113–054113. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 234.Avcioglu G, et al. , Comparison of noninvasive and invasive point-of-care testing methods with reference method for hemoglobin measurement. Journal of Clinical Laboratory Analysis, 2018. 32(3): p. e22309. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 235.Biswas SK, et al. , Anemia Diagnosis on a Simple Paper-based Assay. bioRxiv, 2018: p. 439224. [Google Scholar]
- 236.Kumar MR, Mahadevappa M, and Goswami D. Low cost point of care estimation of Hemoglobin levels. in 2014 International Conference on Medical Imaging, m-Health and Emerging Communication Systems (MedCom). 2014. [Google Scholar]
- 237.Kim O, et al. , Combined reflectance spectroscopy and stochastic modeling approach for noninvasive hemoglobin determination via palpebral conjunctiva. Physiological Reports, 2014. 2(1): p. e00192. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 238.Lamhaut L, et al. , Comparison of the Accuracy of Noninvasive Hemoglobin Monitoring by Spectrophotometry (SpHb) and HemoCue® with Automated Laboratory Hemoglobin Measurement. Anesthesiology, 2011. 115(3): p. 548–554. [DOI] [PubMed] [Google Scholar]
- 239.Mannino RG, et al. , Smartphone app for non-invasive detection of anemia using only patient-sourced photos. Nature Communications, 2018. 9(1): p. 4924. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 240.Paulus J, et al. , Automated quality assessment of retinal fundus photos. International Journal of Computer Assisted Radiology and Surgery, 2010. 5(6): p. 557–564. [DOI] [PubMed] [Google Scholar]
- 241.Pinto C, Parab J, and Naik G, Non-invasive hemoglobin measurement using embedded platform. Sensing and Bio-Sensing Research, 2020. 29: p. 100370. [Google Scholar]
- 242.Tyburski EA, et al. , Disposable platform provides visual and color-based point-of-care anemia self-testing. The Journal of Clinical Investigation, 2014. 124(10): p. 4387–4394. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 243.Timm U, et al. , Novel multi wavelength sensor concept to detect total hemoglobin concentration, methemoglobin and oxygen saturation. Proceedings of SPIE, 2015. 9332: p. 93320J–93320J-9. [Google Scholar]
- 244.McGann PT, et al. , Cost and Reliability of Two Methods of Hemoglobin Identification for Sickle Cell Newborn Screening in the Republic of Angola. Blood, 2012. 120(21): p. 2.22767573 [Google Scholar]
- 245.McGann PT, et al. , A Prospective Pilot Newborn Screening and Treatment Program for Sickle Cell Anemia in the Republic of Angola. Blood, 2012. 120(21): p. 2.22767573 [Google Scholar]
- 246.Alqershi KA, Kadhim AM, and Al-Muaala AMN, Clinical evaluation of a newly developed point-of-care haemoglobinometer. Comparison with colorimetric and cell counter haemoglobin determination methods. New Zealand Journal of Medical Laboratory Science, 2020. 74(1): p. 31–34. [Google Scholar]