. 2022 Feb 8;11(2):223. doi: 10.3390/pathogens11020223

Table 8.

Interlaboratory comparison of the MICs measured when testing E. coli ATCC® 10536 with octenidine.

MIC
(in %)
Lab 1 Lab 2 Lab 3 Lab 4 Lab 5 Lab 6 Lab 7 Lab 8 Lab 9 Lab 10 Lab 11
≥0.032
0.016
0.008
0.004
0.002 1
0.001 1 2 3
0.0005 2 2 5 12 1 2 2 9
0.00025 20 15 18 8 15 2 10 2 15 13
0.000125 8 13 6 8 14 16 20 15 26 12 3
0.00006 12 10 3 2 1 1
0.00003
≤0.000015

The white area represents the proposed QC range. The gray areas represent the parts of the test range outside the QC range. Values outside this QC range are displayed in red and italics.