Skip to main content
. 2022 Feb 9;27(4):1164. doi: 10.3390/molecules27041164

Table 3.

Evaluation of precision of thymol for the greener normal-phase HPTLC method a.

Conc.
(ng/band)
Intraday Precision Interday Precision
Conc.
(ng/band) ± SD
Standard Error CV (%) Conc. (ng/band) ± SD Standard Error CV (%)
50 51.23 ± 0.48 0.19 0.93 50.68 ± 0.50 0.20 0.98
500 504.64 ± 3.25 1.32 0.64 493.65 ± 3.08 1.25 0.62
2000 1987.24 ± 6.84 2.79 0.34 2024.21 ± 7.38 3.01 0.36

a Mean ± SD; n = 6.