Table II.

Mean measures for similarity of stress distribution (%), percentage of interface area underloaded compared to the native case (%), maximum stress (MPa) in medial unicompartmental knee arthroplasty, and the total load measured at the medial bone-implant interface (N). M0 refers to the conventional implant; M1 to M4 refers to the additively manufactured implants as given in Table I.

Test variant Mean Jaccard similarity, % (SD) Mean area underloaded, % (SD) Mean maximum stress, MPa (SD) Mean total load measured, N (SD)
Native N/A N/A 2.0 (0.7) 422.9 (1.1)
M0 19.8 (0.6) 71.1 (1.2) 0.2 (0.0) 386.8 (1.3)
M1 76.9 (1.8) 1.2 (1.0) 1.7 (0.6) 362.6 (1.8)
M2 74.3 (1.1) 3.6 (2.1) 1.7 (0.8) 374.4 (2.2)
M3 71.2 (2.3) 5.0 (1.7) 2.4 (0.9) 423.5 (2.6)
M4 75.4 (3.0) 2.7 (2.2) 1.8 (0.8) 420.0 (2.3)

N/A, not applicable; SD, standard deviation.