. 2022 Feb 16;11(2):91–101. doi: 10.1302/2046-3758.112.BJR-2021-0304.R1

Table III.

Mean measures for similarity of stress distribution (%), percentage of interface area underloaded compared to the native case (%), maximum stress (MPa) in total knee arthroplasty, and the total load measured across the whole bone-implant interface (N). T0 refers to the conventional implant; T1 to T4 refers to the additively manufactured implants as given in Table I.

Test variant Mean Jaccard similarity, % (SD) Mean area underloaded, % (SD) Mean maximum stress, MPa (SD) Mean total load measured, N (SD)
Medial condyle Lateral condyle
Native N/A N/A 2.1 (0.1) 1.7 (0.2) 698.4 (1.4)
T0 20.5 (2.4) 77.0 (1.2) 0.2 (0.0) 0.2 (0.0) 698.8 (0.8)
T1 77.6 (5.0) 0.9 (0.8) 2.7 (0.7) 1.9 (0.2) 699.1 (0.6)
T2 67.8 (12.1) 0.5 (0.4) 1.8 (0.3) 1.4 (0.2) 699.2 (0.2)
T3 74.2 (5.8) 1.7 (0.5) 2.2 (0.1) 1.7 (0.3) 697.8 (1.3)
T4 72.9 (8.1) 0.5 (0.3) 2.0 (0.2) 1.4 (0.2) 698.5 (0.7)

N/A, not applicable; SD, standard deviation.