. 2021 Dec 21;153(4):309–318.e1. doi: 10.1016/j.adaj.2021.08.011

Table 2.

Average detected particles under different test conditions.

DEVICE
ACTIVITY
AVERAGE OF PARTICULATE MATTER, μm
AVERAGE μg/m3 OF TOTAL SUSPENDED PARTICLES
1 2.5 4 10
Microscope and Drape Preoperative 2.72 3.87 5.68 8.75 14.93
Microscope and Drape Aerosol 20.68 35.64 58.28 83.77 106.91
Microscope and Drape Aerosol/HVE 8.18 13.07 20.48 29.49 36.66
Microscope and Drape Aerosol/HVE/extraoral suction device 4.03 6.07 9.02 12.55 16.86
FaceShield Preoperative 3.12 4.66 7.13 11.3 19.13
FaceShield Aerosol 61.15 118.23 243.68 558.16 740.66
FaceShield Aerosol/HVE 21.95 39.81 74.02 134.57 243.86
Face Shield Aerosol/HVE/extraoral suction device 9.45 15.94 28.41 61.5 139.64

HVE: High-volume evacuator.