Table 3.

Comparison of accuracy (RMSEa) and mean bias of SpO2b estimation between devices during the hypoxia exposure phase. There were 215 SaO2 target windows in this phase.

Performance metrics Philips MX450 CheckMe O2+ WristOx2 3150 AP-20 P value
Available SpO2 points, n 215 207 209 214 N/Ac
RMSE, % (95% CI) 2.67 (2.31-3.06) 3.20 (2.85-3.56) 3.33 (2.85-3.86) 2.86 (2.44-3.25) N/A
Mean bias, % 0.49d –0.22 –1.92e –0.3d <.001
Mean |bias|, % 1.92 2.42 2.40 2.00 <.02
Precision, % 2.62d 3.16d 2.73 2.83 <.02

aRMSE: root-mean-square error.

bSpO2: peripheral oxygen saturation.

cN/A: not applicable.

dDifferent from each other.

eDifferent from other values.