Table 3.
Comparison of accuracy (RMSEa) and mean bias of SpO2b estimation between devices during the hypoxia exposure phase. There were 215 SaO2 target windows in this phase.
Performance metrics | Philips MX450 | CheckMe O2+ | WristOx2 3150 | AP-20 | P value |
Available SpO2 points, n | 215 | 207 | 209 | 214 | N/Ac |
RMSE, % (95% CI) | 2.67 (2.31-3.06) | 3.20 (2.85-3.56) | 3.33 (2.85-3.86) | 2.86 (2.44-3.25) | N/A |
Mean bias, % | 0.49d | –0.22 | –1.92e | –0.3d | <.001 |
Mean |bias|, % | 1.92 | 2.42 | 2.40 | 2.00 | <.02 |
Precision, % | 2.62d | 3.16d | 2.73 | 2.83 | <.02 |
aRMSE: root-mean-square error.
bSpO2: peripheral oxygen saturation.
cN/A: not applicable.
dDifferent from each other.
eDifferent from other values.