. 2022 Feb 25;16:489–505. doi: 10.2147/OPTH.S342666

Table 1.

Characteristics of the Three Patient Cohorts for Each of Three Autorefractors; Adaptica 2WIN, Plusoptix A12 and Welch Allyn SPOT. Q1 and Q3 are Interquartile Ranges. K-W is Kruskall-Wallis Expressed as Chi-Squared with 2 Degrees of Freedom. S.D. Is Standard Deviation the Devices Compared with Analysis of Variance (ANOVA) Represented by F Statistic. The Ages, Then the Results from Refined Retinoscopy for Cylinder, Sphere and Anisometropia Defined by Either Spherical Equivalent or Meridional Definition. Finally the Visual Acuity-Compared Refractions are Demonstrated by the Alaska Blind Child Discovery (ABCD) Composite Ellipsoid Metric.

Count <4 Years Median Q1 Q3 K-W Mean S.D. ANOVA
Age 7 5 11 X2(2)=86 8 4 F(2,2713)=43
Age 2WIN 1335 331 7 4 10 p<0.001 7 4 p<0.001
PlusoptiX 628 93 7 5 11 8 4
SPOT 755 30 8 6 12 9 4
Exam Cylinder (maximal) X2(2)=33 F(2,2715)=22
2WIN 0.75 0.3 2 p<0.001 1.27 1.34 p<0.001
PlusoptiX 0.75 0.3 2 1.21 1.27
SPOT 1 0.3 2.75 1.65 1.65
Sphere (right eye Spherical Equivalent) X2(2)=0.8 F(2,2715)=2
2WIN 0.5 −0.5 2 p=0.69 0.75 2.57 p=0.11
PlusoptiX 0.5 0 1.13 0.6 1.44
SPOT 0.63 0 1.25 0.57 1.43
Myopia (meridional) X2(2)=3.6 F(2,2715)=3
2WIN 0 −1.3 1 p=0.16 −0.11 2.7 p=0.5
PlusoptiX 0 −1 0.75 −0.15 1.68
SPOT 0 −1.3 0.5 −0.36 1.76
Hyperopia (meridional) X2(2)=6.3 F(2,2715)=4
2WIN 1.05 0 3 p=0.04 1.54 2.64 p=0.02
PlusoptiX 1 0.5 1.75 1.24 1.7
SPOT 1 0.5 2 1.44 1.69
Anisometropia (meridional) X2(2)=11 F(2,2715)=2
2WIN 0.25 0 0.75 p=0.004 0.65 0.98 p=0.11
PlusoptiX 0.25 0 0.75 0.55 0.97
SPOT 0.25 0 0.75 0.63 1.1
Compare ABCD Ellipsoid Composite X2(2)=82 F(2,2730)=22
2WIN 1.92 1.3 2.83 p<0.001 2.19 1.42 p<0.001
PlusoptiX 1.48 0.9 2.32 1.89 1.57
SPOT 1.46 1 2.16 1.78 1.34