Difference of sleep parameters between polysomnography and the portable EEG/EOG recording system.

PSG Portable EEG/EOG recording system Mann–Whitney Effect size Spearman rank correlation
Mean SD Mean SD P Cohen d P Correlation coefficient
TST (min) 381.8 58.9 394.1 103.9 0.753 0.15 0.693 0.167
TIB (min) 644.1 56.8 673 65.6 0.371 0.47 0.568 0.24
SPT (min) 507.2 95.8 508 92.8 1 0.01 0.01 0.833
SOL (min) 89 81.3 94.7 77.2 0.793 0.07 0.001 0.929
WASO (times) 30.7 9.7 25.5 12.3 0.246 0.47 0.003 0.896
WASO (min) 117.8 50.1 136 69.4 0.6 0.3 0.058 0.69
WASO (%) 29.9 10.3 39.8 26.4 0.674 0.49 0.002 0.905
SE (%) 60.6 7.8 55.5 11.7 0.345 0.51 0.015 0.81
REM (%) 6.7 7.7 13.3 10.9 0.093 0.70 <0.001 0.976
N1 (%) 10.4 3.7 8.8 2.9 0.401 0.48 0.086 0.643
N2 (%) 52.7 6.4 55.9 10.5 0.401 0.37 0.12 0.595
N3 (%) 25 6.3 19.5 4.3 0.036 1.02 0.289 0.429
RWA (%) 5 6 1.9 1.8 0.293 0.70 0.435 0.323
RWA/total SREM (%) 37.3 36.1 17.1 13.6 0.431 0.74 0.028 0,762

TST: total sleep time; TIB: time in bed; SPT: sleep period time; WASO: wake time after sleep onset; SOL: sleep onset latency; SE: sleep efficiency; EEG/EOG: electroencephalography/electrooculography, †: P < 0.05.