Abstract
Hydrophobic acrylic intraocular lens (IOL) is the most popular material in cataract surgery. Posterior capsule opacification (PCO) is a long-term complication of cataract surgery. It can impair vision and adversely affect the prognosis of IOL delamination. The objective of this study was to perform a systematic review and meta-analysis to provide an updated evaluation of long-term complications and visual function after implantation with hydrophobic acrylic and silicone intraocular lenses. PubMed, Embase, and Cochrane Library were searched from January 2000 until March 2021. Randomized controlled trials (RCTs) and retrospective studies were finally included. The main outcomes were PCO value and neodymium-doped yttrium aluminum garnet (Nd : YAG) capsulotomy rate. Subgroup analysis was performed to compare hydrophobic acrylic and silicone IOLs during the follow-up period. Sensitivity analysis was also performed. The meta-analysis included a total of 17 studies. When the follow-up period was considered, the results of the analysis revealed higher PCO value (Group 3: standardized mean difference (SMD), −0.59; 95% confidence interval (CI), −0.90 to −0.28) and Nd : YAG capsulotomy rate (Group 3: risk ratio (RR), 0.60; 95% CI, 0.40–0.89) for hydrophobic acrylic IOLs than silicone IOLs during a long-term (≥6 years) follow-up. In conclusion, both the PCO value and the Nd : YAG capsulotomy rates were higher in hydrophobic acrylic IOLs group than the silicone IOLs group at long-term use (more than 6 years) after implantation.
1. Introduction
Cataract surgery is frequently performed worldwide primarily due to aging [1, 2]. Cataract is prevalent in adults aged between 45 and 50 years [3]. Statistical data pertaining to cataracts in the United States have reported a prevalence of nearly 32% among adults below the age of 65 years and 50% among those in their 40s and 50s [4]. Age-related cataract surgery is also being performed earlier than before [5]. As a result, long-term safety and efficacy of intraocular lens (IOL) implantation have been established [6–8].
Materials of IOLs can be distinguished by their moisture content, chemical composition, refractive index, and tensile strength. Differences in these properties can determine complications and vision [9]. Posterior capsule opacification (PCO) value and neodymium-doped yttrium aluminum garnet (Nd : YAG) capsulotomy rate are typical indicators of the incidence of complications after cataract surgeries [10]. In particular, PCO is a representative long-term complication following cataract surgery. It is caused by fibrosis around the posterior capsule [11–13]. This PCO can lead to impaired vision, contrast sensitivity, and glare [11, 13]. PCO can be easily treated via Nd : YAG capsulotomy [10, 12]. However, Nd : YAG capsulotomy can increase the risk of IOL instability, dislocation, or further complications such as increased intraocular pressure, glaucoma, retinal detachment, and cystic macular edema [10, 14, 15].
Hydrophobic acrylic IOLs are widely used because they can reduce complications such as PCO and optimize vision [10, 16]. Theoretically, hydrophobic acrylic IOLs in bioactive materials are known to prevent serious PCO compared to IOLs in polymethyl methacrylate (PMMA) or silicone materials [17–19]. Several studies [18, 19] have reported that hydrophobic acrylic IOLs can yield a lower PCO value than hydrophilic acrylic IOLs. However, clinical studies [7, 8], including long-term follow-up (over six years), have demonstrated that hydrophobic acrylic IOLs are associated with a relatively higher PCO value or Nd : YAG capsulotomy rate than silicone IOLs. In particular, Rønbeck and Kugelberg [7] have reported a higher degree of survival without Nd : YAG capsulotomy in a 12-year follow-up analysis of silicone IOLs compared with hydrophobic acrylic IOLs at more than 6 to 7 years after cataract surgery. Cheng et al. [20] have stated that clinical trials lasting at least five years are needed to further evaluate the impact of IOL materials on PCO reduction and the use of Nd : YAG capsulotomy. Therefore, we conducted a systematic review and meta-analysis to determine whether hydrophobic acrylic IOLs after cataract surgery might be more effective than silicone IOLs in reducing postsurgical complications during a long-term follow-up.
2. Methods
2.1. Literature Search
This review was conducted following the updated Preferred Reporting Items for Systematic reviews and Meta-Analysis (PRISMA) 2020 statement [21] (PRISMA 2020 checklist is detailed in Supplementary File 1). This study was registered with the International Prospective Register of Systematic Reviews (PROSPERO) database (identifier: CRD42021242394). Reports of randomized controlled trials (RCTs) and retrospective studies comparing hydrophobic acrylic IOLs with silicone IOLs in patients with age-related cataracts were identified via a systematic search of PubMed, Embase, and Cochrane Library. The search period was extended from January 2000 to March 2021 to cover long-term follow-up studies. Search terms used a combination of MeSH/Emtree terms and “natural language terminology,” including cataract, intraocular lens, lens implantation, capsule opacification, hydrophobic acrylic, and silicone (search strategy is detailed in Supplementary File 2). In case of duplicate studies with data extracted from the same population group, only the most recent studies were included. Any disagreements regarding the search strategy were resolved via consensus based on discussion.
2.2. Selection Criteria
Studies fulfilling the following selection criteria were included: (1) patients >45 years of age who had age-related cataract and treated with cataract surgery; (2) interventions using hydrophobic acrylic IOLs; (3) comparison with silicone IOLs; (4) outcomes included at least one of the following outcome variables: PCO value, Nd : YAG capsulotomy rate, visual acuity, anterior capsule opacification (ACO) value, tilt, and decentration; (5) RCTs and retrospective studies. Case studies, pilot studies, grey literature, studies published in languages other than English, patients with congenital or traumatic cataracts, and diabetes requiring medical control were excluded from this study. Studies were selected by two reviewers (Y.R. and Y.N.). The first reviewer (Y.R.) reviewed all titles, abstracts, and full texts. The second reviewer (Y.N.) analyzed studies excluded from the review.
2.3. Data Extraction and Quality Assessment
The following data were extracted from each study: author's name, year of publication, study design, number of eyes, patient's age and gender, follow-up period, IOL materials and designs, and individual study outcomes. Primary outcomes were set at the quantitative PCO value represented by score or grade using evaluative software and Nd : YAG capsulotomy rate to compare the degree of postsurgical complications during the long-term follow-up after cataract surgery. Secondary outcomes were ACO value represented by score or grade or area, visual acuity (best-corrected visual acuity, BCVA) represented by the logarithm of the minimum angle of resolution (log MAR), degree of tilt, and decentration in relation to complications immediately following cataract surgery or visual function. PCO values and Nd : YAG capsulotomy rates were determined and categorized according to the follow-up period. In the case of multiple values, all values that could be included in a subgroup were extracted. In other cases, only the most recent values were extracted. Quality assessment of included RCT studies was performed using the Cochrane group's Risk of Bias (ROB) tool [22]. Retrospective studies were assessed using the Risk of Bias In Nonrandomized Studies–of Interventions (ROBINS-I) tool [22]. All controversies were resolved via consensus based on discussion among reviewers.
2.4. Data Analysis
Sensitivity analysis was performed except for studies with missing SD data. According to Cochrane's handbook [22], missing SDs were replaced with the mean value of SD based on values determined using the same evaluative system. PCO and ACO values with various measurement scales as continuous variables were pooled using standardized mean differences (SMDs) with 95% confidence intervals (CIs). Dichotomous variables of Nd : YAG capsulotomy rate were calculated using relative risks (RRs) with 95% CIs. Outcomes of visual acuity, tilt, and decentration were pooled using mean differences (MDs) with 95% CIs. Meta-analysis was considered statistically significant if P-value was less than 0.05. For heterogeneity, I2 values greater than 75% represented high heterogeneity [23] using a random-effects model. Publication bias was visually evaluated via funnel plots. All data analyses for the meta-analysis were performed using RevMan (version 5.4.1, Cochrane Library).
2.5. Subgroup Analysis
Subgroup analysis was performed to confirm results according to the follow-up period. Based on the study of Rønbeck and Kugelberg [7], the following three groups were created according to the length of the follow-up period: (1) Group 1 (G1), short term, 0 years ≤ follow-up period < 3 years; (2) Group 2 (G2), medium term, 3 years ≤ follow-up period < 6 years; (3) Group 3 (G3), long term, follow-up period ≥ 6 years.
3. Results
3.1. Included Studies
A total of 483 articles were identified in the initial analysis. Of them, 122 duplicated articles were excluded. Based on titles and abstracts, 39 potential studies were screened. Finally, 17 eligible studies [6–8, 24–37] were included in this analysis (excluded studies and reasons for exclusion are detailed in Supplementary File 3). The flow diagram of the selection process is presented in Figure 1.
Figure 1.

PRISMA flow diagram outlining study selection.
3.2. Characteristics of the Included Studies
Characteristics of included studies are listed in Table 1. This meta-analysis included 14 RCTs [7, 8, 25, 27–37] and three retrospective studies [6, 24, 26]. The average age of patients ranged from 61.3 to 78 years. The follow-up period varied from one week to 12 years. A subgroup analysis was performed to determine PCO values and Nd : YAG capsulotomy rates. These subgroups were separated by follow-up periods. Based on PCO values, four studies [8, 27, 28, 34] were included in the short-term group (G1), seven studies [8, 24, 26, 27, 31, 36, 37] were included in the medium-term group (G2), and two studies [6, 8] were included in the long-term group (G3). Based on Nd : YAG capsulotomy rates, the short-term group (G1) included four studies [24, 28, 33, 34], the medium-term group (G2) comprised six studies [24, 27, 29, 31, 32, 37], and the long-term group (G3) had three studies [6–8]. These included studies were conducted in the Netherlands, Germany, United States, Austria, Japan, South Korea, Finland, Italy, Sweden, and Lithuania (characteristics of IOLs included in the meta-analysis are detailed in Supplementary File 4).
Table 1.
Characteristics of RCTs or retrospective studies included in the meta-analysis.
| Study | Study design | Country | IOL group | Eyes | Age | Follow-up |
|---|---|---|---|---|---|---|
| Abhilakh Missier et al. [24] | Retrospective study | Netherlands | Hydrophobic acrylic | 107 | 74 ± 14 | 3 years |
| Silicone | 107 | |||||
|
| ||||||
| Baumeister et al. [25] | RCT | Germany | Hydrophobic acrylic | 28 | 74 ± 7 | 1 week |
| Silicone | 28 | 6, 12 months | ||||
|
| ||||||
| Daynes et al. [26] | Retrospective study | USA | Hydrophobic acrylic | 60 | 70 | 3 years |
| Silicone | 51 | 77 | ||||
|
| ||||||
| Findl et al. [27] | RCT | Austria | Hydrophobic acrylic | 53 | 78 ± 4 | 1 year |
| Silicone | 53 | 3 years | ||||
|
| ||||||
| Hayashi et al. [28] | RCT | Japan | Hydrophobic acrylic | 96 | 68.8 ± 10.5 | 1 week |
| Silicone | 83 | 71.0 ± 8.9 | 3, 6, 12, 18 months 2 years | |||
|
| ||||||
| Hayashi et al. [29] | RCT | Japan | Hydrophobic acrylic | 100 | 71.4 ± 6.5 | 3 years |
| Silicone | 100 | |||||
|
| ||||||
| Kim et al. [30] | RCT | Korea | Hydrophobic acrylic | 25 | 63.7 ± 9.2 | 1, 3, 6 months |
| Silicone | 47 | 61.3 ± 10.4 | ||||
|
| ||||||
| Kohnen et al. [31] | RCT | Germany | Hydrophobic acrylic | 60 | 73.9 | 3 years |
| Silicone | 60 | |||||
|
| ||||||
| Ernest et al. [32] | RCT | USA | Hydrophobic acrylic | 83 | 74 | 3 years |
| Silicone | 73 | |||||
|
| ||||||
| Pohjalainen et al. [33] | RCT | Finland | Hydrophobic acrylic | 40 | 67.1 ± 14.1 | 2.4 years |
| Silicone | 40 | 67.2 ± 13.9 | ||||
|
| ||||||
| Prosdocimo et al. [34] | RCT | Italy | Hydrophobic acrylic | 38 | 71 | 18 months |
| Silicone | 40 | |||||
|
| ||||||
| Rønbeck et al. [7] | RCT | Sweden | Hydrophobic acrylic | 62 | 73.1 | 12 years |
| Silicone | 64 | |||||
|
| ||||||
| Sacu et al. [35] | RCT | Austria | Hydrophobic acrylic | 53 | 78 ± 4 | 1 year |
| Silicone | 53 | |||||
|
| ||||||
| Vock et al. [6] | Retrospective study | Austria | Hydrophobic acrylic | 98 | M: 66.4 ± 10.1 | 10 years |
| F: 68.1 ± 10.1 | ||||||
| Silicone | 44 | M: 65.6 ± 7.8 | ||||
| F: 69.8 ± 6.5 | ||||||
|
| ||||||
| Vock, Crnej et al. [8] | RCT | Austria | Hydrophobic acrylic | 53 | 75 ± 9 | 6 years |
| Silicone | 53 | 75 | ||||
|
| ||||||
| Wejde et al. [36] | RCT | Sweden | Hydrophobic acrylic | 59 | 75 | 3 years |
| Silicone | 60 | 73 | ||||
|
| ||||||
| Zemaitiene et al. [37] | RCT | Lithuania | Hydrophobic acrylic | 34 | 67.6 ± 7.7 | 3 years |
| Silicone | 30 | |||||
All included patients had age-related cataract; age is reported in years. F = female; IOL = intraocular lens; M = male; RCT = randomized controlled trial.
3.3. Assessment of Risk of Bias
Figure 2 summarizes the risk of bias in 14 RCTs using the ROB tool. Investigators used an envelope [28, 30], a randomization scheme [25], or a computerized random number generator [7, 8, 27, 29, 31] for random assignment of the 14 RCTs included in the present meta-analysis. Of these 14 RCTs, five [8, 27–29, 35] were double-blind, one study [25] was single-blind, and two studies [31, 37] were impossible to blind. In the case of single-blind or nonblinded trials, the risk of performance bias was deemed high. When random assignment and blinding methods were not specified, they were considered to have an unclear risk. Figure 3 summarizes the risk of bias in three retrospective studies using the ROBINS-I tool. Patients were recruited by follow-up visits [24, 26] or invitations [6] of patients conducted by the same surgeon. The risk of participant selection bias was deemed high when only patients who met the preliminary criteria were recruited by the same surgeon retrospectively [26]. All studies reported the number and reason of dropout patients (bias of each study is detailed in Supplementary Files 5 and 6).
Figure 2.

Risk of bias assessment of RCTs.
Figure 3.

Risk of bias assessment of non-RCTs.
3.4. Comparison of the Degree of Complications Based on Long-Term Follow-Up after Cataract Surgery
3.4.1. PCO Value
PCO values of hydrophobic acrylic and silicone IOLs were comparatively analyzed in 10 studies [6, 8, 24, 26–28, 31, 34, 36, 37] comprising 1,138 eyes. A random-effects model was used due to the high heterogeneity (I2 = 80%) of studies. The overall effect on PCO value showed no statistically significant difference between hydrophobic acrylic and silicone IOLs when the follow-up period was not considered ([SMD], −0.23; 95% CI, −0.50 to 0.05; P=0.11). The forest plot is detailed in Supplementary File 7. Subgroup analysis during the follow-up period revealed a high heterogeneity (I2 = 79%). Therefore, the random-effects model was used. Short-term (G1, 0 years ≤ follow-up period < 3 years) and medium-term (G2, 3 years ≤ follow-up period < 6 years) groups showed no significant difference in PCO value between hydrophobic acrylic and silicone IOLs (G1, [SMD], −0.15; 95% CI, −0.61 to −0.30; P=0.51; G2, [SMD], 0.08; 95% CI, −0.22 to 0.39; P=0.60). However, in the long term (G3, follow-up period ≥6 years), hydrophobic acrylic IOLs were associated with relatively higher PCO values than silicone IOLs, showing a statistically significant difference (G3 [SMD], −0.59; 95% CI, −0.90 to −0.28; P=0.001, Figure 4).
Figure 4.

Comparison of subgroup effects on PCO value between hydrophobic acrylic and silicone IOLs. In the long-term group (G3, follow-up period ≥6 years), hydrophobic acrylic IOLs showed significantly high PCO values than silicone IOLs. Chi2 = chi-square statistic; CI = confidence interval; df = degrees of freedom; I2 = I-squared, heterogeneity statistic; IOL = intraocular lens; IV = inverse variance; PCO = posterior capsule opacification; SMD = standard mean difference; Z = Z-statistic.
3.4.2. Nd : YAG Capsulotomy Rate
The meta-analysis included 12 studies [6–8, 24, 27–29, 31–34, 37] involving 1,541 eyes. The overall effect showed an intermediate degree of heterogeneity (I2 = 70%). Therefore, the fixed-effects model was used. The overall effect without considering the follow-up period showed no statistically significant difference in the Nd : YAG capsulotomy rate between hydrophobic acrylic and silicone IOLs ([RR], 1.21; 95% CI, 0.9–1.56; P=0.14). The forest plot is detailed in Supplementary File 7. Subgroup analysis during the follow-up period revealed an intermediate degree of heterogeneity (I2 = 74%). Thus, a fixed-effects model was applied. Short-term (G1, 0 years ≤ follow-up period < 3 years) and medium-term (G2, 3 years ≤ follow-up period < 6 years) groups with hydrophobic acrylic IOLs showed lower Nd : YAG capsulotomy rates than those with silicone IOLs (G1, [RR], 3.08; 95% CI, 1.57–6.07; P=0.001, G2, [RR], 2.12; 95% CI, 1.45–3.12; P < 0.001). However, in the long-term group (G3, follow-up period ≥6 years), hydrophobic acrylic IOLs resulted in higher Nd : YAG capsulotomy rates than silicone IOLs, showing a statistically significant difference between these two IOLs (G3 [RR], 0.60; 95% CI, 0.40–0.89; P=0.01, Figure 5).
Figure 5.

Comparison of subgroup effects on Nd : YAG capsulotomy rates between hydrophobic acrylic and silicone IOLs. In the long-term group (G3, follow-up period ≥6 years), hydrophobic acrylic IOLs were associated with significantly higher Nd : YAG capsulotomy rates than silicone IOLs. Chi2 = chi-square statistic; CI = confidence interval; df = degrees of freedom; I2 = I-squared, heterogeneity statistic; IOL = intraocular lens; M-H = Mantel-Haenszel estimate; RR = risk ratio; Z = Z-statistic.
3.5. Complications Immediately following Cataract Surgery and Visual Function
3.5.1. ACO Value
The meta-analysis included four studies [26, 27, 35, 37] with 384 eyes to determine the ACO value. The overall effect showed a high heterogeneity (I2 = 81%). Therefore, the random-effects model was applied. In the forest plot, ACO values of hydrophobic acrylic IOLs were relatively lower than those of silicone IOLs, showing no statistically significant difference ([SMD], 0.34; 95% CI, −0.14 to 0.83; P=0.17, Figure 6).
Figure 6.

Comparison of ACO values between hydrophobic acrylic and silicone IOLs, showing no statistically significant differences between the two IOLs. Chi2 = chi-square statistic; CI = confidence interval; df = degrees of freedom, I2 = I-squared, heterogeneity statistic; IOL = intraocular lens; IV = inverse variance; SMD = standard mean difference; Z = Z-statistic.
3.5.2. Visual Acuity (BCVA)
The meta-analysis included five studies [6, 26, 30, 31, 37] with 481 eyes to determine visual acuity. No statistically significant heterogeneity (I2 = 0%) was observed between included studies. Therefore, the fixed-effects model was used. The overall effect showed no statistically significant difference in visual acuity between hydrophobic acrylic and silicone IOLs ([MD], −0.00; 95% CI, −0.02 to 0.01; P=0.92, Figure 7).
Figure 7.

Comparison of visual acuity (BCVA) between hydrophobic acrylic and silicone IOLs, showing no statistically significant differences between the two IOLs. BCVA = best-corrected visual acuity; Chi2 = chi-square statistic; CI = confidence interval; df = degrees of freedom; I2 = I-squared, heterogeneity statistic; IOL = intraocular lens; IV = inverse variance; MD = mean difference; Z = Z-statistic.
3.5.3. Tilt and Decentration
Two studies [25, 30] with 128 eyes were included to analyze tilt and decentration, respectively. Both outcomes showed no statistically significant heterogeneity between studies (tilt, I2 = 0%; decentration, I2 = 0%). Thus, fixed-effects models were applied. Overall effects showed no statistically significant differences between hydrophobic acrylic and silicone IOLs (tilt, [MD], −0.06; 95% CI, −0.43 to 0.31; P=0.75, Figure 8; decentration [MD], 0.02; 95% CI, −0.04 to 0.08; P=0.50, Figure 9).
Figure 8.

Comparison of tilt between hydrophobic acrylic and silicone IOLs, showing no statistically significant differences between the two IOLs. Chi2 = chi-square statistic; CI = confidence interval; df = degrees of freedom; I2 = I-squared, heterogeneity statistic; IOL = intraocular lens; IV = inverse variance; MD = mean difference; Z = Z-statistic.
Figure 9.

Comparison of decentration between hydrophobic acrylic and silicone IOLs, showing no statistically significant differences between the two IOLs. Chi2 = chi-square statistic; CI = confidence interval; df = degrees of freedom; I2 = I-squared, heterogeneity statistic; IOL = intraocular lens; IV = inverse variance; MD = mean difference; Z = Z-statistic.
3.6. Sensitivity Analysis
Sensitivity analysis was performed except for three [27, 31, 36] that did not report SDs. Analysis revealed no significant change in overall results (Supplementary File 8).
3.7. Publication Bias
Publication bias was evaluated by visually examining the funnel plot. The funnel plot showed asymmetry in Nd : YAG capsulotomy rate, suggesting some degree of publication bias (Supplementary File 9).
4. Discussion
In the present study, systematic review and meta-analysis were conducted to evaluate complications during long-term follow-up and visual function of hydrophobic acrylic IOLs compared with silicone IOLs. The contribution of this study can be summarized as follows. We evaluated the effects of complications and visual function, including long-term clinical studies with follow-up of more than six years after implantation of hydrophobic acrylic or silicone IOLs. We also found that compared with silicone IOLs, hydrophobic acrylic IOLs were better in terms of the degree of PCO [17–19]. However, hydrophobic acrylic IOLs were associated with higher PCO values and Nd : YAG capsulotomy rates over a 6-year follow-up. Therefore, a long-term (≥6 years) use of hydrophobic acrylic IOLs could affect PCO and Nd : YAG capsulotomy more than such use of silicone IOLs.
Subgroup analysis during the follow-up period revealed higher PCO value and Nd : YAG capsulotomy rates in the group carrying long-term (≥6 years) hydrophobic acrylic IOLs compared with those bearing silicone IOLs. This finding was inconsistent with previous studies [17–19] reporting a lower PCO value with hydrophobic acrylic IOLs than that with silicone IOLs. The barrier effect on PCO is generated by a stably formed capsule bending inhibiting the movement of lens epithelial cells (LECs) to the posterior capsule, which is primarily superior in sharp-edge IOLs [38, 39]. However, if the continuous proliferation of LECs is delayed over a specific duration, a Soemmering's Ring is formed, which abrades the barrier effect of the sharp edge [39]. The hydrophobic acrylic IOLs in this study all had sharp edges. In contrast, silicone IOLs partially exhibited round edges. Nonetheless, compared with silicone IOLs, hydrophobic acrylic IOLs exhibited higher PCO values and Nd : YAG capsulotomy rates, implying that the barrier effect of sharp-edge hydrophobic acrylic IOL was lost due to a long-term (≥6 years) use. Thus, from a long-term perspective, it can be interpreted that the properties of the material itself had a greater impact on the PCO than the effects of the edge design. Compared with hydrophobic acrylic, silicone can mediate the adhesion between IOL and capsule by combining collagen IV and vitronectin attachment proteins [40]. Silicone can also resist the formation of Soemmering's Ring [39]. Therefore, it could help prevent PCO longer during the long-term use than hydrophobic acrylic.
The PCO value did not vary significantly between hydrophobic acrylic and silicone IOLs in the short-term (G1) or the medium-term (G2) follow-up. However, the incidence of Nd : YAG capsulotomy rate was lower in the case of hydrophobic acrylic IOLs during short-term (G1) and medium-term (G2) follow-ups. Although Nd : YAG capsulotomy is the only treatment for PCO, the PCO value and the Nd : YAG capsulotomy rate did not show consistency, which was contrary to other studies [7, 41]. This result might be attributed to differences in reaching Nd : YAG capsulotomy diagnosis depending on the degree of PCO. According to Ling et al. [41], the diagnosis of PCO prior to performing Nd : YAG capsulotomy is not always established. It may vary depending on the assessment. Clinical studies analyzed in this meta-analysis used a variety of evaluation systems, including subjective methods for evaluating PCO levels. Unfortunately, no standardized method is currently available to evaluate the PCO value before Nd : YAG capsulotomy in clinical practice [41]. Therefore, advanced methods of PCO standardization and clinical trials with subsequent Nd : YAG capsulotomy are needed.
There was no significant difference in ACO value between hydrophobic acrylic and silicone IOLs. Hydrophobic acrylic IOLs had relatively lower ACO values than silicone IOLs ([SMD] = 0.34), indicating an intermediate effect size (0.2 ≤ [SMD] < 0.5) [42]. This might be due to a more pronounced effect of similar properties between haptic materials of the two IOLs on ACO than optic materials of the IOLs. Silicone IOLs included in our ACO analysis all had three pieces made of polyvinylidene fluoride (PVDF) or PMMA haptics. Loop memory in PVDF has properties similar to PMMA haptic of hydrophobic acrylic IOLs [43]. However, a high heterogeneity (I2 = 81%) between ACO studies included in the meta-analysis was found. This interpretation is marginal due to the small number of studies. No further analysis of heterogeneity has been made. However, the high heterogeneity might be attributed to a combination of factors and scales that affect ACO.
Comparing the effect size of hydrophobic acrylic and silicone IOLs in terms of visual function after cataract surgery, visual acuity was statistically similar between the two groups. Previous meta-analyses [11, 44, 45] comparing typical IOL materials (PMMA, silicone, and acrylic) have revealed no significant differences in visual acuity. Our study results are consistent with these prior studies, suggesting the absence of a significant effect on the visual acuity of these two IOL materials. There were no significant differences in tilt or decentration between hydrophobic acrylic and silicone IOL materials either. Forward and backward movement of IOL due to tilt and decentration can affect refraction and aberration of eyes [25]. This effect depends on the spherical degree of IOLs, which has recently been complemented by the emergence of aspherical IOLs [46]. All hydrophobic acrylic and silicone IOLs included in this study were spherical, suggesting no difference in optical performance [25, 47].
This meta-analysis has some limitations. Since most clinical studies related to IOL materials mainly reported results of PCO and Nd : YAG capsulotomy, a subgroup analysis was feasible only for PCO value and Nd : YAG capsulotomy rate during the follow-up period. Therefore, an adequate number of clinical trials related to visual function and complications other than PCO are needed in the future. Furthermore, results of ACO suggested a high heterogeneity (I2 = 81%). No further analysis has been made to decrease the heterogeneity. Another subgroup or sensitivity analysis, such as an additional analysis based on edge design, haptic material, optical size, presence of aspheric lens, and surgical technique used [48, 49], will be necessary in the future. Nd : YAG capsulotomy irradiated with low energy laser affects the morphology of IOL. Thus, further meta-analysis studies should be done to determine damage and structure changes of IOL after being hit by the laser [50].
5. Conclusion
Hydrophobic acrylic IOLs are associated with higher PCO values and Nd : YAG capsulotomy rates than silicone IOLs when they are used for a long term (more than 6 years). However, both hydrophobic acrylic and silicone IOLs can lead to similar visual functions.
Acknowledgments
This research was supported by the Training Program for Advanced Medical Device Industry Professional Manpower of Korea Health Industry Development Institute (KHIDI) funded by the Ministry of Health and Welfare, Republic of Korea.
Data Availability
All data are included within this article and its supplementary files.
Disclosure
The organization had no role in the design or conduct of this research.
Conflicts of Interest
The authors declare that there are no conflicts of interest regarding the publication of this article.
Supplementary Materials
Supplementary File 1. PRISMA 2020 checklist. Supplementary File 2. Details of the search strategy. Supplementary File 3. Details of excluded studies and reasons. Supplementary File 4. Characteristics of IOLs included studies. Supplementary File 5. Risk of bias assessment of randomized controlled trials. Supplementary File 6. Risk of bias assessment of nonrandomized controlled trials. Supplementary File 7. Forest plots of the overall effect of PCO value and Nd : YAG capsulotomy rate. Supplementary File 8. Sensitivity analysis except the studies missing SDs. Supplementary File 9. Funnel plots of publication bias.
References
- 1.Asbell P., Dualan I., Mindel J., Brocks D., Ahmad M., Epstein S. Age-related cataract. The Lancet . 2005;365(9459):599–609. doi: 10.1016/s0140-6736(05)70803-5. [DOI] [PubMed] [Google Scholar]
- 2.Chang J. R., Koo E., Agrón E., et al. Risk factors associated with incident cataracts and cataract surgery in the age-related eye disease study (AREDS) Ophthalmology . 2011;118(11):2113–2119. doi: 10.1016/j.ophtha.2011.03.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Liu Y. C., Wilkins M., Kim T., Malyugin B., Mehta J. S. Cataracts. The Lancet . 2017;390(10094):600–612. doi: 10.1016/s0140-6736(17)30544-5. [DOI] [PubMed] [Google Scholar]
- 4.National Eye Institute. Cataract data and statistics. 2021. https://www.nei.nih.gov/learn-about-eye-health/resources-for-health-educators/eye-health-data-and-statistics/cataract-data-and-statistics/cataract-tables .
- 5.Kanclerz P., Yildirim T. M., Khoramnia R. A review of late intraocular lens opacifications. Current Opinion in Ophthalmology . 2021;32(1):31–44. doi: 10.1097/icu.0000000000000719. [DOI] [PubMed] [Google Scholar]
- 6.Vock L., Menapace R., Stifter E., Georgopoulos M., Sacu S., Bühl W. Posterior capsule opacification and neodymium: YAG laser capsulotomy rates with a round-edged silicone and a sharp-edged hydrophobic acrylic intraocular lens 10 years after surgery. Journal of Cataract and Refractive Surgery . 2009;35(3):459–465. doi: 10.1016/j.jcrs.2008.11.044. [DOI] [PubMed] [Google Scholar]
- 7.Rønbeck M., Kugelberg M. Posterior capsule opacification with 3 intraocular lenses: 12-year prospective study. Journal of Cataract and Refractive Surgery . 2014;40:70–76. doi: 10.1016/j.jcrs.2013.07.039. [DOI] [PubMed] [Google Scholar]
- 8.Vock L., Crnej A., Findl O., et al. Posterior capsule opacification in silicone and hydrophobic acrylic intraocular lenses with sharp-edge optics six years after surgery. American Journal of Ophthalmology . 2009;147(4):683–690. doi: 10.1016/j.ajo.2008.11.006. [DOI] [PubMed] [Google Scholar]
- 9.Nanu R. V., Ungureanu E., Vrapciu S. L., et al. An overview of the influence and design of biomaterial of the intraocular implant of the posterior capsule opacification. Romanian Journal of Ophthalmology . 2018;62(3):188–193. doi: 10.22336/rjo.2018.29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Thom H., Ender F., Samavedam S., et al. Effect of AcrySof versus other intraocular lens properties on the risk of Nd: YAG capsulotomy after cataract surgery: a systematic literature review and network meta-analysis. PLoS One . 2019;14 doi: 10.1371/journal.pone.0220498.e0220498 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Findl O., Buehl W., Bauer P., Sycha T. Interventions for preventing posterior capsule opacification. Cochrane Database of Systemic Reviwes . 2010;17 doi: 10.1002/14651858.CD003738.pub3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Buehl W., Findl O. Effect of intraocular lens design on posterior capsule opacification. Journal of Cataract and Refractive Surgery . 2008;34(11):1976–1985. doi: 10.1016/j.jcrs.2008.07.029. [DOI] [PubMed] [Google Scholar]
- 13.Wormstone I. M., Wormstone Y. M., Smith A., Eldred J. A. Posterior capsule opacification: what’s in the bag? Progress in Retinal and Eye Research . 2021;82 doi: 10.1016/j.preteyeres.2020.100905. [DOI] [PubMed] [Google Scholar]
- 14.Kim B. J., Kim J. H., Kim S. J., Han Y. S., Park J. M., Chung I. Y. Two cases of WIOL-CF IOL dislocation after Nd: YAG laser capsulotomy. Journal of the Korean Ophthalmological Society . 2015;56(3):443–446. doi: 10.3341/jkos.2015.56.3.443. [DOI] [Google Scholar]
- 15.Billotte C., Berdeaux G. Adverse clinical consequences of neodymium: YAG laser treatment of posterior capsule opacification. Journal of Cataract and Refractive Surgery . 2004;30(10):2064–2071. doi: 10.1016/j.jcrs.2004.05.003. [DOI] [PubMed] [Google Scholar]
- 16.Lundström M., Barry P., Henry Y., Rosen P., Stenevi U. Evidence-based guidelines for cataract surgery: guidelines based on data in the European Registry of Quality Outcomes for Cataract and Refractive Surgery database. Journal of Cataract and Refractive Surgery . 2012;38:1086–1093. doi: 10.1016/j.jcrs.2012.03.006. [DOI] [PubMed] [Google Scholar]
- 17.Vasavada A. R., Raj S. M., Shah G. D., Nanavaty M. A. Posterior capsule opacification after lens implantation: incidence, risk factors and management. Expert Review of Ophthalmology . 2013;8(2):141–149. doi: 10.1586/eop.12.80. [DOI] [Google Scholar]
- 18.Alió J. L., Chipont E., BenEzra D., Fakhry M. A. International ocular inflammation society, study group of uveitic cataract surgery, “comparative performance of intraocular lenses in eyes with cataract and uveitis. Journal of Cataract & Refractive Surgery . 2002;28(12):2096–2108. doi: 10.1016/s0886-3350(02)01452-9. [DOI] [PubMed] [Google Scholar]
- 19.Vasavada A. R., Raj S. M., Shah A., Shah G., Vasavada V., Vasavada V. Comparison of posterior capsule opacification with hydrophobic acrylic and hydrophilic acrylic intraocular lenses. Journal of Cataract and Refractive Surgery . 2011;37(6):1050–1059. doi: 10.1016/j.jcrs.2010.12.060. [DOI] [PubMed] [Google Scholar]
- 20.Cheng J. W., Wei R. L., Cai J. P., et al. Efficacy of different intraocular lens materials and optic edge designs in preventing posterior capsular opacification: a meta-analysis. American Journal of Ophthalmology . 2007;143(3):428–436. doi: 10.1016/j.ajo.2006.11.045. [DOI] [PubMed] [Google Scholar]
- 21.Page M. J., Moher D., Bossuyt P. M. PRISMA 2020 explanation and elaboration: updated guidance and exemplars for reporting systematic reviews. BMJ . 2021;372 doi: 10.1136/bmj.n160. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.The Cochrane Collaboration. Cochrane Handbook for Systematic Reviews of Interventions. Version 5.1.0 . 2021. https://handbook-5-1.cochrane.org/chapter_16/16_1_3_missing_standard_deviations.htm . [Google Scholar]
- 23.Huedo Medina T. B., Sánchez Meca J., Marín Martínez F., Botella J. Assessing heterogeneity in meta-analysis: Q statistic or I2 index? Psychological Methods . 2006;11(11):193–206. doi: 10.1037/1082-989x.11.2.193. [DOI] [PubMed] [Google Scholar]
- 24.Abhilakh Missier K. A., Nuijts R. M. M. A., Tjia K. F. Posterior capsule opacification: silicone plate-haptic versus AcrySof intraocular lenses. Journal of Cataract and Refractive Surgery . 2003;29(8):1569–1574. doi: 10.1016/s0886-3350(02)02046-1. [DOI] [PubMed] [Google Scholar]
- 25.Baumeister M., Neidhardt B., Strobel J., Kohnen T. Tilt and decentration of three-piece foldable high-refractive silicone and hydrophobic acrylic intraocular lenses with 6-mm optics in an intraindividual comparison. American Journal of Ophthalmology . 2005;140(6):1051–1058. doi: 10.1016/j.ajo.2005.07.026. [DOI] [PubMed] [Google Scholar]
- 26.Daynes T., Spencer T. S., Doan K., Mamalis N., Olson R. J. Three-year clinical comparison of 3-piece AcrySof and SI-40 silicone intraocular lenses. Journal of Cataract and Refractive Surgery . 2002;28(7):1124–1129. doi: 10.1016/s0886-3350(02)01460-8. [DOI] [PubMed] [Google Scholar]
- 27.Findl O., Menapace R., Sacu S., Buehl W., Rainer G. Effect of optic material on posterior capsule opacification in intraocular lenses with sharp-edge optics randomized clinical trial. Ophthalmology . 2005;112(1):67–72. doi: 10.1016/j.ophtha.2004.07.032. [DOI] [PubMed] [Google Scholar]
- 28.Hayashi K., Hayashi H., Nakao F., Hayashi F. Changes in posterior capsule opacification after poly(methyl methacrylate), silicone, and acrylic intraocular lens implantation. Journal of Cataract and Refractive Surgery . 2001;27(6):817–824. doi: 10.1016/s0886-3350(01)00787-8. [DOI] [PubMed] [Google Scholar]
- 29.Hayashi K., Hayashi H. Influence on posterior capsule opacification and visual function of intraocular lens optic material. American Journal of Ophthalmology . 2007;144(2):195–202. doi: 10.1016/j.ajo.2007.04.044. [DOI] [PubMed] [Google Scholar]
- 30.Kim J. S., Shyn K. H. Biometry of 3 types of intraocular lenses using Scheimpflug photography. Journal of Cataract and Refractive Surgery . 2001;27(4):533–536. doi: 10.1016/s0886-3350(00)00649-0. [DOI] [PubMed] [Google Scholar]
- 31.Kohnen T., Fabian E., Gerl R., et al. Optic edge design as long-term factor for posterior capsular opacification rates. Ophthalmology . 2008;115(8):1308–1314. doi: 10.1016/j.ophtha.2008.01.002. [DOI] [PubMed] [Google Scholar]
- 32.Ernest P. H. Posterior capsule opacification and neodymium: YAG capsulotomy rates with AcrySof acrylic and PhacoFlex II silicone intraocular lenses. Journal of Cataract and Refractive Surgery . 2003;29(8):1546–1550. doi: 10.1016/s0886-3350(02)01974-0. [DOI] [PubMed] [Google Scholar]
- 33.Pohjalainen T., Vesti E., Uusitalo R. J., Laatikainen L. Posterior capsular opacification in pseudophakic eyes with a silicone or acrylic intraocular lens. European Journal of Ophthalmology . 2002;12(3):212–218. doi: 10.1177/112067210201200307. [DOI] [PubMed] [Google Scholar]
- 34.Prosdocimo G., Tassinari G., Sala M., et al. Posterior capsule opacification after phacoemulsification: silicone CeeOn edge versus acrylate AcrySof intraocular lens. Journal of Cataract and Refractive Surgery . 2003;29(8):1551–1555. doi: 10.1016/s0886-3350(02)02051-5. [DOI] [PubMed] [Google Scholar]
- 35.Sacu S., Menapace R., Findl O. Effect of optic material and haptic design on anterior capsule opacification and capsulorrhexis contraction. American Journal of Ophthalmology . 2006;141(3):488–493. doi: 10.1016/j.ajo.2005.10.041. [DOI] [PubMed] [Google Scholar]
- 36.Wejde G., Kugelberg M., Zetterström C. Position of anterior capsulorhexis and posterior capsule opacification. Acta Ophthalmologica Scandinavica . 2004;82(5):531–534. doi: 10.1111/j.1600-0420.2004.00322.x. [DOI] [PubMed] [Google Scholar]
- 37.Zemaitienė R., Jašinskas V. Prevention of posterior capsule opacification with 3 intraocular lens models: a prospective, randomized, long-term clinical trial. Medicina . 2011;47:595–599. [PubMed] [Google Scholar]
- 38.Nishi O., Nishi K., Akura J. Speed of capsular bend formation at the optic edge of acrylic, silicone, and poly(methyl methacrylate) lenses. Journal of Cataract and Refractive Surgery . 2002;28(3):431–437. doi: 10.1016/s0886-3350(01)01094-x. [DOI] [PubMed] [Google Scholar]
- 39.Nishi O., Yamamoto N., Nishi K., Nishi Y. Contact inhibition of migrating lens epithelial cells at the capsular bend created by a sharp-edged intraocular lens after cataract surgery. Journal of Cataract and Refractive Surgery . 2007;33(6):1065–1070. doi: 10.1016/j.jcrs.2007.02.022. [DOI] [PubMed] [Google Scholar]
- 40.Linnola R. J., Sund M., Ylönen R., Pihlajaniemi T. Adhesion of soluble fibronectin, vitronectin, and collagen type IV to intraocular lens materials. Journal of Cataract and Refractive Surgery . 2003;29(1):146–152. doi: 10.1016/s0886-3350(02)01422-0. [DOI] [PubMed] [Google Scholar]
- 41.Ling R., Borkenstein E. M., Borkenstein A. F. Evaluation of Nd: YAG laser capsulotomy rates in a real-life population. Clinical Ophthalmology . 2020;14:3249–3257. doi: 10.2147/opth.s276329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Takeshima N., Sozu T., Tajika A., Ogawa Y., Hayasaka Y., Furukawa T. A. Which is more generalizable, powerful and interpretable in meta-analyses, mean difference or standardized mean difference? BMC Medical Research Methodology . 2014;14 doi: 10.1186/1471-2288-14-30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Izak A. M., Werner L., Apple D. J., Macky T. A., Trivedi R. H., Pandey S. K. Loop memory of haptic materials in posterior chamber intraocular lenses. Journal of Cataract and Refractive Surgery . 2002;28(7):1229–1235. doi: 10.1016/s0886-3350(02)01326-3. [DOI] [PubMed] [Google Scholar]
- 44.Li N., Chen X., Zhang J., et al. Effect of AcrySof versus silicone or polymethyl methacrylate intraocular lens on posterior capsule opacification. Ophthalmology . 2008;115(5):830–838. doi: 10.1016/j.ophtha.2007.06.037. [DOI] [PubMed] [Google Scholar]
- 45.Zhao Y., Yang K., Li J., Huang Y., Zhu S. Comparison of hydrophobic and hydrophilic intraocular lens in preventing posterior capsule opacification after cataract surgery: an updated meta-analysis. Medicine (Baltimore) . 2017;96 doi: 10.1097/MD.0000000000008301.e8301 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Schuster A. K., Tesarz J., Vossmerbaeumer U. The impact on vision of aspheric to spherical monofocal intraocular lenses in cataract surgery. Ophthalmology . 2013;120(11):2166–2175. doi: 10.1016/j.ophtha.2013.04.011. [DOI] [PubMed] [Google Scholar]
- 47.Raj S. M., Vasavada A. R., Kaid J. S., Vasavada V. A., Vasavada V. A. Post-operative capsular opacification. Nepalese Journal of Ophthalmology . 2009;1:43–59. doi: 10.3126/nepjoph.v1i1.3673. [DOI] [PubMed] [Google Scholar]
- 48.Werner L., Pandey S. K., Escobar-Gomez M., Visessook N., Peng Q., Apple D. J. Anterior capsule opacification 11 the authors have no financial or proprietary interest in any product mentioned in this paper. Ophthalmology . 2000;107(3):463–471. doi: 10.1016/s0161-6420(99)00088-3. [DOI] [PubMed] [Google Scholar]
- 49.Pérez-Vives C. Biomaterial influence on intraocular lens performance: an overview. Journal of Ophthalmology . 2018;2018:17. doi: 10.1155/2018/2687385.2687385 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Meduri A., Severo AA., De Maria A., et al. PMMA intraocular lenses changes after treatment with Nd: Yag laser: a scanning electron microscopy and X-ray spectrometry study. Applied Sciences . 2020;10 [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplementary File 1. PRISMA 2020 checklist. Supplementary File 2. Details of the search strategy. Supplementary File 3. Details of excluded studies and reasons. Supplementary File 4. Characteristics of IOLs included studies. Supplementary File 5. Risk of bias assessment of randomized controlled trials. Supplementary File 6. Risk of bias assessment of nonrandomized controlled trials. Supplementary File 7. Forest plots of the overall effect of PCO value and Nd : YAG capsulotomy rate. Supplementary File 8. Sensitivity analysis except the studies missing SDs. Supplementary File 9. Funnel plots of publication bias.