Skip to main content
. 2022 Mar 4;12:3602. doi: 10.1038/s41598-022-07696-1

Figure 2.

Figure 2

Bland–Altman plot of K1 (a) and K2 (b) from GALILEI G6 and ANTERION. The solid line represents the mean difference, while the dotted lines on each side show the upper and lower 95% LoA. The K1 and K2 plot demonstrate that only four and five eyes respectively of the total 96 eyes were out of the 95% LoAs, indicating excellent K1 and K2 measurement agreement between the two devices. Created by SPSS (Version 22, SPSS Inc., IBM, Chicago, USA).