Table 2.

Comparison and Intraclass correlation coefficient of various biometric parameters between GALILEI G6 and ANTERION.

Parameter GALILEI G6 ANTERION Mean difference ± SD p value ICC Confident limit
Mean ± SD Mean ± SD lower upper
ACD (mm) 2.93 ± 0.36 2.94 ± 0.38 0.00 ± 0.09 0.630a 0.970 0.954 0.980
AL (mm) 23.30 ± 0.99 23.32 ± 0.99 − 0.03 ± 0.03 < 0.001a 0.999 0.996 1.000
K1 (D) 43.61 ± 1.56 43.34 ± 1.53 0.27 ± 0.46 < 0.001a 0.942 0.859 0.970
K2 (D) 44.65 ± 1.58 44.33 ± 1.56 0.32 ± 0.45 < 0.001a 0.939 0.807 0.972
LT (mm) 4.70 ± 0.36 4.85 ± 0.37 − 0.15 ± 0.15 < 0.001a 0.849 0.335 0.943
WTW (mm) 11.81 ± 0.40 11.66 ± 0.47 0.15 ± 0.21 < 0.001a 0.835 0.570 0.921
IOL power (D) 21.33 ± 2.63 21.55 ± 2.67 − 0.22 ± 0.61 0.001b 0.971 0.951 0.982

Values are presented as mean ± SD.

ap-value from independent t-test .

bp-value from Mann–Whitney U test .