Skip to main content
. 2022 Feb 22;29(Pt 2):581–590. doi: 10.1107/S1600577522000984

Table 1. Summary of the beamline parameters, including source, optical configuration, X-ray beam characteristics and the experimental setup for ID23-2 at the ESRF.

Beamline name ID23-2
Source type 1 × 1.6 m 20.2 mm-period undulator
Horizontal emittance (pm rad) 132
Vertical emittance (pm rad) 5
Monochromator Si(111)
Energy (keV) 14.2
Horizontal photon source size (µm RMS) 30.3
Vertical photon source size (µm RMS) 4.5
Horizontal photon divergence (µrad RMS) 6.9
Vertical photon divergence (µrad RMS) 5.6
 
Vertical focusing
Focusing elements 1D CRLs in two transfocators
Lens material Be
Transfocator Large beam Small beam
Axis 1 (mm) 2 mm alignment pinhole 2 mm alignment pinhole
Axis 2 [No. of lenses × radius (µm)] 1 × 200 1 × 200
Axis 3 (µm) 5 × 200 12 × 200
Axis 4 (µm) 1 × 500 1 × 500
Axis 5 (µm) 1 × 300 1 × 300
Axes 6–7 (µm) Unused Unused
Axis 8 (µm) 2 mm alignment pinhole 2 mm alignment pinhole
Transfocator efficiency (%) 93 88
Vertical demagnification ratio 2.4×
 
Horizontal focusing
Mirror assembly Large beam Small beam
Horizontal focusing mirror (HFM) Bendable tangential mirror Bendable tangential mirror
Incident angle on HFM (mrad) 18 18
Slope error of HFM (µrad RMS) 0.50 0.57
Horizontal demagnification ratio 21× 91×
 
Sample position
Beam size (H × V, FWHM) (µm) 4.7 × 6.0 1.5 × 3.0
Flux (photons s–1) 1.8 × 1013 1.5 × 1013
Diffractometer MD3Up
Sample mounting FLEX HCD with rapid-exchange single and double grippers
Detector type CMOS Hybrid Pixel-Array
Detector model Dectris PILATUS3 X 2M (450 µm Si sensor)