TABLE 4.

Comparison of Coronary Artery and Stent Lumen ROI Data (Surface, HU, and Noise) Between the 2 Systems and Different SPCCT Kernels

EID-DLCT SPCCT XCD vs Sharp Detailed 2 vs Sharp
XCD Detailed 2 Sharp P P
Coronary artery
 Surface, mm2 Obs 1 12.0 (5.4) 10.8 (6.2) 11.0 (1.9) 0.63 0.13
Obs 2 5.5 (1.3) 6.4 (1.8) 6.4 (1.8) 0.15 0.32
 Density, HU Obs 1 510 (194) 439.3 (243.8) 446.2 (236.1) 0.63 0.46
Obs 2 521 (161) 450 (245) 455 (247) 0.75 0.11
 SD Obs 1 65.3 (17.7) 60.6 (17.4) 135.2 (65.5) 0.002 0.001
Obs 2 56.9 (2.3) 49 (6.5) 157 (23) 0.003 0.001
Stent lumen
 Surface, mm2 Obs 1 3.2 (2.1) 3.2 (3.2) 4.3 (2.6) 0.24 0.65
Obs 2 6.3 (4.5) 2.2 (2.2) 2.2 (2.1) 0.002 0.53
 HU, HU Obs 1 629 (449) 519 (321) 512 (311) 0.01 0.04
Obs 2 608 (723) 535 (303) 520 (309) 0.008 0.008
 SD Obs 1 56.6 (26.0) 70.4 (50.7) 122.5 (75.3) 0.002 0.004
Obs 2 88 (62.9) 96.1 (41.8) 98.9 (55.4) 0.24 0.028
Δ S-C*
Obs 1 148 (304) 109 (90) 84 (61) 0.04 0.019
Obs 2 210 (377) 124 (92) 107 (106) 0.023 0.005

Continuous variables are presented as median (interquartile range). Data in bold indicate significant values.

* Medians of the absolute values of the Δ S-Cs.

EID-DLCT indicates energy integrating detectors dual-layer computed tomography; HU, Hounsfield unit; ROI, region of interest; SPCCT, spectral photon counting computed tomography; XCD, Xres cardiac detailed.