Table 7.

RT-qPCR and KU-LAMP testing of 319 clinical samples categorized into the groups of Ct value bins.

Samples RT-qPCR KU-LAMP Mismatch KU-LAMP
Specificity (%)
KU-LAMP
Sensitivity (%)
Allplex Cobas NeoPlex Total
Negative 25 25 48 98 98 0 100
Positive Ct < 25 13 13 15 41 41 0 100
Ct = 25–30 19 15 18 52 51 1 98.92
Ct = 30–35 17 36 34 87 75 12 86.21
Ct > 35 12 6 23 41 36 5 87.80
Total 86 95 138 319 301 18 94.36