TABLE 1.

The full‐width half‐maximum (mm) and penumbra for 4‐mm scans shown in Figure 3; the delta column shows the difference between TMR TPS values and the measured values (Msmt). Tolerance from TG‐178 is ±1 mm of TPS values; action level is ±1.2 mm of TPS values 1

TMR 10 TPS values Film Microdiamond W2 scintillator
Collimation: 4 mm Msmt (mm) (mm) Msmt (mm) (mm) Msmt (mm) (mm)
FWHM x profile 6.16 6.18 (6.0) a 0.06 5.97 (6.2) a −0.19 6.11 −0.05
y profile 6.16 6.22 (5.5) a 0.06 6.30 (6.2) a 0.14 6.17 −0.10
z profile 5.04 5.02 (4.9) a −0.02 4.96 (5.0) a −0.08 5.07 −0.03
Penumbra x profile 2.82 2.87 (2.7) a 0.05 3.04 (2.9) a 0.22 2.97 0.15
y profile 2.82 3.10 (2.9) a 0.28 3.02 (2.9) a 0.20 2.99 0.17
z profile 1.52 1.49 (1.5) a −0.03 1.58 (1.6) a 0.06 1.83 0.31

Abbreviations: FWHM, full‐width, half‐maximum; TPS, treatment planning systems; Msmt, measured values (Msmt).

a

Data shown in parentheses are from Maraghechi et al.15, included for comparison.