TABLE 3.

The full‐width half‐maximum (mm) and penumbra for 16‐mm scans shown in Figure 3; the delta column shows the difference between TMR TPS values and the measured values. Tolerance from TG‐178 is ±1 mm of TPS values; Action level is ±1.2 mm of TPS values 1

Film Microdiamond W2 scintillator
Collimation: 16 mm TMR 10 TPS values Msmt (mm) (mm) Msmt (mm) (mm) Msmt (mm) (mm)
FWHM x profile 21.75 21.4 (21.6) a −0.35 21.73 (11.1) a −0.02 21.71 −0.01
y profile 21.75 21.46 (21.5) a −0.29 21.77 (11.1) a 0.02 21.58 −0.06
z profile 17.44 17.43 (17.3) a −0.01 17.44 (9.7) a 0.00 17.46 −0.02
Penumbra x profile 9.03 8.96 (9) a −0.07 9.00 (4.1) a −0.03 8.91 −0.12
y profile 9.03 9.21 (8.6) a 0.18 9.09 (4.2) a 0.06 9.07 0.04
z profile 2.55 2.62 (2.6) a 0.07 2.52 (2.4) a −0.03 2.76 0.21

Abbreviations: FWHM, full‐width, half‐maximum; TPS, treatment planning systems; Msmt, measured values (Msmt).

a

Data shown in parentheses are from Maraghechi et al.15, included for comparison.