TABLE 4.

Chamber measurements for daily adapt‐to‐shape (ATS) and adapt‐to‐position (ATP) plans on head phantom

ATP workflow ATS workflow
Sup GTV Inf GTV Sup GTV Inf GTV
Adaptive plan (cGy) 2201.0 2263.0 2201.6 2205.6
Measurement (cGy) 2186.5 2255.0 2195.5 2194.0
% Difference 0.7% 0.4% 0.3% 0.5%