TABLE 3.

Comparison of the noMag and wMag methods for small fields

Small field comparison
Location noMag method wMag method
d = 10 cm, CAX 0.9% 0.5%
d = 20 cm, +0.5 cm X‐OAX 2.2% 1.6%
d = 20 cm, ‐0.5 cm X‐OAX 0.3% 0.8%
d = 5 cm, +0.5 cm Y‐OAX 1.4% 1.1%
d = 5 cm, ‐0.5 cm Y‐OAX 0.7% 0.4%

Note: All points are within 2% which is within MPPG5a tolerances except for single OAX noMag comparison.