TABLE 3.
Comparison of the noMag and wMag methods for small fields
Small field comparison | ||
---|---|---|
Location | noMag method | wMag method |
d = 10 cm, CAX | 0.9% | 0.5% |
d = 20 cm, +0.5 cm X‐OAX | 2.2% | 1.6% |
d = 20 cm, ‐0.5 cm X‐OAX | 0.3% | 0.8% |
d = 5 cm, +0.5 cm Y‐OAX | 1.4% | 1.1% |
d = 5 cm, ‐0.5 cm Y‐OAX | 0.7% | 0.4% |
Note: All points are within 2% which is within MPPG5a tolerances except for single OAX noMag comparison.