Skip to main content
. 2022 Mar 10;2022(3):CD013208. doi: 10.1002/14651858.CD013208.pub2

Bwana 2019a.

Study characteristics
Patient Sampling In order to assess the performance of the Xpert HIV‐1 quantitative assay on GeneXpert platform, 100 plasma samples were tested on both the platform and on the Abbott m2000 assay and the results were compared. Both the qualitative and the quantitative studies of the performance of the GeneXpert platform were cross‐sectional evaluations of samples obtained from facilities across the country.
Patient characteristics and setting HIV‐positive adults on antiretroviral therapy (ART); field in Kenya; contacted author who confirmed that samples were from those on ART.
Index tests Xpert HIV‐1 quantitative assay (Cepheid, Sunnyvale, CA, USA); done in peripheral lab on fresh plasma samples from the field.
Target condition and reference standard(s) High viral load > 1000 copies/mL; Abbott m2000.
Flow and timing In order to assess the performance of the Xpert HIV‐1 quantitative assay on GeneXpert platform, 100 plasma samples were tested on both the platform and on the Abbott m2000 assay and the results were compared. For viral load estimation on the GeneXpert platform, a total of 1.2 mL of plasma was added into the Xpert HIV‐1 Viral Load cartridge using a calibrated pipette. The cartridge was closed and loaded onto the machine. Test results were observed and recorded after 90 minutes. For comparison, viral load estimation was done on the Abbott m2000 using plasma according to manufacturer’s instructions. Test results were observed and recorded after 5 hours.
Comparative  
Notes  
Methodological quality
Item Authors' judgement Risk of bias Applicability concerns
DOMAIN 1: Patient Selection
Was a consecutive or random sample of patients enrolled? Unclear    
Was a case‐control design avoided? Unclear    
Did the study avoid inappropriate exclusions? Unclear    
Could the selection of patients have introduced bias?   Unclear risk  
Are there concerns that the included patients and setting do not match the review question?     Low concern
DOMAIN 2: Index Test (All tests)
Were the index test results interpreted without knowledge of the results of the reference standard? Unclear    
If a threshold was used, was it pre‐specified? Yes    
Could the conduct or interpretation of the index test have introduced bias?   Unclear risk  
Are there concerns that the index test, its conduct, or interpretation differ from the review question?     Low concern
DOMAIN 3: Reference Standard
Is the reference standards likely to correctly classify the target condition? Yes    
Were the reference standard results interpreted without knowledge of the results of the index tests? Unclear    
Could the reference standard, its conduct, or its interpretation have introduced bias?   Unclear risk  
Are there concerns that the target condition as defined by the reference standard does not match the question?     Low concern
DOMAIN 4: Flow and Timing
Was there an appropriate interval between index test and reference standard? Unclear    
Did all patients receive the same reference standard? Yes    
Were all patients included in the analysis? Yes    
Could the patient flow have introduced bias?   Low risk