Skip to main content
. 2022 Mar 10;2022(3):CD013208. doi: 10.1002/14651858.CD013208.pub2

Goel 2017c.

Study characteristics
Patient Sampling In Zimbabwe, the performance of the SAMBA HIV‐1 Semi‐Q Test was evaluated by testing plasma specimens from adults attending the Opportunistic Infection Clinic at Harare Hospital. A total of 193 fresh specimens were tested with the SAMBA HIV‐1 Semi‐Q Test as performed on the SAMBA I platform and with the Roche CAP/CTM v2.0 assay at the National Reference Microbiology Laboratory in Harare.
Patient characteristics and setting Samples from HIV‐positive adults attending an HIV clinic in Harare, Zimbabwe.
Index tests SAMBA HIV‐1 Semi‐Q Test on SAMBA I platform on fresh plasma samples in a central reference laboratory.
Target condition and reference standard(s) High HIV viral load > 1000 copies/mL; Roche CAP/CTM v2.0 assay and Abbott HIV‐1 RealTime Assay (discrepant results). 7 discrepant results original Roche testing: 6 discrepant results were concordant with tie‐breaker testing (Abbott = Roche); 1+ Roche and 1‐ SAMBA. Results of the reference test were based on tie‐breaker results but unlikely to introduce bias as it was only one differing result.
Flow and timing A total of 193 fresh specimens were tested with the SAMBA HIV‐1 Semi‐Q Test as performed on the SAMBA I platform and with the Roche CAP/CTM v2.0 assay at the National Reference Microbiology Laboratory in Harare. There were 7 discrepant results original Roche testing: 6 discrepant results were concordant with tie‐breaker testing (Abbott = Roche); 1+ Roche and 1‐ SAMBA.
Comparative  
Notes  
Methodological quality
Item Authors' judgement Risk of bias Applicability concerns
DOMAIN 1: Patient Selection
Was a consecutive or random sample of patients enrolled? Unclear    
Was a case‐control design avoided? Yes    
Did the study avoid inappropriate exclusions? Unclear    
Could the selection of patients have introduced bias?   Unclear risk  
Are there concerns that the included patients and setting do not match the review question?     Unclear
DOMAIN 2: Index Test (All tests)
Were the index test results interpreted without knowledge of the results of the reference standard? Unclear    
If a threshold was used, was it pre‐specified? Yes    
Could the conduct or interpretation of the index test have introduced bias?   Unclear risk  
Are there concerns that the index test, its conduct, or interpretation differ from the review question?     High
DOMAIN 3: Reference Standard
Is the reference standards likely to correctly classify the target condition? Yes    
Were the reference standard results interpreted without knowledge of the results of the index tests? Unclear    
Could the reference standard, its conduct, or its interpretation have introduced bias?   Unclear risk  
Are there concerns that the target condition as defined by the reference standard does not match the question?     Low concern
DOMAIN 4: Flow and Timing
Was there an appropriate interval between index test and reference standard? Unclear    
Did all patients receive the same reference standard? No    
Were all patients included in the analysis? Yes    
Could the patient flow have introduced bias?   Unclear risk