Skip to main content
. 2022 Mar 10;2022(3):CD013208. doi: 10.1002/14651858.CD013208.pub2

Mor 2015.

Study characteristics
Patient Sampling Study included samples collected between 2013 and 2014 from patients (confirmed HIV‐1 carriers) attending the AIDS clinic of the Sheba Medical Center. Plasma from whole blood samples collected in EDTA‐containing tubes was separated by centrifugation (1100 g for 5 min). Plasma samples (n = 404) spanning the full range of HIV‐1 viral loads were selected based on the NUCLISENS v2.0 results.
Patient characteristics and setting Samples from HIV‐1 positive patients (some on ART, others not); Sheba clinic in Israel
Index tests The Xpert HIV‐1 Viral Load assay on the GeneXpert platform (Cepheid Inc); frozen plasma samples in a laboratory
Target condition and reference standard(s) High HIV viral load results > 40 copies/mL, Abbott RealTime HIV‐1 assay
Flow and timing Plasma from whole blood samples collected in EDTA‐containing tubes was separated by centrifugation (1100 g for 5 min). An aliquot (0.5 mL) was initially tested with the NUCLISENS v2.0 assay as part of the regular monitoring of HIV‐1 copy numbers. Separate aliquots of the plasma samples were stored frozen at 20 °C in volumes required for the different assays, with a single freeze‐thaw cycle prior to analysis on the different platforms. On the day of analysis, the aliquots were thawed.
Comparative  
Notes  
Methodological quality
Item Authors' judgement Risk of bias Applicability concerns
DOMAIN 1: Patient Selection
Was a consecutive or random sample of patients enrolled? Unclear    
Was a case‐control design avoided? Yes    
Did the study avoid inappropriate exclusions? Unclear    
Could the selection of patients have introduced bias?   Unclear risk  
Are there concerns that the included patients and setting do not match the review question?     High
DOMAIN 2: Index Test (All tests)
Were the index test results interpreted without knowledge of the results of the reference standard? Unclear    
If a threshold was used, was it pre‐specified? Yes    
Could the conduct or interpretation of the index test have introduced bias?   Unclear risk  
Are there concerns that the index test, its conduct, or interpretation differ from the review question?     High
DOMAIN 3: Reference Standard
Is the reference standards likely to correctly classify the target condition? Yes    
Were the reference standard results interpreted without knowledge of the results of the index tests? Unclear    
Could the reference standard, its conduct, or its interpretation have introduced bias?   Unclear risk  
Are there concerns that the target condition as defined by the reference standard does not match the question?     Low concern
DOMAIN 4: Flow and Timing
Was there an appropriate interval between index test and reference standard? Yes    
Did all patients receive the same reference standard? Yes    
Were all patients included in the analysis? No    
Could the patient flow have introduced bias?   Low risk