Table 2.

Comparison between major existing breast photoacoustic tomography (PAT) systems.

PAM-02 Twente PAM 2 OptoSonics LOUISA-3D SBH-PACT Imagio
Transducer/
Detector Geometry
2D Matrix Array Hemisphere Spherical Arc Ring Linear Array
Receiver number 600
(20 × 30)
384 512 96 512 128
Central frequency 2 MHz 1 MHz 2 MHz 50 kHz—6 MHz frequency range 2.25 MHz 0.1–12 MHz frequency range
Bandwidth 130% 100% 70% N/A 95% N/A
Illumination Geometry Rectangles Behind and in-front of PA detector 9 fiber bundles Cone beam Arc Donut Two Rectangular Windows
Laser Wavelengths 700–900 nm 755 and 1064 nm 750–800 nm 757, 797 nm 750, 1064 nm 757, 1064 nm
Resolution 1 mm 1.06 mm (lateral)
0.96 mm (elevational)
0.42 mm −0.3 mm 0.255 mm 0.73–0.81 mm (lateral)
0.42–0.47 mm
(axial)
Volumetric Imaging Yes, via rotation Yes Yes Yes, via rotation Yes, via linear scanning No
Breast Compression Yes No No No Yes No
Co-registered with US Yes No No Yes No Yes
FDA Status N/A N/A N/A N/A N/A Approved