Performance of the CRISIS Ventilator with a set Respiratory System compliance of 30mL/cmH2O. Here the tolerances with an adult lung with severely decreased compliance for a given inflow rate is demonstrated. Listed resultant values represent Mean ± the standard deviations.

Set Flow Rate (L/min) Goal Tidal Volume (mL) Goal Respiratory Rate Measured Tidal Volume (mL) Measured Respiratory Rate PIP (cmH2O) PEEP (cmH2O)
20 250 Low 254.1 ± 8.3 19.2 ± 1.8 12.25 ± 0.2 2.96 ± 0.2
High 250.7 ± 2.1 31.2 ± 3.2 18.63 ± 2.3 7.28 ± 1.5
350 Low 347.6 ± 2.9 14.6 ± 1.7 15.73 ± 0.4 3.38 ± 0.4
High 351.1 ± 3.3 22.4 ± 3.5 27.32 ± 4.4 10.40 ± 2.7
450 Low 449.4 ± 3.8 12.1 ± 1.0 19.60 ± 0.2 3.98 ± 0.6
High 447.7 ± 5.6 19.5 ± 3.5 28.72 ± 4.0 9.59 ± 2.3
550 Low 550.1 ± 2.6 10.1 ± 0.2 23.41 ± 0.6 4.73 ± 1.1
High 547.0 ± 2.7 16.1 ± 4.0 32.82 ± 2.8 10.49 ± 1.7
650 Low 649.6 ± 0.9 9.9 ± 0.4 28.45 ± 2.5 5.34 ± 1.1
High 646.7 ± 2.1 13.7 ± 2.4 35.42 ± 5.8 11.24 ± 2.0
30 250 Low 247.0 ± 4.7 24.0 ± 2.0 13.80 ± 0.5 3.98 ± 0.5
High 248.2 ± 8.8 38.0 ± 2.3 17.03 ± 1.2 7.05 ± 0.9
350 Low 350.9 ± 1.3 17.5 ± 1.2 17.79 ± 0.8 4.40 ± 0.8
High 344.2 ± 6.7 30.4 ± 1.9 28.30 ± 5.1 12.17 ± 3.4
450 Low 448.0 ± 5.1 16.6 ± 2.3 26.08 ± 6.1 5.04 ± 0.7
High 448.3 ± 6.7 26.9 ± 2.7 31.43 ± 3.3 13.23 ± 2.6
550 Low 554.9 ± 3.2 12.0 ± 0.4 25.43 ± 0.8 5.70 ± 0.9
High 551.2 ± 4.4 20.9 ± 2.9 37.71 ± 1.7 14.05 ± 1.6
650 Low 649.6 ± 1.6 11.3 ± 0.6 28.95 ± 0.9 6.30 ± 1.0
High 655.7 ± 3.4 18.4 ± 2.1 40.17 ± 3.9 13.72 ± 2.2