Table 4.

Performance of the CRISIS ventilator with a set respiratory system compliance of 100mL/cmH2O (Healthy Lung). Here the tolerances for a healthy adult lung for a given inflow rate is demonstrated. Listed resultant values represent MEAN ± the standard deviations.

Set Flow Rate (L/min) Goal Tidal Volume (mL) Goal Respiratory Rate Measured Tidal Volume (mL) Measured Respiratory Rate PIP (cmH2O) PEEP (cmH2O)
20 250 Low 257.0 ± 13.2 16.2 ± 1.1 8.3 ± 1.7 4.5 ± 1.3
High 250.8 ± 7.1 18.7 ± 1.6 7.8 ± 1.6 4.2 ± 1.5
350 Low 342.6 ± 24.1 14.4 ± 1.0 8.5 ± 0.7 4.0 ± 0.8
High 350.3 ± 10.6 15.8 ± 0.8 10.9 ± 1.0 5.6 ± 0.6
450 Low 446.9 ± 18.0 10.5 ± 1.7 8.4 ± 0.3 3.1 ± 0.1
High 447.2 ± 4.5 12.3 ± 2.0 13.0 ± 1.2 6.2 ± 0.6
550 Low 545.4 ± 15.9 8.6 ± 1.3 9.6 ± 0.5 3.2 ± 0.4
High 553.3 ± 2.9 11.2 ± 2.0 14.7 ± 1.7 6.5 ± 0.8
650 Low 658.4 ± 15.7 7.9 ± 2.0 12.0 ± 2.8 4.2 ± 1.8
High 664.6 ± 5.5 10.0 ± 1.8 14.4 ± 2.8 5.8 ± 1.6
30 250 Low 269.2 ± 24.1 19.5 ± 3.4 9.4 ± 1.3 5.7 ± 1.3
High 253.9 ± 21.4 23.2 ± 2.5 8.7 ± 0.2 5.0 ± 0.1
350 Low 348.6 ± 7.3 15.0 ± 4.0 10.4 ± 3.2 4.9 ± 2.5
High 339.5 ± 11.4 20.5 ± 1.5 10.4 ± 0.3 5.6 ± 0.4
450 Low 445.5 ± 5.5 11.8 ± 4.0 10.3 ± 0.3 4.2 ± 0.5
High 446.4 ± 7.4 16.8 ± 1.0 13.2 ± 1.7 6.8 ± 1.1
550 Low 555.8 ± 8.3 9.3 ± 1.3 11.0 ± 0.6 3.9 ± 0.5
High 541.1 ± 15.5 15.1 ± 1.3 15.3 ± 2.2 7.5 ± 1.4
650 Low 646.5 ± 3.7 8.3 ± 0.7 12.0 ± 0.6 3.9 ± 0.3
High 641.9 ± 10.8 14.0 ± 1.7 17.5 ± 1.5 8.2 ± 1.1