Table 5:
E1–E3 nursing performance during simulation
Table E1: nursing completion of key practice guidelines for impella patient during simulation sessions (n=9) | |||||
Task | Complete | Incomplete | |||
Two-dimensional (2D) echo verification of placement | |||||
Simulation #1 | 11% (1) | 89% (8) | |||
Simulation #2 | 100% (9) | 0% (0) | |||
CM marking verified | |||||
Simulation #1 | 44% (4) | 55% (5) | |||
Simulation #2 | 100% (9) | 0% (0) | |||
Tightening of Toughy Borst valve | |||||
Simulation #1 | 0% (0) | 100% (9) | |||
Simulation #2 | 78% (7) | 22% (2) | |||
Transfer to standard configuration | |||||
Simulation #1 | 89% (8) | 11% (1) | |||
Simulation #2 | 89% (8) | 11% (1) | |||
Purge solution verified and hung | |||||
Simulation #1 | 0% (0) | 100% (9) | |||
Simulation #2 | 78% (7) | 22% (2) |
Table E2: average time to complete critical tasks for Impella patient during simulation (n=9) | |||||
Activity | Simulation 1 | Simulation 2 | |||
2D echo verification | 5 min 32 s | 1 min 32 s | |||
CM marking | 5 min 13 s | 3 min 55 s | |||
Standard configuration | 4 min 3 s | 4 min 20 s | |||
Purge solution verified | No groups fully completed* | 2 min 18 s | |||
Purge solution hung | No groups fully completed* | 3 min 13 s |
Table E3: nursing recognition and management of alarms for impella patient during simulation sessions (n=9) | |||||
Recognised alarm (suction, aorta/ventricle) |
Correctly identified cause of alarm | Correct identification of action(s) to resolve alarm | Notified physician with recommendation | Notified physician without recommendation | |
Simulation #1 | 100% (9) | 33% (3) (suction alarms only) |
33% (3) (suction alarms only) |
33% (3) (suction alarms only) |
67% (6) |
Simulation #2 | 100% (9) | 100% (9) | 100% (9) | 100% (9) | 0% (0) |
*During simulation #1, all groups partially completed the task of verifying the purge solution change to dextrose plus heparin by simply hanging the solution. However, none of groups changed the information in the Impella console; therefore, the console remained programmed as having only dextrose infusing as the purge solution and would not recognise or calculate the amount of heparin in use and delivered to the patient.