Figure 2.

Figure 2

Liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS). The experiment begins with a protein separation by liquid chromatography (LC) followed by ionization and mass spectrometry analysis. Triple quadrupole mass spectrometer comprises of three quadrupoles Q1, Q2, and Q3. Quadrupoles Q1 and Q3 serve as mass filters. The middle quadrupole Q2 functions as a collision cell. The specific precursor ions isolated by Q1 and accumulated in Q2 collide with neutral gas molecules, such as nitrogen generating the product ions. The product ions reach the third quadrupole assembly Q3. The quadrupole Q1 allows ions with a specific m/z ratio range to pass to the detector (D). P—pump; C –column; ES—electrospray, a liquid-to-aerosol converting device.