Skip to main content
. 2022 Feb 25;12(3):142. doi: 10.3390/bios12030142

Table 1.

Comparison of the analytical performance of the CSPE/CNF–AuNP/Aptamer with the other aptasensors and immunosensors for the biomarkers related to SARS-CoV-2.

Biosensor Biomarker Method Linear Range Limit of Detection Response Time Ref.
Carbon electrode/Graphene oxide/SP RBD SP (78.3 kDa) SWV 1–1000 ng/mL 0.11 ng/mL 45 min [44]
CB/CSPE combined with Magnetic bead-based immunosensor 2019-nCoV-NP (51.1 kDa) DPV 0.19–11.7 nM
(0.01–0.6 μg/mL)
0.15 nM
(8 ng/mL)
30 min [11]
SP (78.3 kDa) DPV 0.04–10 μg/mL 19 ng/mL 30 min
Gold electrode/Aptamer combined with a labeled aptamer 2019-nCoV-NP (51.1 kDa) DPV 0.00048–0.97 nM
(0.025–50 ng/mL)
(0.0048–0.97 nM)
0.00016 nM (8.33 pg/mL) 60 min [23]
Pad/Aptamer 2019-nCoV-NP (51.1 kDa) SPR 0.5–16 ng/mL 1 ng/mL (20 pM) 110 min [21]
ITO/gC3N4-CdS/Chitosan-Aptamer SARS-CoV-2-RBD (35 kDa) PEC 0.5–32 nM
17.5–1120 nM
0.12 nM
4.2 ng/mL
40 min [18]
CSPE/Graphene/Spike IgG antibody 0.25 fg/mL–1 ng/mL 0.25 fg/mL 5 [12]
CSPE/CNF–AuNP/Aptamer SARS-CoV-2-RBD (35 KDa) EIS 0.01–64 nM
(0.35–2240 ng/mL)
0.007 nM
(0.24 ng/mL)
40 min This work

SP: Spike protein; CB/CSPE: Carbon black-modified screen-printed electrode; ITO: Indium Tin Oxide electrode;g-C3N4: Graphitic carbon nitride; CdS: Cadmium sulfide; PEC: Photo electrochemical method; SPR: Surface plasmon resonance; SWV: Square-wave voltammetry; 2019-nCoV-NP: SARS-CoV-2 nucleocapsid protein; DPV: Differential pulse voltammetry.