Table 3.

Proposed MRI protocol for patients with malignant pleural mesothelioma. TSE: turbo spin echo; HASTE: half-Fourier acquired single shot turbo spin-echo; VIBE: Volumetric interpolated breath-hold examination; DWI: diffusion weighted imaging; SPAIR: spectral attenuated inversion recovery; FOV: field of view; TR: repetition time; TE: echo time; FA: flip angle.

Sequence Manufacturer Acronyms Typical Contrast Average Acquisition Time Spatial Resolution Scan Plane Scan Parameters Field Strength B0
Morphology
2D Echo-planar Fast Spin Echo Sequence HASTE
(SIEMENS)
T2-weighted Expiration Breath-Hold
≈ 40 s (in 2 different breath-hold)
FOV = 440 mm
Thickness = 5 mm
Plane = Coronal
TR = 2860 ms
TE = 93 ms
FA = 160 deg
Matrix = (182 × 256)
1.5T
2D Echo-planar Fast Spin Echo Sequence HASTE
(SIEMENS)
T2-weighted Expiration Breath-Hold
≈ 40 s (in 2 different breath-hold)
FOV = 420 mm
Thickness = 5 mm
Plane = Sagittal
TR = 2860 ms
TE = 93 ms
FA = 160 deg
Matrix = (182 × 256)
1.5T
2D Echo-planar Fast Spin Echo Sequence HASTE
(SIEMENS)
T2-weighted Expiration Breath-Hold
≈ 40 s (in 2 different breath-hold)
FOV = 360 mm
Thickness = 5 mm
Plane = Axial
TR = 2860 ms
TE = 93 ms
FA = 160 deg
Matrix = (170 × 256)
1.5T
2D Echo-planar Fast Spin Echo Sequence
(Fat Saturated)
HASTE
(SIEMENS)
T2-weighted
(Fat saturated)
Expiration Breath-Hold
≈ 40 s (in 2 different breath-hold)
FOV = 440 mm
Thickness = 5 mm
Plane = Coronal
TR = 2860 ms
TE = 93 ms
FA = 160 deg
Matrix = (182 × 256)
1.5T
2D Echo-planar Fast Spin Echo Sequence
(Fat Saturated)
HASTE
(SIEMENS)
T2-weighted
(Fat saturated)
Expiration Breath-Hold
≈ 40 s (in 2 different breath-hold)
FOV = 440 mm
Thickness = 5 mm
Plane = Axial
TR = 2860 ms
TE = 93 ms
FA = 160 deg
Matrix= (170 × 256)
1.5T
2D Turbo Spin Echo
(TSE)
T1-weighted Free breathing
(Navigator)
≈ 2:30 min
FOV = 440 mm
Thickness = 5 mm
Plane = Coronal
TE =491 ms
TE = 19 ms
FA = 135 deg
Matrix = (179 × 256)
1.5T
3D Rapid Acquisition Spoiled Gradient Echo VIBE
(SIEMENS)
T1-weighted
(Fat saturated)
Expiration Breath-Hold
≈ 26 s
FOV = 390 mm
Thickness = 2.5 mm
Plane = Axial
TR = 4.55 ms
TE = 2 ms
FA = 10 deg
Matrix = (153 × 224)
1.5T
DWI
2D Single-Shot Echo Planar Imaging (EPI) DWI T2-weighted diffusion-weighted
(SPAIR fat saturation)
Free breathing (Navigator)
7–10 min
FOV = 360 mm
Thickness = 5 mm
Plane = Axial
TR = 11,500
TE = 90
BW = 1371
B = 0, 50, 400, 800 s/mm2
Matrix = (148 × 192)
1.5T
Contrast Gadolinium Enhanced
3D Rapid Acquisition Spoiled Gradient Echo
Post-contrast
(40 s, 80 s, 3′, 4:30′, 5′)
VIBE
(SIEMENS)
T1-weighted
(Fat saturated)
Expiration Breath-Hold
≈ 26 s
FOV = 390 mm
Thickness = 2.5 mm
Plane = Axial
TR = 4.55 ms
TE = 2 ms
FA = 10 deg
Matrix = (153 × 224)
1.5T