. 2022 Feb 26;12(3):598. doi: 10.3390/diagnostics12030598

Table 1.

Anterior corneal astigmatism difference between the two devices.

Biometer A
(179 Eyes)
Biometer B
(179 Eyes)
A–B p *
Astigmatism magnitude (D) 1.25 ± 0.88 1.27 ± 0.85 −0.02 ± 0.41 0.439
Axis (degree) 75 ± 50 85 ± 56 −11 ± 54 0.009 #
J0 −0.05 ± 0.51 −0.03 ± 0.57 −0.01 ± 0.82 0.841
J45 0.03 ± 0.57 0.03 ± 0.51 0.01 ± 0.75 0.902

Note: A = ANTERION; B = IOLMaster 700; D = diopter. * Calculated using paired sample t-test. # Values with a significance level of 5% are set in bold.