Table 1.

MRI acquisition parameters for the “low-SNR” and “high-SNR” multi-echo spin-echo sequences. Please note that although 10 echo times were initially sampled, only the first 7 echoes were used for the T2 fitting because of insufficient SNRs in the last echoes. The choice of echo times was guided by the Osteoarthritis Initiative study [39]. Abbreviations: MESE = Multi-echo spin-echo, SENSE = Sensitivity Encoding.

“Low-SNR” Sequence “High-SNR” Sequence
Sequence Type 2D MESE
Orientation Axial
Repetition Time [ms] 500
Echo Times [ms] {10+n10| n=0,1,..6}
Field of View [mm] 140 × 140
Acquisition Matrix 512 × 512
Reconstruction Matrix 512 × 512
Scan percentage [%] 100
Flip angle [°] 90
Number of Signal Averages [n] 1 4
SENSE Factor 3 1
Slices [n] 1
Slice Thickness [mm] 2
Duration [min:s] 2:15 26:34