Skip to main content
. 2022 Mar 22;14:493–516. doi: 10.2147/NSS.S348795

Table 1.

Sleep-Wake Summary Agreement

Outcome Device n Device Mean (SD) Reference Mean (SD)
TIB Fatigue Science Readiband 112 459.0 (85.2) 437.4 (78.9)
Fitbit Inspire HR 106 459.8 (78.6) 443.3 (73.6)
Oura Ring 111 460.0 (90.3) 439.9 (74.0)
Polar Vantage V Titan 103 452.0 (75.6) 445.5 (69.4)
TST Actiwatch 113 406.1 (77.4) 401.3 (82.6)
Fatigue Science Readiband 112 406.0 (89.4) 401.0 (82.9)
Fitbit Inspire HR 106 406.4 (70.1) 406.6 (78.8)
Oura Ring 111 412.5 (77.5) 403.1 (78.8)
Polar Vantage V Titan 103 418.5 (74.0) 411.6 (72.1)
SE Actiwatch 113 92.4 (3.2) 91.2 (6.9)
Fatigue Science Readiband 112 88.0 (7.5) 91.3 (7.0)
Fitbit Inspire HR 106 94.6 (2.9) 91.3 (7.0)
Oura Ring 111 89.8 (4.9) 91.2 (7.0)
Polar Vantage V Titan 103 92.5 (4.6) 92.2 (4.6)
SOL Actiwatch 113 2.3 (1.7) 8.5 (5.6)
Fatigue Science Readiband 112 18.8 (17.3) 8.5 (5.6)
Fitbit Inspire HR 106 4.9 (3.4) 8.6 (5.6)
Oura Ring 111 7.8 (5.5) 8.5 (5.6)
Polar Vantage V Titan 103 0.6 (1.0) 8.5 (5.4)
WASO Actiwatch 113 30.2 (12.7) 28.1 (22.5)
Fatigue Science Readiband 112 21.2 (23.8) 27.9 (22.5)
Fitbit Inspire HR 106 38.0 (13.1) 28.1 (21.9)
Oura Ring 111 39.7 (32.0) 28.4 (22.5)
Polar Vantage V Titan 103 25.8 (8.6) 25.4 (17.5)

Notes: Sleep-wake summary agreement between devices and reference. The sleep diary served as reference for TIB, and the Dreem headband served as reference for TST, SE, SOL, and WASO. All nights with available data for both the device and reference were included in analyses, with the total number of nights (n) indicated in each row. Mean and SD are shown for each device and reference.

Abbreviations: SD, standard deviation; SE, sleep efficiency; SOL, sleep onset latency; TIB, time in bed; TST, total sleep time; WASO, wake after sleep onset.