Published in final edited form as: Exp Eye Res. 2022 Jan 22;217:108937. doi: 10.1016/j.exer.2022.108937

Table 5.

The absolute and percentage differences between calibrated LenStar and ultrasound values for anterior chamber depth (ACD), lens thickness (LT), vitreous chamber depth (VCD), and axial length (AL) for the validation arm data.

Mean ± SD difference
calibrated LenStar –
ultrasound) (mm)
95% LoA Mean ± SD
percentage difference
in relation to ultrasound
95% CIs in percentage
difference
Lower
bound
Upper
bound
Lower
bound
Upper
bound
ACD −0.005 ± 0.054 −0.110 0.101 −0.048 ± 1.820 −0.153 0.057
LT 0.004 ± 0.074 −0.142 0.150 −0.034 ± 2.021 −0.151 0.082
VCD −0.010 ± 0.089 −0.184 0.165 −0.052 ± 0.903 −0.104 0.0003
AL −0.004 ± 0.094 −0.187 0.180 −0.027 ± 0.566 −0.060 0.005