Published in final edited form as: Exp Eye Res. 2022 Jan 22;217:108937. doi: 10.1016/j.exer.2022.108937

B).

Difference between retrospectively predicted LenStar value and analytic arm LenStar data. The predictive values (not shown) are calculated by fitting the validation arm ultrasound values into linear equations created using the parameters in Appendix 4A.

Mean ± SD difference
(Predictive LenStar –
Measured LenStar)
(mm)
95% LoA Mean ± SD
percentage difference
95% CIs in percentage
difference
Lower
bound
Upper
bound
Lower
bound
Upper
bound
ACD 0.001 ± 0.047 −0.092 0.094 0.053 ± 1.716 −0.046 0.152
LT 0.002 ± 0.068 −0.132 0.136 0.066 ± 1.916 −0.045 0.176
VCD 0.006 ± 0.092 −0.174 0.185 0.064 ± 0.964 0.008 0.120
AL 0.007 ± 0.097 −0.183 0.198 0.047 ± 0.610 0.012 0.083