Table 2.
Calculated recovery coefficient and contrast recovery and background noise for the images of NEMA IQ phantom for different image reconstruction and radionuclide combinations.
| Sphere size | Radionuclide | Recovery coefficient | Contrast recovery [%] | |||
|---|---|---|---|---|---|---|
| OSEM | PRC | Relative difference [%] | OSEM | PRC | ||
| 10 mm | 18F | 0.54 | 0.53 | −2.34 | 42.8 | 41.2 |
| 68Ga | 0.37 | 0.49 | 32.51 | 28.4 | 42.2 | |
| 124I | 0.39 | 0.53 | 36.52 | 23.8 | 41.6 | |
| 13 mm | 18F | 0.62 | 0.61 | −1.8 | 52.5 | 51.1 |
| 68Ga | 0.56 | 0.68 | 21.5 | 49.3 | 63.0 | |
| 124I | 0.50 | 0.63 | 25.0 | 38.1 | 53.9 | |
| 17 mm | 18F | 0.73 | 0.72 | −1.8 | 66.6 | 64.9 |
| 68Ga | 0.61 | 0.68 | 10.1 | 55.8 | 62.9 | |
| 124I | 0.64 | 0.73 | 15.4 | 54.5 | 66.8 | |
| 22 mm | 18F | 0.82 | 0.80 | −1.7 | 77.2 | 75.5 |
| 68Ga | 0.67 | 0.73 | 7.7 | 62.7 | 68.6 | |
| 124I | 0.75 | 0.81 | 8.2 | 68.8 | 76.5 | |
| 28 mm | 18F | 0.84 | 0.81 | −2.9 | 79.6 | 76.6 |
| 68Ga | 0.72 | 0.79 | 8.7 | 68.5 | 75.8 | |
| 124I | 0.77 | 0.83 | 7.8 | 71.5 | 79.1 | |
| 37 mm | 18F | 0.87 | 0.85 | −1.8 | 83.9 | 81.9 |
| 68Ga | 0.78 | 0.82 | 5.5 | 75.1 | 80.0 | |
| 124I | 0.85 | 0.88 | 4.4 | 80.7 | 85.3 | |
| Background noise [%] | ||||||
| 18F | 10.1 | 9.30 | −8.3 | |||
| 68Ga | 17.7 | 15.7 | −11.0 | |||
| 124I | 9.7 | 9.35 | −3.8 | |||
18F images (both OSEM and PRC) were reconstructed with 3 iterations and 12 subsets, while 68Ga and 124I images with 2 iterations and 12 subsets for OSEM and 8 iterations as 12 subsets for PRC.