Table 4. Clinical evaluation results.
| Measures | Group | Test results, Median (IQR) | P value | Post-hoc analysisa | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Unaided | LiNX 3D (premium HA) | Nevara1 (basic HA) | CS50+ (premium PSAP) | Bean (basic PSAP) | Unaided vs. LiNX 3D | Unaided vs. Nevara1 | Unaided vs. CS50+ | Unaided vs. Bean | |||
| Sound-filed audiometry (dB HL) | MHL | 40.625 (38.75, 43.75) | 31.25 (28.75, 35) | 36.625 (32.5, 40) | 31.875 (30, 36.25) | 30.625 (28.75, 32.5) | < 0.001*** | < 0.001*** | < 0.001*** | 0.002** | < 0.001*** |
| MDHL | 48.75 (45, 55) | 35 (32.4, 40) | 45 (41.25, 50) | 40.625 (35, 45) | 37.5 (33.75, 42.5) | < 0.001*** | < 0.001*** | < 0.001*** | < 0.001*** | < 0.001*** | |
| MSHL | 65 (58.75, 71.25) | 45 (41.25, 51.25) | 57.5 (53.75, 63.75) | 48.75 (42.5, 53.75) | 47.5 (45. 55) | < 0.001*** | < 0.001*** | < 0.001*** | < 0.001*** | < 0.001*** | |
| WRS (%) | MHL | 92 (92, 96) | 96 (92, 96) | 92 (92, 96) | 92 (84, 96) | 96 (96, 100) | 0.005** | 0.024* | < 0.001*** | 1.000 | 0.177 |
| MDHL | 88 (80, 92) | 92 (84, 96) | 88 (84, 92) | 84 (80, 96) | 92 (84, 96) | 0.001** | < 0.001*** | < 0.001*** | 1.000 | 0.027* | |
| MSHL | 60 (24, 68) | 88 (52, 92) | 76 (40, 84) | 68 (52, 88) | 80 (52, 92) | < 0.001*** | < 0.001*** | < 0.001*** | 0.004** | < 0.001*** | |
| K-HINT (SNR) | MHL | −0.7 (−1.8, 0.3) | −0.6 (−1.6, 1) | 1.4 (−0.3, 2.4) | −0.2 (−1.6, 0.5) | −0.5 (−1.1, 0.3) | 0.053 | 0.002** | < 0.001*** | < 0.001*** | < 0.001*** |
| MDHL | 1 (−0.1, 2.8) | 0.9 (−0.7, 3.8) | 1.4 (−0.1, 2.8) | −0.1 (−1.8, 1.9) | −0.4 (−1.3, 1.3) | 0.007** | > 0.99 | > 0.99 | 0.021* | 0.011* | |
| MSHL | 4.7 (2.6, 12.1) | 5.1 (1.8, 10) | 3.3 (2.2, 7.4) | 3 (0.7, 10.8) | 3.3 (0.8, 6) | 0.002** | 0.202 | 0.002** | 0.179 | 0.001** | |
IQR = inter-quartile range, HA = hearing aid, PSAP = personal sound amplification product, MHL = mild hearing loss, MDHL = moderate hearing loss, MSHL = moderately severe hearing loss, WRS = word recognition score, K-HINT = Korean version of the Hearing In Noise Test.
aThe comparison result of each device was corrected by the Bonferroni correction method for type I error increase by multiple comparisons.
*P < 0.05, **P < 0.01, ***P < 0.001.